Comparison of Early Strut Coverage Between Zotarolimus- and Everolimus-Eluting Stents Using Optical Coherence. Tomography

被引:37
作者
Kim, Seunghwan [1 ]
Kim, Jung-Sun [1 ]
Shin, Dong-Ho [1 ]
Kim, Byeong-Keuk [1 ]
Ko, Young-Guk [1 ]
Choi, Donghoon [1 ]
Cho, Yun-Kyeong [2 ]
Nam, Chang-Wook [2 ]
Hur, Seung-Ho [2 ]
Jang, Yangsoo [1 ,3 ]
Hong, Myeong-Ki [1 ,3 ]
机构
[1] Yonsei Univ, Coll Med, Severance Cardiovasc Hosp, Div Cardiol, Seoul 120749, South Korea
[2] Keimyung Univ, Dongsan Med Ctr, Taegu, South Korea
[3] Yonsei Univ, Coll Med, Severance Biomed Sci Inst, Seoul 120749, South Korea
关键词
NEOINTIMAL COVERAGE; CORONARY STENTS; IMPLANTATION; THROMBOSIS; RESOLUTE; TRIAL;
D O I
10.1016/j.amjcard.2012.08.037
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
There have been no optical coherence tomographic (OCT) data directly comparing the pattern of strut coverage between the 2 second-generation drug-eluting stents in the early period. The aim of this prospective study was to evaluate early strut coverage using optical coherence tomography 3 months after Resolute zotarolimus-eluting stent (ZES-R) or everolimus-eluting stent (EES) implantation in de novo coronary artery lesions. A total of 40 patients who were suitable for the OCT procedure and consented to the study protocol were randomized 1:1 to receive either ZES-R or EES. Among these patients, 35 stented lesions (18 ZES-R, 17 EES) in 34 patients were evaluated by optical coherence tomography immediately and 3 months after stent implantation. Neointimal hyperplasia thickness, percentage of uncovered struts, and the proportion of malapposed struts were measured, at 1-mm intervals. An uncovered strut was defined as having a neointimal hyperplasia thickness of 0 mu m. At the 3-month OCT evaluation, mean neointimal hyperplasia thickness (ZES-R vs EES 74 +/- 41 vs 75 35 gm, p = 0.89) and mean percentage of uncovered struts (ZES-R vs EES 6.2 +/- 6.9 vs 4.7 +/- 5.1%, p = 0.62) were not significantly different between the groups. The percentage of malapposed struts was also similar between the groups (0.7 +/- 2.2% for ZES-R and 0.7 +/- 1.7% for EES, p = 0.64). Thrombi Were documented in 3 stents (1 [5.6%] in a ZES-R vs 2 [11.8%] in EES, p = 0.60). In conclusion, early stent strut coverage on the basis of serial OCT evaluation was comparable between ZES-R and EES 3 months after stent implantation. (c) 2013 Elsevier Inc. All rights reserved. (Am J Cardiol 2013;111:1-5)
引用
收藏
页码:1 / 5
页数:5
相关论文
共 19 条
[1]   Incomplete stent apposition and very late stent thrombosis after drug-eluting stent implantation [J].
Cook, Stephane ;
Wenaweser, Peter ;
Togni, Mario ;
Billinger, Michael ;
Morger, Cyrill ;
Seiler, Christian ;
Vogel, Rolf ;
Hess, Otto ;
Meier, Bernhard ;
Windecker, Stephan .
CIRCULATION, 2007, 115 (18) :2426-2434
[2]   Pathological mechanisms of fatal late coronary stent thrombosis in humans [J].
Farb, A ;
Burke, AP ;
Kolodgie, FD ;
Virmani, R .
CIRCULATION, 2003, 108 (14) :1701-1706
[3]   Pathological correlates of late drug-eluting stent thrombosis - Strut coverage as a marker of endothelialization [J].
Finn, Aloke V. ;
Joner, Michael ;
Nakazawa, Gaku ;
Kolodgie, Frank ;
Newell, John ;
John, Mike C. ;
Gold, Herman K. ;
Virmani, Renu .
CIRCULATION, 2007, 115 (18) :2435-2441
[4]   Impact of Drug Release Kinetics on Vascular Response to Different Zotarolimus-Eluting Stents Implanted in Patients With Long Coronary Stenoses The LongOCT Study (Optical Coherence Tomography in Long Lesions) [J].
