Quantitative optical densitometry with scanning-laser film digitizers

被引:47
作者
Dempsey, JF [1 ]
Low, DA [1 ]
Kirov, AS [1 ]
Williamson, JF [1 ]
机构
[1] Washington Univ, Sch Med, Edward Mallinckrodt Inst Radiol, Radiat Oncol Ctr, St Louis, MO 63110 USA
关键词
laser film digitizer; film digitization; film densitometry; film dosimetry; electrophoresis gel quantitation;
D O I
10.1118/1.598664
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A new process for eliminating two types of artifacts inherent in commercially available transmission scanning-laser film digitizers is presented. The first kind of artifact results in nonreproducible interference-pattern fluctuations as large as 7%. The second kind results in spreading of transmitted light from low-to-high optical density (OD) in regions with rapidly varying ODs, producing errors as large as 50%. These OD artifacts cause the loss of precision for films with low-OD regions (first type) and the loss of accuracy for films with regions of high-OD near high-OD gradients (second type).:Test radiochromic films, produced by uniform exposure:to a 6 MV photon beam and a high dose rate Ir-192 brachytherapy source, along with test radiographic films were used to characterize the artifacts of a commercially available scanning-laser film digitizer. The interference-pattern artifact was eliminated by digitizing the films on a masked diffusing ground-glass scanning bed. The light-transmission artifact was eliminated through discrete-fast-Fourier-transform (DFFT) de convolution of transmission profiles with measured digitizer line-spread functions. Obtaining precise: OD distributions after the DFFT deconvolution required prior removal of the interference-pattern artifact and application of a low-pass Wiener noise filter. Light-transmission artifacts are particularly significant for applications requiring measurement of high-gradient OD distributions, such as brachytherapy or conformal photon-beam film dosimetry and quantitation of two-dimensional electrophoresis gels. Errors as large as 15%-35% occur in OD distributions representative of these applications. The data collection and correction:process developed in this study successfully removes these artifacts. (C) 1999 American Association of Physicists in Medicine. [S0094-2405(99)00308-9].
引用
收藏
页码:1721 / 1731
页数:11
相关论文
共 22 条
[11]   REGULATION OF PHOSPHATIDYLCHOLINE BIOSYNTHESIS IN ACTIVATED ALVEOLAR TYPE-II CELLS [J].
MILLER, BE ;
HOOK, GER .
AMERICAN JOURNAL OF RESPIRATORY CELL AND MOLECULAR BIOLOGY, 1989, 1 (02) :127-136
[12]   PHOTON ENERGY-DEPENDENCE OF THE SENSITIVITY OF RADIOCHROMIC FILM AND COMPARISON WITH SILVER-HALIDE FILM AND LIF TLDS USED FOR BRACHYTHERAPY DOSIMETRY [J].
MUENCH, PJ ;
MEIGOONI, AS ;
NATH, R ;
MCLAUGHLIN, WL .
MEDICAL PHYSICS, 1991, 18 (04) :769-775
[13]   Radiochromic film dosimetry: Recommendations of AAPM Radiation Therapy Committee Task Group 55 [J].
Niroomand-Rad, A ;
Blackwell, CR ;
Coursey, BM ;
Gall, KP ;
Galvin, JM ;
McLaughlin, WL ;
Meigooni, AS ;
Nath, R ;
Rodgers, JE ;
Soares, CG .
MEDICAL PHYSICS, 1998, 25 (11) :2093-2115
[14]  
OFARRELL PH, 1975, J BIOL CHEM, V250, P4007
[15]   THE USE OF RADIOCHROMIC FILM IN TREATMENT VERIFICATION OF DYNAMIC STEREOTAXIC RADIOSURGERY [J].
RAMANI, R ;
LIGHTSTONE, AW ;
MASON, DLD ;
OBRIEN, PF .
MEDICAL PHYSICS, 1994, 21 (03) :389-392
[16]   Optical density dependence on postirradiation temperature and time for MD-55-2 type radiochromic film [J].
Reinstein, LE ;
Gluckman, GR .
MEDICAL PHYSICS, 1999, 26 (03) :478-484
[17]  
REINSTEIN LE, 1995, MED PHYS, V22, P1011
[18]   CALIBRATION OF OPHTHALMIC APPLICATORS AT NIST - A REVISED APPROACH [J].
SOARES, CG .
MEDICAL PHYSICS, 1991, 18 (04) :787-793
[19]  
TRUEBLOOD JH, 1993, DIGITAL IMAGING
[20]   FILM DOSIMETRY OF MEGAVOLTAGE PHOTON BEAMS - A PRACTICAL METHOD OF ISODENSITY-TO-ISODOSE CURVE CONVERSION [J].
WILLIAMSON, JF ;
KHAN, FM ;
SHARMA, SC .
MEDICAL PHYSICS, 1981, 8 (01) :94-98