ELECTRICALLY HEATED WIRE-LOOP, IN-TORCH VAPORIZATION SAMPLE INTRODUCTION SYSTEM FOR INDUCTIVELY-COUPLED PLASMA-ATOMIC EMISSION-SPECTROMETRY WITH PHOTODIODE-ARRAY DETECTION

被引:15
作者
KARANASSIOS, V [1 ]
BATEMAN, KP [1 ]
SPIERS, GA [1 ]
机构
[1] UNIV GUELPH,GUELPH N1G 2W1,ON,CANADA
基金
加拿大自然科学与工程研究理事会;
关键词
D O I
10.1016/0584-8547(94)80075-8
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
A new sample introduction system that can be used for the analysis of microliter volumes of liquids by inductively coupled plasma atomic emission spectrometry (ICP-AES) is described. The system consists of an electrically heated W wire-loop which is attached to a thermocouple ceramic insulator. Ceramic and wire-loop are inserted manually into the central tube of a modified ICP torch, with the top of the wire-loop positioned about 10 cm below the plasma. An inexpensive variac, which plugs into a wall socket, is used as a power supply. This in-torch vaporization (ITV) sample introduction system was characterized using an ICP optical emission spectrometer with a photodiode array detector. In this pilot study aimed at testing the validity of the approach, spectral interference effects arising from W emission lines are documented for Be, Cu, Mn, Sc, Sr, V, Y and Zn and preliminary analytical performance characteristics using an unoptimized wire-loop/torch design are presented. Calibration curves were linear for over three orders of magnitude for Sr, Y and Be; the non-linear calibration curves obtained for Cu were attributed to contamination from the power transfer cables, for Mg to contamination from the ceramic and for Ca to airborne particles adhering to the wire-loop. As well, Au-coated W wire-loops proved to bp sensitive to Hg-0-vapor in air. Precision varied between 1.9 and 4.5% (10 mu l injection of single element standards) and, estimated absolute detection limits were (in pg) Zn (710), V (20), Mn (10), Y (10), Sc (9), Be (1) and Sr (0.4).
引用
收藏
页码:847 / 865
页数:19
相关论文
共 65 条
[21]   ELECTROTHERMAL VAPORIZATION SAMPLE INTRODUCTION SYSTEM FOR THE INDUCTIVELY COUPLED PLASMA [J].
HULL, DR ;
HORLICK, G .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1984, 39 (06) :843-850
[22]  
Ingle J.D., 1988, SPECTROCHEMICAL ANAL
[23]   INTERPRETING SOME ANALYTICAL CHARACTERISTICS OF THERMAL DISPERSION METHODS USED FOR SAMPLE INTRODUCTION IN ATOMIC SPECTROMETRY [J].
KANTOR, T .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1988, 43 (9-11) :1299-1320
[24]   DETERMINATION OF MERCURY-VAPOR IN AIR USING ELECTRICALLY HEATED, GOLD-COATED WIRE-LOOPS, IN-TORCH VAPORIZATION SAMPLE INTRODUCTION AND INDUCTIVELY-COUPLED PLASMA-ATOMIC EMISSION-SPECTROMETRY WITH PHOTODIODE-ARRAY DETECTION [J].
KARANASSIOS, V ;
BATEMAN, KP ;
SPIERS, GA .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1994, 49 (09) :867-874
[25]  
KARANASSIOS V, 1990, SPECTROCHIM ACTA REV, V13, P89
[26]   SPECTRAL CHARACTERISTICS OF A NEW SPECTROMETER DESIGN FOR ATOMIC EMISSION-SPECTROSCOPY [J].
KARANASSIOS, V ;
HORLICK, G .
APPLIED SPECTROSCOPY, 1986, 40 (06) :813-821
[27]   ELIMINATION OF SOME SPECTRAL INTERFERENCES AND MATRIX EFFECTS IN INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY USING DIRECT SAMPLE INSERTION TECHNIQUES [J].
KARANASSIOS, V ;
HORLICK, G .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1989, 44 (12) :1387-1396
[28]   BACKGROUND SPECTRAL CHARACTERISTICS IN DIRECT SAMPLE INSERTION INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
KARANASSIOS, V ;
HORLICK, G .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1989, 44 (12) :1361-1385
[29]   ELECTROTHERMAL VAPORIZATION SAMPLE INTRODUCTION SYSTEM FOR THE ANALYSIS OF PELLETIZED SOLIDS BY INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION-SPECTROMETRY [J].
KARANASSIOS, V ;
REN, JM ;
SALIN, ED .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1991, 6 (07) :527-533
[30]  
KARANASSIOS V, 1993, 39TH CAN SPECTR C QU