Stereotactic radiotherapy of extracranial targets:: CT-simulation and accuracy of treatment in the stereotactic body frame

被引:181
作者
Wulf, J [1 ]
Hädinger, U [1 ]
Oppitz, U [1 ]
Olshausen, B [1 ]
Flentje, M [1 ]
机构
[1] Univ Wurzburg, Dept Radiotherapy, D-97080 Wurzburg, Germany
关键词
stereotactic radiotherapy; body radiosurgery; stereotactic body frame; conformal radiotherapy; accuracy of treatment; quality assurance;
D O I
10.1016/S0167-8140(00)00226-7
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background and purpose: Evaluation of set-up accuracy and analysis of target reproducibility in the stereotactic body frame (SBF), designed by Blomgren and Lax from Karolinska Hospital, Stockholm. Different types of targets were analyzed for the risk of target deviation. The correlation of target deviation to bony structures was analyzed to evaluate the value of bones as reference structures for isocenter verification. Materials and methods: Thirty patients with 32 targets were treated in the SBF for primary or metastatic peripheral lung cancer, liver metastases, abdominal and pelvic tumor recurrences or bone metastases. Set-up accuracy and target mobility were evaluated by CT-simulation and port films. The contours of the target at isocenter level, bony structures and body outline were compared by matching the CT-slices for treatment planning and simulation using the stereotactic coordinates of the SBF as external reference system. The matching procedure was performed by using a 3D treatment planning program. Results: Set-up accuracy represented by bony structures revealed standard deviations (SD) of 3.5 mm in longitudinal, 2.2 mm in anterior-posterior and 3.9 mm in lateral directions. Target reproducibility showed a SD of 4.4 mm in longitudinal, 3.4 mm ap and 3.3 mm in lateral direction prior to correction. Correlation of target deviation to bones ranged from 33% (soft tissue targets) to 100% (bones). Conclusion: A security margin of 5 mm for PTV definition is sufficient, if CT simulation is performed prior to each treatment to correct larger target deviations or set-up errors. Isocenter verification relative to bony structures is only safe for bony targets but not for soft tissue targets. (C) 2000 Elsevier Science Ireland Ltd. All rights reserved.
引用
收藏
页码:225 / 236
页数:12
相关论文
共 28 条
[1]   Prostate target volume variations during a course of radiotherapy [J].
Antolak, JA ;
Rosen, II ;
Childress, CH ;
Zagars, GK ;
Pollack, A .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1998, 42 (03) :661-672
[2]   Improvement of CT-based treatment-planning models of abdominal targets using static exhale imaging [J].
Balter, JM ;
Lam, KL ;
McGinn, CJ ;
Lawrence, TS ;
Ten Haken, RK .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1998, 41 (04) :939-943
[3]   Uncertainties in CT-based radiation therapy treatment planning associated with patient breathing [J].
Balter, JM ;
TenHaken, RK ;
Lawrence, TS ;
Lam, KL ;
Robertson, JM .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1996, 36 (01) :167-174
[4]   Impact of cradle immobilization on setup reproducibility during external beam radiation therapy for lung cancer [J].
Bentel, GC ;
Marks, LB ;
Krishnamurthy, R .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1997, 38 (03) :527-531
[5]   MAXIMIZING SETUP ACCURACY USING PORTAL IMAGES AS APPLIED TO A CONFORMAL BOOST TECHNIQUE FOR PROSTATIC-CANCER [J].
BIJHOLD, J ;
LEBESQUE, JV ;
HART, AAM ;
VIJLBRIEF, RE .
RADIOTHERAPY AND ONCOLOGY, 1992, 24 (04) :261-271
[6]   STEREOTAXIC HIGH-DOSE FRACTION RADIATION-THERAPY OF EXTRACRANIAL TUMORS USING AN ACCELERATOR - CLINICAL-EXPERIENCE OF THE FIRST 31 PATIENTS [J].
BLOMGREN, H ;
LAX, I ;
NASLUND, I ;
SVANSTROM, R .
ACTA ONCOLOGICA, 1995, 34 (06) :861-870
[7]  
Blomgren H, 1998, Journal of Radiosurgery, V1, P63
[8]   What margins should be added to the clinical target volume in radiotherapy treatment planning for lung cancer? [J].
Ekberg, L ;
Holmberg, O ;
Wittgren, L ;
Bjelkengren, G ;
Landberg, T .
RADIOTHERAPY AND ONCOLOGY, 1998, 48 (01) :71-77
[9]   Set-up error in supine-positioned patients immobilized with two different modalities during conformal radiotherapy of prostate cancer [J].
Fiorino, C ;
Reni, M ;
Bolognesi, A ;
Bonini, A ;
Cattaneo, GM ;
Calandrino, R .
RADIOTHERAPY AND ONCOLOGY, 1998, 49 (02) :133-141
[10]   Extracranial stereotactic radiation therapy:: Set-up accuracy of patients treated for liver metastases [J].
Herfarth, KK ;
Debus, J ;
Lohr, F ;
Bahner, ML ;
Fritz, P ;
Höss, A ;
Schlegel, W ;
Wannenmacher, MF .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2000, 46 (02) :329-335