Quantification of mitral regurgitation with MR phase-velocity mapping using a control volume method

被引:27
作者
Chatzimavroudis, GP [1 ]
Oshinski, JN [1 ]
Pettigrew, RI [1 ]
Walker, PG [1 ]
Franch, RH [1 ]
Yoganathan, AP [1 ]
机构
[1] Georgia Inst Technol, Sch Chem Engn, Cardiovasc Fluid Mech Lab, Inst Bioengineering & Biosci, Atlanta, GA 30332 USA
来源
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING | 1998年 / 8卷 / 03期
关键词
mitral regurgitation; control volume; regurgitant volume; aortic outflow;
D O I
10.1002/jmri.1880080310
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Reliable diagnosis and quantification of mitral regurgitation are important for patient management and for optimizing the time for surgery. Previous methods have often provided suboptimal results. The aim of this in vitro study was to evaluate MR phase-velocity mapping in quantifying the mitral regurgitant volume (MRV) using a control volume (CV) method, A number of contiguous slices were acquired with all three velocity components measured. A CV was then selected, encompassing the regurgitant orifice. Mass conservation dictates that the net inflow into the CV should be equal to the regurgitant now. Results showed that a CV, the boundary voxels of which excluded the region of now acceleration and aliasing at the orifice, provided accurate measurements of the regurgitant now. A smaller CV provided erroneous results because of now acceleration and velocity aliasing close to the orifice. A large Cv generally provided inaccurate results because of reduced velocity sensitivity far from the orifice. Aortic outflow, orifice shape, and valve geometry did not affect the accuracy of the CV measurements. The CV method is a promising approach to the problem of quantification of the MRV.
引用
收藏
页码:577 / 582
页数:6
相关论文
共 25 条
[1]   Slice location dependence of aortic regurgitation measurements with MR phase velocity mapping [J].
Chatzimavroudis, GP ;
Walker, PG ;
Oshinski, JN ;
Franch, RH ;
Pettigrew, RI ;
Yoganathan, AP .
MAGNETIC RESONANCE IN MEDICINE, 1997, 37 (04) :545-551
[2]   PULSED DOPPLER ECHOCARDIOGRAPHY IN THE DIAGNOSIS AND ESTIMATION OF SEVERITY OF AORTIC-INSUFFICIENCY [J].
CIOBANU, M ;
ABBASI, AS ;
ALLEN, M ;
HERMER, A ;
SPELLBERG, R .
AMERICAN JOURNAL OF CARDIOLOGY, 1982, 49 (02) :339-343
[3]   LIMITATIONS OF QUALITATIVE ANGIOGRAPHIC GRADING IN AORTIC OR MITRAL REGURGITATION [J].
CROFT, CH ;
LIPSCOMB, K ;
MATHIS, K ;
FIRTH, BG ;
NICOD, P ;
TILTON, G ;
WINNIFORD, MD ;
HILLIS, LD .
AMERICAN JOURNAL OF CARDIOLOGY, 1984, 53 (11) :1593-1598
[4]  
Dulce MC, 1992, RADIOLOGY, V185, P234
[5]   INVIVO VALIDATION OF MR VELOCITY IMAGING [J].
FIRMIN, DN ;
NAYLER, GL ;
KLIPSTEIN, RH ;
UNDERWOOD, SR ;
REES, RSO ;
LONGMORE, DB .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1987, 11 (05) :751-756
[6]   QUANTIFICATION OF MITRAL REGURGITATION BY VELOCITY-ENCODED CINE NUCLEAR-MAGNETIC-RESONANCE IMAGING [J].
FUJITA, N ;
CHAZOUILLERES, AF ;
HARTIALA, JJ ;
OSULLIVAN, M ;
HEIDENREICH, P ;
KAPLAN, JD ;
SAKUMA, H ;
FOSTER, E ;
CAPUTO, GR ;
HIGGINS, CB .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1994, 23 (04) :951-958
[7]   REAL-TIME FLOOD FLOW IMAGING BY SPIRAL SCAN PHASE-VELOCITY MAPPING [J].
GATEHOUSE, PD ;
FIRMIN, DN ;
COLLINS, S ;
LONGMORE, DB .
MAGNETIC RESONANCE IN MEDICINE, 1994, 31 (05) :504-512
[8]   QUANTITATIVE ASSESSMENT OF AORTIC REGURGITATION BY MAGNETIC-RESONANCE-IMAGING [J].
GLOBITS, S ;
FRANK, H ;
MAYR, H ;
NEUHOLD, A ;
GLOGAR, D .
EUROPEAN HEART JOURNAL, 1992, 13 (01) :78-83
[9]   QUANTITATIVE ASSESSMENT OF AORTIC REGURGITATION BY COMBINED 2-DIMENSIONAL, CONTINUOUS-WAVE AND COLOR FLOW DOPPLER MEASUREMENTS [J].
HOLM, S ;
ERIKSSON, P ;
KARP, K ;
OSTERMAN, G ;
TEIEN, D .
JOURNAL OF INTERNAL MEDICINE, 1992, 231 (02) :115-121
[10]  
Hopmeyer J, 1996, J Am Soc Echocardiogr, V9, P44, DOI 10.1016/S0894-7317(96)90103-0