Nondestructive three-dimensional analysis of electrode to modiolus proximity

被引:48
作者
Husstedt, HW [1 ]
Aschendorff, A
Richter, B
Laszig, R
Schumacher, M
机构
[1] Univ Freiburg, Dept Neuroradiol, Freiburg, Germany
[2] Univ Freiburg, Dept Otorhinolaryngol, Freiburg, Germany
关键词
cochlear implant; cone-beam computed tomography; electrode-modiolus relationship; three-dimensional digital subtraction angiography;
D O I
10.1097/00129492-200201000-00012
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Purpose: To propose a nondestructive method for three-dimensional analysis of inner ear morphology after cochlear implantation for isolated petrous bones. Materials and Methods: After implantation of cochlear implant electrode arrays, fresh temporal bones were investigated on a new C-ami-based radiographic device permitting the generation of cross-sectional images and of three-dimensional models from multiple two-dimensional radiographic images taken under different projections (cone-beam computed tomography). Cross-sectional images and multiplanar reformations with a slice thickness of 0.15 mm were acquired. The relationship of the electrode to the modiolus was analyzed (distance between electrode and the modiolus: position of the electrode inside the tympanic or vestibular scale). Histologic preparation was used as a gold standard. Results: In all cases, cone-beam computed tomography gave similar information concerning the position of the electrode compared with histologic analysis (tympanic scale versus vestibular scale, proximity versus distance to the modiolus). Perforation of the electrode from the tympanic to the vestibular scale could be assessed three-dimensionally in the cross-sectional images. In contrast to histology, cross-sectional imaging based on radiography is performed in less than 10 minutes and needs no preparation of the object. Because it is nondestructive, it can be repeated and used as a control after position-correcting maneuvers. Conclusion: Cross-sectional imaging based on radiography is a valuable tool for the analysis of the electrode-modiolus relationship after cochlear implantation in isolated temporal bones. which may confirm histologic analysis.
引用
收藏
页码:49 / 52
页数:4
相关论文
共 17 条
[1]  
Anxionnat R, 1998, J NEURORADIOLOGY, V25, P251
[2]   Second-generation three-dimensional reconstruction for rotational three-dimensional angiography [J].
Bidaut, LM ;
Laurent, C ;
Piotin, M ;
Gailloud, P ;
Muster, M ;
Fasel, JHD ;
Rüfenacht, DA ;
Terrier, F .
ACADEMIC RADIOLOGY, 1998, 5 (12) :836-849
[3]   FACTORS PREDICTING POSTOPERATIVE SENTENCE SCORES IN POSTLINGUISTICALLY DEAF ADULT COCHLEAR IMPLANT PATIENTS [J].
BLAMEY, PJ ;
PYMAN, BC ;
CLARK, GM ;
DOWELL, RC ;
GORDON, M ;
BROWN, AM ;
HOLLOW, RD .
ANNALS OF OTOLOGY RHINOLOGY AND LARYNGOLOGY, 1992, 101 (04) :342-348
[4]  
Donnelly M J, 1995, Ann Otol Rhinol Laryngol Suppl, V166, P409
[5]   The Clarion electrode positioner: Temporal bone studies [J].
Fayad, JN ;
Luxford, W ;
Linthicum, FH .
AMERICAN JOURNAL OF OTOLOGY, 2000, 21 (02) :226-229
[6]   PRACTICAL CONE-BEAM ALGORITHM [J].
FELDKAMP, LA ;
DAVIS, LC ;
KRESS, JW .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1984, 1 (06) :612-619
[7]  
HUSSTEDT HW, 2001, IN PRESS J NEUROIMAG
[8]  
Ketten DR, 1998, ANN OTO RHINOL LARYN, V107, P1
[9]   A new volumetric CT machine for dental imaging based on the cone-beam technique: preliminary results [J].
Mozzo, P ;
Procacci, C ;
Tacconi, A ;
Martini, PT ;
Andreis, IAB .
EUROPEAN RADIOLOGY, 1998, 8 (09) :1558-1564
[10]   Cochlear structures after implantation of a perimodiolar electrode array [J].
Richter, B ;
Jaekel, K ;
Aschendorff, A ;
Marangos, N ;
Laszig, R .
LARYNGOSCOPE, 2001, 111 (05) :837-843