CALIBRATION OF OPHTHALMIC APPLICATORS AT NIST - A REVISED APPROACH

被引:62
作者
SOARES, CG
机构
[1] National Institute of Standards and Technology, Gaithersburg, Maryland
关键词
D O I
10.1118/1.596632
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A revised approach to the problem of measuring a surface-absorbed dose from beta-particle-emitting ophthalmic applicators is presented. The technique chosen employs an extrapolation chamber equipped with a 4-mm-diam collecting electrode to make current measurements at air gaps from 0.08 to 0.20 mm at 0.02-mm intervals. These data yield a linear relationship between current and air gap, the slope of which is used to determine average surface-absorbed-dose rate over the central area of the source. For additional information about the distribution of the activity over the source surface, autoradiographs using calibrated radiochromic dye films are analyzed to map the dose-rate profile across the surface of the applicator. Experiments varying several parameters of the extrapolation chamber measurement, including collecting electrode area, voltage gradient, range of air gaps used, and entrance foil material, are described. Also treated are calibrations of, and a description of the use of, radiochromic dye films for source profiling. Experiments for determining correction factors for the extrapolation chamber measurements are described, and an assessment of the uncertainties associated with these measurements is given.
引用
收藏
页码:787 / 793
页数:7
相关论文
共 18 条
[1]  
AINSWORTH A, 1989, UNPUB MAR NIST WORKS
[2]   DETERMINATION OF SURFACE DOSE-RATE FROM A SR-90 OPHTHALMIC APPLICATOR [J].
ALI, MM ;
KHAN, FM .
MEDICAL PHYSICS, 1990, 17 (03) :416-421
[3]  
BERGER MJ, 1983, NBSIR822550A REP
[4]   SATURATION CORRECTIONS FOR PLANE-PARALLEL IONIZATION CHAMBERS [J].
BOHM, J .
PHYSICS IN MEDICINE AND BIOLOGY, 1976, 21 (05) :754-759
[5]  
DEASY JO, UNPUB EXTRAPOLATION
[6]   CALIBRATION OF SR-90 OPHTHALMIC APPLICATORS [J].
GOETSCH, SJ .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1989, 16 (06) :1653-1653
[7]   SURFACE DOSE-RATE CALIBRATION OF SR-90 PLANE OPHTHALMIC APPLICATORS [J].
GOETSCH, SJ ;
SUNDERLAND, KS .
MEDICAL PHYSICS, 1991, 18 (02) :161-166
[8]   EXTRAPOLATION CHAMBER FOR THE MEASUREMENT OF BETA-SOURCES [J].
LOEVINGER, R .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1953, 24 (10) :907-914
[9]  
LOEVINGER R, 1913, INT J APPL RADIAT IS, V10, P113
[10]  
MCLAUGHLIN WL, IN PRESS NUCL INSTRU