Numerical investigation of the intravascular coronary stent flexibility

被引:103
作者
Petrini, L
Migliavacca, F
Auricchio, F
Dubini, G
机构
[1] Univ Pavia, Dipartimento Meccan Strutturale, I-27100 Pavia, Italy
[2] Politecn Milan, Lab Biol Struct Mech, Dipartimento Bioingn & Ingn Strutturale, I-20133 Milan, Italy
关键词
stent flexibility; coronary stent; finite element method; mechanical response; mathematical model;
D O I
10.1016/j.jbiomech.2003.09.002
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Nowadays stent therapy is widely adopted to treat atherosclerotic vessel diseases. The high commercial value of these devices and the high prototypation costs require the use of finite element analyses, instead of classical trial and error technique, to design and verify new models. In this paper, we explore the advantages of the finite element method (FEM) in order to investigate new generation stent performance in terms of flexibility. Indeed, the ability of the stent to bend in order to accommodate curvatures and angles of vessels during delivery is one of the most significant prerequisites for optimal stent performance. Two different FEM models, resembling two new generation intravascular stents, were developed. The main model dimensions were obtained by means of a stereo microscope, analyzing one Cordis BX-Velocity and one Carbostent Sirius coronary stent. Bending tests under displacement control in the unexpanded and expanded configuration were carried out. The curvature index, defined as the ratio between the sum of rotation angles at the extremes and the length of the stent, yielded comparative information about the capability of the device to be delivered into tortuous vessels and to conform to their contours. Results, expressed in terms of the bending moment-curvature index, demonstrated a different response for the two models. In particular the Cordis model showed a higher flexibility. Lower flexibility in the expanded configurations for both models was detected. However this flexibility depends on how the contact takes place between the different parts of the struts. (C) 2003 Published by Elsevier Ltd.
引用
收藏
页码:495 / 501
页数:7
相关论文
共 18 条
[1]  
Auricchio F., 2001, COMPUT METHOD BIOMEC, V4, P249, DOI DOI 10.1080/10255840108908007
[2]   Coronary restenosis: A review of current insights and therapies [J].
Bailey, SR .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2002, 55 (02) :265-271
[3]  
BARTORELLI AL, 2000, HDB CORONARY STENTS, P177
[4]   Physical properties of endovascular stents: An experimental comparison [J].
Duda, SH ;
Wiskirchen, J ;
Tepe, G ;
Bitzer, M ;
Kaulich, TW ;
Stoeckel, D ;
Claussen, CD .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2000, 11 (05) :645-654
[5]   Mechanical behaviour modelling of balloon-expandable stents [J].
Dumoulin, C ;
Cochelin, B .
JOURNAL OF BIOMECHANICS, 2000, 33 (11) :1461-1470
[6]   Mechanical properties of metallic stents: How do these properties influence the choice of stent for specific lesions? [J].
Dyet, JF ;
Watts, WG ;
Ettles, DF ;
Nicholson, AA .
CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2000, 23 (01) :47-54
[7]   Mechanical properties of coronary stents determined by using finite element analysis [J].
Etave, F ;
Finet, G ;
Boivin, M ;
Boyer, JC ;
Rioufol, G ;
Thollet, G .
JOURNAL OF BIOMECHANICS, 2001, 34 (08) :1065-1075
[8]   STRENGTH, ELASTICITY, AND PLASTICITY OF EXPANDABLE METAL STENTS - IN-VITRO STUDIES WITH 3 TYPES OF STRESS [J].
FLUECKIGER, F ;
STERNTHAL, H ;
KLEIN, GE ;
ASCHAUER, M ;
SZOLAR, D ;
KLEINHAPPL, G .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 1994, 5 (05) :745-750
[9]   Computational fluid dynamics and stent design [J].
Frank, AO ;
Walsh, PW ;
Moore, JE .
ARTIFICIAL ORGANS, 2002, 26 (07) :614-621
[10]   Treating atherosclerosis: local drug delivery from laboratory studies to clinical trials [J].
Gershlick, AH .
ATHEROSCLEROSIS, 2002, 160 (02) :259-271