Quantitative optical densitometry with scanning-laser film digitizers

被引:47
作者
Dempsey, JF [1 ]
Low, DA [1 ]
Kirov, AS [1 ]
Williamson, JF [1 ]
机构
[1] Washington Univ, Sch Med, Edward Mallinckrodt Inst Radiol, Radiat Oncol Ctr, St Louis, MO 63110 USA
关键词
laser film digitizer; film digitization; film densitometry; film dosimetry; electrophoresis gel quantitation;
D O I
10.1118/1.598664
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A new process for eliminating two types of artifacts inherent in commercially available transmission scanning-laser film digitizers is presented. The first kind of artifact results in nonreproducible interference-pattern fluctuations as large as 7%. The second kind results in spreading of transmitted light from low-to-high optical density (OD) in regions with rapidly varying ODs, producing errors as large as 50%. These OD artifacts cause the loss of precision for films with low-OD regions (first type) and the loss of accuracy for films with regions of high-OD near high-OD gradients (second type).:Test radiochromic films, produced by uniform exposure:to a 6 MV photon beam and a high dose rate Ir-192 brachytherapy source, along with test radiographic films were used to characterize the artifacts of a commercially available scanning-laser film digitizer. The interference-pattern artifact was eliminated by digitizing the films on a masked diffusing ground-glass scanning bed. The light-transmission artifact was eliminated through discrete-fast-Fourier-transform (DFFT) de convolution of transmission profiles with measured digitizer line-spread functions. Obtaining precise: OD distributions after the DFFT deconvolution required prior removal of the interference-pattern artifact and application of a low-pass Wiener noise filter. Light-transmission artifacts are particularly significant for applications requiring measurement of high-gradient OD distributions, such as brachytherapy or conformal photon-beam film dosimetry and quantitation of two-dimensional electrophoresis gels. Errors as large as 15%-35% occur in OD distributions representative of these applications. The data collection and correction:process developed in this study successfully removes these artifacts. (C) 1999 American Association of Physicists in Medicine. [S0094-2405(99)00308-9].
引用
收藏
页码:1721 / 1731
页数:11
相关论文
共 22 条
[1]   Dosimetry of a radioactive coronary balloon dilitation catheter for treatment of neointimal hyperplasia [J].
Amols, HI ;
Reinstein, LE ;
Weinberger, J .
MEDICAL PHYSICS, 1996, 23 (10) :1783-1788
[2]   HIGH-SENSITIVITY GAFCHROMIC FILM DOSIMETRY FOR I-125 SEED [J].
CHIUTSAO, ST ;
DELAZERDA, A ;
LIN, J ;
KIM, JH .
MEDICAL PHYSICS, 1994, 21 (05) :651-657
[3]  
Dutreix J, 1969, Ann N Y Acad Sci, V161, P33
[4]  
GARRELS JI, 1989, J BIOL CHEM, V264, P5269
[5]   ACCEPTANCE TESTING AND QUALITY ASSURANCE OF AUTOMATED SCANNING FILM DENSITOMETERS USED IN THE DOSIMETRY OF ELECTRON AND PHOTON THERAPY BEAMS [J].
HOLMES, TW ;
MCCULLOUGH, EC .
MEDICAL PHYSICS, 1983, 10 (05) :698-700
[6]   DENSITOMETRIC QUANTITATION OF HIGH-RESOLUTION AGAROSE-GEL PROTEIN ELECTROPHORESIS [J].
HOWERTON, DA ;
CHECK, IJ ;
HUNTER, RL .
AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1986, 85 (02) :213-218
[7]  
Lim J. S., 1990, Two-dimensional signal and image processing
[8]   SENSITOMETRY OF THE RESPONSE OF A NEW RADIOCHROMIC FILM DOSIMETER TO GAMMA-RADIATION AND ELECTRON-BEAMS [J].
MCLAUGHLIN, WL ;
YUNDONG, C ;
SOARES, CG ;
MILLER, A ;
VANDYK, G ;
LEWIS, DF .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1991, 302 (01) :165-176
[9]   TESTS FOR EVALUATING LASER FILM DIGITIZERS [J].
MEEDER, RJJ ;
JAFFRAY, DA ;
MUNRO, P .
MEDICAL PHYSICS, 1995, 22 (05) :635-642
[10]   Dosimetric characteristics of an improved radiochromic film [J].
Meigooni, AS ;
Sanders, MF ;
Ibbott, GS ;
Szeglin, SR .
MEDICAL PHYSICS, 1996, 23 (11) :1883-1888