THE VACUUM-ULTRAVIOLET SCANNING PHOTOELECTRON MICROSCOPE AT MAX-LAB

被引:26
作者
JOHANSSON, U [1 ]
NYHOLM, R [1 ]
TORNEVIK, C [1 ]
FLODSTROM, A [1 ]
机构
[1] ROYAL INST TECHNOL,S-10044 STOCKHOLM,SWEDEN
关键词
D O I
10.1063/1.1145986
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The performance of a scanning photoelectron microscope at the MAX I storage ring in Lund is presented. The microscope utilizes undulator radiation in the energy range 15-150 eV and is comprised of a plane-grating monochromator with a Kirkpatrick-Baez objective and a gracing incidence ellipsoidal focusing mirror. The instrument with its high photon flux, 109-1010 photons/s, and narrow bandwidth, better than 0.2 eV, is excellently suited for high-resolution core-level spectroscopy and imaging of samples with lateral inhomogeneties in the micrometer range. © 1995 American Institute of Physics.
引用
收藏
页码:1398 / 1400
页数:3
相关论文
共 14 条
[11]   AN ELECTROSTATIC MICROSCOPE FOR SYNCHROTRON RADIATION X-RAY ABSORPTION MICROSPECTROSCOPY [J].
TONNER, BP ;
HARP, GR ;
KORANDA, SF ;
ZHANG, J .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1992, 63 (01) :564-568
[12]   ENERGY-FILTERED IMAGING WITH ELECTROSTATIC OPTICS FOR PHOTOELECTRON MICROSCOPY [J].
TONNER, BP .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1990, 291 (1-2) :60-66
[13]   GRAZING-INCIDENCE OPTICS FOR SOFT-X-RAY MICROSCOPY [J].
VOSS, J ;
KUNZ, C ;
MOEWES, A ;
STORJOHANN, I .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1992, 63 (01) :569-573
[14]  
VOSS J, 1992, XRAY MICROSCOPY, V3, P235