Speed and amplitude of lung tumor motion precisely detected in four-dimensional setup and in real-time tumor-tracking radiotherapy

被引:161
作者
Shirato, H
Suzuki, K
Sharp, GC
Fujita, K
Onimaru, R
Fujino, M
Kato, N
Osaka, Y
Kinoshita, R
Taguchi, H
Onodera, S
Miyasaka, K
机构
[1] Hokkaido Univ, Sch Med, Dept Radiol, Sapporo, Hokkaido 0608636, Japan
[2] Harvard Univ, Massachusetts Gen Hosp, Sch Med, Dept Radiat Oncol, Boston, MA USA
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 2006年 / 64卷 / 04期
关键词
four-dimensional radiotherapy; setup; real-time tracking; four-dimensional setup;
D O I
10.1016/j.ijrobp.2005.11.016
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background: To reduce the uncertainty of registration for lung tumors, we have developed a four-dimensional (4D) setup system using a real-time tumor-tracking radiotherapy system. Methods and Materials: During treatment planning and daily setup in the treatment room, the trajectory of the internal fiducial marker was recorded for 1 to 2 min at the rate of 30 times per second by the real-time tumor-tracking radiotherapy system. To maximize gating efficiency, the patient's position on the treatment couch was adjusted using the 4D setup system with fine on-line remote control of the treatment couch. Results: The trajectory of the marker detected in the 4D setup system was well visualized and used for daily setup. Various degrees of interfractional and intrafractional changes in the absolute amplitude and speed of the internal marker were detected. Readjustments were necessary during each treatment session, prompted by baseline shifting of the tumor position. Conclusion: The 4D setup system was shown to be useful for reducing the uncertainty of tumor motion and for increasing the efficiency of gated irradiation. Considering the interfractional and intrafractional changes in speed and amplitude detected in this study, intercepting radiotherapy is the safe and cost-effective method for 4D radiotherapy using real-time tracking technology. (C) 2006 Elsevier Inc.
引用
收藏
页码:1229 / 1236
页数:8
相关论文
共 22 条
[21]   A finite state model for respiratory motion analysis in image guided radiation therapy [J].
Wu, HM ;
Sharp, GC ;
Salzberg, B ;
Kaeli, D ;
Shirato, H ;
Jiang, SB .
PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (23) :5357-5372
[22]   Stereotactic radiotherapy of extracranial targets:: CT-simulation and accuracy of treatment in the stereotactic body frame [J].
Wulf, J ;
Hädinger, U ;
Oppitz, U ;
Olshausen, B ;
Flentje, M .
RADIOTHERAPY AND ONCOLOGY, 2000, 57 (02) :225-236