Guagliumi, Giulio ;
Ikejima, Hideyuki ;
Sirbu, Vasile ;
Bezerra, Hiram ;
Musumeci, Giuseppe ;
Lortkipanidze, Nikoloz ;
Fiocca, Luigi ;
Tahara, Satoko ;
Vassileva, Angelina ;
Matiashvili, Aleksandre ;
Valsecchi, Orazio ;
Costa, Marco .
JACC-CARDIOVASCULAR INTERVENTIONS, 2011, 4 (07) :778-785
[5]   Tissue coverage of a hydrophilic polymer-coated zotarolimus-eluting stent vs. a fluoropolymer-coated everolimus-eluting stent at 13-month follow-up: an optical coherence tomography substudy from the RESOLUTE All Comers trial [J].
Gutierrez-Chico, Juan Luis ;
van Geuns, Robert Jan ;
Regar, Evelyn ;
van der Giessen, Willem J. ;
Kelbaek, Henning ;
Saunamaki, Kari ;
Escaned, Javier ;
Gonzalo, Nieves ;
di Mario, Carlo ;
Borgia, Francesco ;
Nueesch, Eveline ;
Garcia-Garcia, Hector M. ;
Silber, Sigmund ;
Windecker, Stephan ;
Serruys, Patrick W. .
EUROPEAN HEART JOURNAL, 2011, 32 (19) :2454-2463
[6]   Optical coherence evaluation of everolimus-eluting stents 8 months after implantation [J].
Inoue, Takumi ;
Shite, Junya ;
Yoon, Junghan ;
Shinke, Toshiro ;
Otake, Hiromasa ;
Sawada, Takahiro ;
Kawamori, Hiroyuki ;
Katoh, Hiroki ;
Miyoshi, Naoki ;
Yoshino, Naoki ;
Kozuki, Amane ;
Hariki, Hirotoshi ;
Hirata, Ken-ichi .
HEART, 2011, 97 (17) :1379-1384
[7]   Endothelial cell recovery between comparator polymer-based drug-eluting stents [J].
Joner, Michael ;
Nakazawa, Gaku ;
Finn, Aloke V. ;
Quee, Shawn Chin ;
Coleman, Leslie ;
Acampado, Eduardo ;
Wilson, Patricia S. ;
Skorija, Kristi ;
Cheng, Qi ;
Xu, Xin ;
Gold, Herman K. ;
Kolodgie, Frank D. ;
Virmani, Renu .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2008, 52 (05) :333-342
[8]   Comparison of Optical Coherence Tomographic Assessment between First- and Second-Generation Drug-Eluting Stents [J].
Kim, Byeong-Keuk ;
Kim, Jung-Sun ;
Park, Junbeom ;
Ko, Young-Guk ;
Choi, Donghoon ;
Jang, Yangsoo ;
Hong, Myeong-Ki .
YONSEI MEDICAL JOURNAL, 2012, 53 (03) :524-529
[9]   ComparisOn of neointimal coVerage betwEen zotaRolimus-eluting stent and everolimus-eluting stent using Optical Coherence Tomography (COVER OCT) [J].
Kim, Jung-Sun ;
Kim, Byeong-Kuk ;
Jang, Ik-Kyung ;
Shin, Dong-Ho ;
Ko, Young-Guk ;
Choi, Donghoon ;
Hong, Myeong-Ki ;
Cho, Yun-Kyeong ;
Nam, Chang-Wook ;
Hur, Seung-Ho ;
Choi, Jin-Ho ;
Song, Young Bin ;
Hahn, Joo Yong ;
Choi, Seung Hyuk ;
Gwon, Hyeon Cheol ;
Jang, Yangsoo .
AMERICAN HEART JOURNAL, 2012, 163 (04) :601-607
[10]   Assessment of coronary arterial thrombus by optical coherence tomography [J].
Kume, Teruyoshi ;
Akasaka, Takashi ;
Kawamoto, Takahiro ;
Ogasawara, Yasuo ;
Watanabe, Nozorni ;
Toyota, Eiji ;
Neishi, Yoji ;
Sukmawan, Renan ;
Sadahira, Yoshito ;
Yoshida, Kiyoshi .
AMERICAN JOURNAL OF CARDIOLOGY, 2006, 97 (12) :1713-1717