EFFECT OF THE SAMPLE SOLVENT AND INSTRUMENT DESIGN ON THE REPRODUCIBILITY OF RETENTION TIMES AND PEAK SHAPES IN PACKED-COLUMN SUPERCRITICAL-FLUID CHROMATOGRAPHY

被引:7
作者
SMITH, RM
BRIGGS, DA
机构
[1] Department of Chemistry, Loughborough University of Technology, Loughborough
关键词
D O I
10.1016/0021-9673(94)80291-2
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Factors affecting the reproducibility of retention times and peak shapes in supercritical fluid chromatography on a cyano-bonded silica packed column have been studied. These included the inclusion of low proportions of modifier in the eluent and the solvent used to inject the analyte. With carbon dioxide as the eluent, polar sample solvents were found to cause residual effects, which changed subsequent separations. These effects were lost when an eluent modifier was present suggesting that they resulted from temporary masking of silanol groups on the silica surface. If the mobile phase was cooled to near the critical point between the oven and a spectroscopic detector, small changes in conditions caused baseline fluctuations, which was considered to be due to changes in the refractive index of the solution.
引用
收藏
页码:161 / 171
页数:11
相关论文
共 30 条
[21]   IMPACT OF PHASE-EQUILIBRIA ON THE BEHAVIOR OF CYLINDER-STORED CO2-MODIFIER MIXTURES USED AS SUPERCRITICAL FLUIDS [J].
SCHWEIGHARDT, FK ;
MATHIAS, PM .
JOURNAL OF CHROMATOGRAPHIC SCIENCE, 1993, 31 (06) :207-211
[22]   CHARACTERIZATION OF POLYMER-COATED C-18 PACKING MATERIALS FOR REVERSED-PHASE LIQUID-CHROMATOGRAPHY [J].
SHIROTA, O ;
OHTSU, Y ;
NAKATA, O .
JOURNAL OF CHROMATOGRAPHIC SCIENCE, 1990, 28 (11) :553-558
[23]   MODIFICATION OF CONVENTIONAL HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY EQUIPMENT FOR USE IN CARBON-DIOXIDE SUPERCRITICAL FLUID CHROMATOGRAPHY [J].
SIMPSON, RC ;
GANT, JR ;
BROWN, PR .
JOURNAL OF CHROMATOGRAPHY, 1986, 371 :109-119
[24]   RETENTION IN SUPERCRITICAL FLUID CHROMATOGRAPHY ON CYANO-BONDED SILICA COLUMNS [J].
SMITH, RM ;
COCKS, S ;
SANAGI, MM ;
BRIGGS, DA ;
EVANS, VG .
ANALYST, 1991, 116 (12) :1281-1285
[25]   RETENTION INDEXES IN SUPERCRITICAL FLUID CHROMATOGRAPHY ON A POLYSTYRENE-DIVINYLBENZENE COLUMN [J].
SMITH, RM ;
SANAGI, MM .
CHROMATOGRAPHIA, 1988, 26 :77-84
[26]  
SMITH RM, 1988, SUPERCRITICAL FLUID, pCH2
[27]  
SMITH RM, UNPUB J CHROMATOGR
[28]  
Snyder L.R., 1979, INTRO MODERN LIQUID, P298
[29]  
SNYDER LR, 1979, INTRO MODERN LIQUID, P374
[30]   GRADIENT ELUTION WITH NORMAL PHASES ON SILICA - A COMPARISON BETWEEN HIGH-PERFORMANCE LIQUID AND SUPERCRITICAL FLUID CHROMATOGRAPHY [J].
STEUER, W ;
SCHINDLER, M ;
ERNI, F .
JOURNAL OF CHROMATOGRAPHY, 1988, 454 :253-259