What margins should be added to the clinical target volume in radiotherapy treatment planning for lung cancer?

被引:228
作者
Ekberg, L [1 ]
Holmberg, O
Wittgren, L
Bjelkengren, G
Landberg, T
机构
[1] Malmo Univ Hosp, Dept Oncol, S-20502 Malmo, Sweden
[2] Malmo Univ Hosp, Dept Radiat Phys, S-20502 Malmo, Sweden
[3] St Lukes Hosp, Dept Phys, Dublin 6, Ireland
关键词
margins; lung cancer; patient set-up; organ movement; portal imaging; fluoroscopy;
D O I
10.1016/S0167-8140(98)00046-2
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background: The planning target volume in radiotherapy treatment planning takes into account both movements of the clinical target volume (CTV) and set-up deviations. Materials and methods: A group of patients who received radiotherapy for lung cancer were studied. In order to measure the CTV movements due to respiration and other internal organ motions, fluoroscopy was performed for 20 patients. To study the accuracy and reproducibility of patient and beam set-up, 553 electronic portal images from 20 patients were evaluated. Discrepancies between planned and actual field positions were measured and the systematic and random errors were identified. The combined effect of these geometrical variations was evaluated. Results: The average CTV movement with quiet respiration was about 2.4 mm in the medio-lateral and dorso-ventral directions. Movement in the cranio-caudal direction was on average 3.9 mm with a range of 0 - 12 mm. The systematic set-up errors were on average 2.0 mm in the transversal plane and 3.0 mm in the cranio-caudal direction. The random errors can be described by their standard deviations of 3.2 and 2.6 mm. In this study, the combined effect of the two parameters (CTV movement and set-up deviations) varied between 7.5 and 10.3 mm in different anatomical directions. Conclusions: In our daily clinical routine, we use a margin of 11 mm in the transversal plane and 15 mm cranially and caudally, also taking into account other unquantified variations and uncertainties. (C) 1998 Elsevier Science Ireland Ltd. All rights reserved.
引用
收藏
页码:71 / 77
页数:7
相关论文
共 32 条
  • [1] AALTONENBRAHME P, 1997, ACTA ONCOL S10, V36
  • [2] [Anonymous], 1993, ICRU REP, V50
  • [3] 3-DIMENSIONAL CONFORMAL RADIATION-THERAPY MAY IMPROVE THE THERAPEUTIC RATIO OF HIGH-DOSE RADIATION-THERAPY FOR LUNG-CANCER
    ARMSTRONG, JG
    BURMAN, C
    LEIBEL, S
    FONTENLA, D
    KUTCHER, G
    ZELEFSKY, M
    FUKS, Z
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1993, 26 (04): : 685 - 689
  • [4] Three dimensional planning target volumes: A model and a software tool
    AustinSeymour, M
    Kalet, I
    McDonald, J
    KromhoutSchiro, S
    Jacky, J
    Hummel, S
    Unger, J
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1995, 33 (05): : 1073 - 1080
  • [5] A VERIFICATION PROCEDURE TO IMPROVE PATIENT SET-UP ACCURACY USING PORTAL IMAGES
    BEL, A
    VANHERK, M
    BARTELINK, H
    LEBESQUE, JV
    [J]. RADIOTHERAPY AND ONCOLOGY, 1993, 29 (02) : 253 - 260
  • [6] TRANSFER ERRORS OF PLANNING CT TO SIMULATOR - A POSSIBLE SOURCE OF SETUP INACCURACIES
    BEL, A
    BARTELINK, H
    VIJLBRIEF, RE
    LEBESQUE, JV
    [J]. RADIOTHERAPY AND ONCOLOGY, 1994, 31 (02) : 176 - 180
  • [7] MAXIMIZING SETUP ACCURACY USING PORTAL IMAGES AS APPLIED TO A CONFORMAL BOOST TECHNIQUE FOR PROSTATIC-CANCER
    BIJHOLD, J
    LEBESQUE, JV
    HART, AAM
    VIJLBRIEF, RE
    [J]. RADIOTHERAPY AND ONCOLOGY, 1992, 24 (04) : 261 - 271
  • [8] A REVIEW OF ELECTRONIC PORTAL IMAGING DEVICES (EPIDS)
    BOYER, AL
    ANTONUK, L
    FENSTER, A
    VANHERK, M
    MEERTENS, H
    MUNRO, P
    REINSTEIN, LE
    WONG, J
    [J]. MEDICAL PHYSICS, 1992, 19 (01) : 1 - 16
  • [9] ROUTINE CLINICAL ONLINE PORTAL IMAGING FOLLOWED BY IMMEDIATE FIELD ADJUSTMENT USING A TELE-CONTROLLED PATIENT COUCH
    DENEVE, W
    VANDENHEUVEL, F
    DEBEUKELEER, M
    COGHE, M
    THON, L
    DEROOVER, P
    VANLANCKER, M
    STORME, G
    [J]. RADIOTHERAPY AND ONCOLOGY, 1992, 24 (01) : 45 - 54
  • [10] SIZES AND SOURCES OF FIELD PLACEMENT ERROR IN ROUTINE IRRADIATION FOR PROSTATE-CANCER
    DUNSCOMBE, P
    LOOSE, S
    LESZCZYNSKI, K
    [J]. RADIOTHERAPY AND ONCOLOGY, 1993, 26 (02) : 174 - 176