Diagnosis of Hepatic Metastasis: Comparison of Respiration-Triggered Diffusion-Weighted Echo-Planar MRI and Five T2-Weighted Turbo Spin-Echo Sequences

被引:120
作者
Bruegel, Melanie [1 ]
Gaa, Jochen [1 ]
Waldt, Simone [1 ]
Woertler, Klaus [1 ]
Holzapfel, Konstantin [1 ]
Kiefer, Berthold [2 ]
Rummeny, Ernst J. [1 ]
机构
[1] Tech Univ Munich, Dept Radiol, Klinikum Rechts Isar, D-81675 Munich, Germany
[2] Siemens Med Solut, Erlangen, Germany
关键词
detection; diffusion MRI; echo-planar imaging; hepatic metastasis;
D O I
10.2214/AJR.07.3279
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 [临床医学]; 100207 [影像医学与核医学]; 1009 [特种医学];
摘要
OBJECTIVE. The purpose of this study was to compare the value of respiration-triggered diffusion-weighted (DW) single-shot echo-planar MRI (EPI) and five variants of T2-weighted turbo spin-echo (TSE) sequences in the diagnosis of hepatic metastasis. MATERIALS AND METHODS. Fifty-two patients with extrahepatic primary malignant tumors underwent 1.5-T MRI that included DW EPI and the following variants of T2weighted TSE techniques: breath-hold fat-suppressed HASTE, breath-hold fat-supressed TSE, respiration-triggered fat-suppressed TSE, breath-hold STIR, and respiration-triggered STIR. Images were reviewed independently by two blinded observers who used a 5-point confidence scale to identify lesions. Results were correlated with surgical and histopathologic findings and follow-up imaging findings. The accuracy of each technique was measured with free-response receiver operating characteristic analysis. RESULTS. A total of 118 hepatic metastatic lesions (mean diameter, 12.8 mm; range, 3-84 mm) were evaluated. Accuracy values were higher (p < 0.001) with DW EPI (0.91-0.92) than with the T2-weighted TSE techniques (0.47-0.67). Imaging with the HASTE sequence (0.47-0.52) was less accurate (p < 0.05) than imaging with the breath-hold TSE, breath-hold STIR, respiration-triggered TSE, and respiration-triggered STIR sequences (0.59-0.67). Sensitivity was higher (p < 0.001) with DW EPI (0.88-0.91) than with T2-weighted TSE techniques (0.45-0.62). For small (<= 10 mm) metastatic lesions only, the differences in sensitivity between DW EPI (0.85) and T2-weighted TSE techniques (0.26-0.44) were even more pronounced. CONCLUSION. DW EPI was more sensitive and more accurate than imaging with T2weighted TSE techniques. Because of the black-blood effect on vessels and low susceptibility to motion artifacts, DW EPI was particularly useful for the detection of small (<= 10 mm) metastatic lesions.
引用
收藏
页码:1421 / 1429
页数:9
相关论文
共 32 条
[1]
Liver: T2-weighted MR imaging with breath-hold fast-recovery optimized fast spin-echo compared with breath-hold half-Fourier and non-breath-hold respiratory-triggered fast spin-echo pulse sequences [J].
Augui, J ;
Vignaux, O ;
Argaud, C ;
Coste, J ;
Gouya, H ;
Legmann, P .
RADIOLOGY, 2002, 223 (03) :853-859
[2]
Detection of colorectal liver metastases: a prospective multicenter trial comparing unenhanced MRI, MnDPDP-enhanced MRI, and spiral CT [J].
Bartolozzi, C ;
Donati, F ;
Cioni, D ;
Procacci, C ;
Morana, G ;
Chiesa, A ;
Grazioli, L ;
Cittadini, G ;
Cittadini, G ;
Giovagnoni, A ;
Gandini, G ;
Maass, J ;
Lencioni, R .
EUROPEAN RADIOLOGY, 2004, 14 (01) :14-20
[3]
Hepatic lesion detection at MR imaging: A comparative study with four sequences [J].
Blakeborough, A ;
Ward, J ;
Wilson, D ;
Griffiths, M ;
Kajiya, Y ;
Guthrie, JA ;
Robinson, PJA .
RADIOLOGY, 1997, 203 (03) :759-765
[4]
Characterization of focal liver lesions by ADC measurements using a respiratory triggered diffusion-weighted single-shot echo-planar MR imaging technique [J].
Bruegel, Melanie ;
Holzapfel, Konstantin ;
Gaa, Jochen ;
Woertler, Klaus ;
Waldt, Simone ;
Kiefer, Berthold ;
Stemmer, Alto ;
Ganter, Carl ;
Rummeny, Ernst J. .
EUROPEAN RADIOLOGY, 2008, 18 (03) :477-485
[5]
Single breath-hold diffusion-weighted imaging of the abdomen [J].
Chow, LC ;
Bammer, R ;
Moseley, ME ;
Sommer, FG .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2003, 18 (03) :377-382
[6]
Improved focal liver lesion detection:: comparison of single-shot diffusion-weighted echoplanar and single-shot T2 weighted turbo spin echo techniques [J].
Coenegrachts, K. ;
Delanote, J. ;
Ter Beek, L. ;
Haspeslagh, M. ;
Bipat, S. ;
Stoker, J. ;
Van Kerkhove, F. ;
Steyaert, L. ;
Rigauts, H. ;
Casselman, J. W. .
BRITISH JOURNAL OF RADIOLOGY, 2007, 80 (955) :524-531
[7]
Spectrum of MRI appearances of untreated metastases of the liver [J].
Danet, IM ;
Semelka, RC ;
Leonardou, P ;
Braga, L ;
Vaidean, G ;
Woosley, JT ;
Kanematsu, M .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2003, 181 (03) :809-817
[8]
Liver masses: Replacement of conventional T2-weighted spin-echo MR imaging with breath-hold MR imaging [J].
Gaa, J ;
Hatabu, H ;
Jenkins, RL ;
Finn, JP ;
Edelman, RR .
RADIOLOGY, 1996, 200 (02) :459-464
[9]
Generalized Autocalibrating Partially Parallel Acquisitions (GRAPPA) [J].
Griswold, MA ;
Jakob, PM ;
Heidemann, RM ;
Nittka, M ;
Jellus, V ;
Wang, JM ;
Kiefer, B ;
Haase, A .
MAGNETIC RESONANCE IN MEDICINE, 2002, 47 (06) :1202-1210
[10]
CT and MR imaging of benign hepatic and biliary tumors [J].
Horton, KM ;
Bluemke, DA ;
Hruban, RH ;
Soyer, P ;
Fishman, EK .
RADIOGRAPHICS, 1999, 19 (02) :431-451