In vitro simulation - Early results of stereotaxy for pedicle screw placement

被引:58
作者
Carl, AL
Khanuja, HS
Sachs, BL
Gatto, CA
vomLehn, J
Vosburgh, K
Schenck, J
Lorensen, W
Rohling, K
Disler, D
机构
[1] GE CO,CTR CORP RES & DEV,SCHENECTADY,NY 12301
[2] ALBANY MED CTR,DEPT RADIOL,ALBANY,NY
关键词
Doppler; in vitro simulation; pedicle screw; stereotaxy; ultrasound; FIXATION; FRAMELESS; COMPLICATIONS; NEUROSURGERY; INSERTION; ACCURACY; SPINE;
D O I
10.1097/00007632-199705150-00018
中图分类号
R74 [神经病学与精神病学];
学科分类号
100204 [神经病学];
摘要
Study Design. Frameless stereotaxy with doppler ultrasound and three dimensional computer model registration is assessed in vitro for pedicle screw placement. Objective. To identify feasibility of pedicle screw navigation and placement using this technology. Summary of Background Data. Inaccurate pedicle screw placement can lead to neurovascular injury or suboptimal fixation. Present techniques in pedicle screw placement involve only confirmation of hole orientation. Method. Forty-four pedicle screws were placed in lumbosacral models and cadaver specimens. Accuracy was assessed with a computed tomography scan and vertebral cross sectioning. Results. All screws were intrapedicular. Accuracy of anterior cortical fixation was 1.5 mm, with a range of 2.5 mm. Conclusion. In vitro frameless stereotaxy is accurate for pedicle screw placement. This technology adds a component of navigation to pedicle screw placement.
引用
收藏
页码:1160 / 1164
页数:5
相关论文
共 22 条
[1]
COMPUTER-ASSISTED PEDICLE SCREW FIXATION - A FEASIBILITY STUDY [J].
AMIOT, LP ;
LABELLE, H ;
DEGUISE, JA ;
SATI, M ;
BRODEUR, P ;
RIVARD, CH .
SPINE, 1995, 20 (10) :1208-1212
[2]
USE OF A FRAMELESS, ARMLESS STEREOTAXIC WAND FOR BRAIN-TUMOR LOCALIZATION WITH 2-DIMENSIONAL AND 3-DIMENSIONAL NEUROIMAGING [J].
BARNETT, GH ;
KORMOS, DW ;
STEINER, CP ;
WEISENBERGER, J .
NEUROSURGERY, 1993, 33 (04) :674-678
[3]
Camille RR., 1986, CLIN ORTHOP RELAT R, V203, P7
[4]
TRANSPEDICULAR SCREW-ROD FIXATION OF THE LUMBAR SPINE - OPERATIVE TECHNIQUE AND OUTCOME IN 104 CASES [J].
DICKMAN, CA ;
FESSLER, RG ;
MACMILLAN, M ;
HAID, RW .
JOURNAL OF NEUROSURGERY, 1992, 77 (06) :860-870
[5]
COMPLICATIONS ASSOCIATED WITH THE TECHNIQUE OF PEDICLE SCREW FIXATION - A SELECTED SURVEY OF ABS MEMBERS [J].
ESSES, SI ;
SACHS, BL ;
DREYZIN, V .
SPINE, 1993, 18 (15) :2231-2239
[6]
GALLOWAY RC, 1991, BIOMED INSTRUM TECHN, V255, P457
[7]
GALLOWAY RL, 1990, CRIT REV BIOMED ENG, V18, P181
[8]
GEORGE DC, 1991, SPINE, V16, P181, DOI 10.1097/00007632-199116020-00015
[9]
ACCURACY OF PEDICULAR SCREW PLACEMENT INVIVO [J].
GERTZBEIN, SD ;
ROBBINS, SE .
SPINE, 1990, 15 (01) :11-14
[10]
APPLICATION OF FRAMELESS STEREOTAXY TO PEDICLE SCREW FIXATION OF THE SPINE [J].
KALFAS, IH ;
KORMOS, DW ;
MURPHY, MA ;
MCKENZIE, RL ;
BARNETT, GH ;
BELL, GR ;
STEINER, CP ;
TRIMBLE, MB ;
WEISENBERGER, JP .
JOURNAL OF NEUROSURGERY, 1995, 83 (04) :641-647