Visualization of sample introduction in liquid chromatography columns - The effect of the frit diameter

被引:34
作者
Broyles, BS
Shalliker, RA
Guiochon, G [1 ]
机构
[1] Univ Tennessee, Dept Chem, Knoxville, TN 37996 USA
[2] Oak Ridge Natl Lab, Div Chem & Analyt Sci, Oak Ridge, TN 37831 USA
关键词
frit diameter; visualization of sample introduction; sample introduction;
D O I
10.1016/S0021-9673(99)00737-2
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A previously developed on-column detection technique using 35 mm SLR cameras [J. Chromatogr. A 826 (1998) 1] was employed to visualize colored sample bands as they elute through frits of differing diameter. Head fittings containing a 4.0 mm frit and a 15.9 mm frit were mounted in a 17 mm I.D. glass column packed with C-18 silica with an average particle size of 21 mu m. A carbon tetrachloride mobile phase of matching refractive index to that of the silica provides clarity along the column diameter during band migration. The photographs of the migrating sample zones were scanned and analyzed with appropriate imaging software. The smallest diameter frit induced severely parabolic sample distributions at the column inlet compared to the larger frit. Local axial dispersion coefficient values, expressed as local reduced plate height, were calculated as well as local zone velocities at the column inlet. The results demonstrate clearly the need to match the diameter of the inlet frit to the I.D. of the column so as to avoid the initial onset of zone broadening due to the frit. (C) 1999 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:367 / 382
页数:16
相关论文
共 8 条
[1]  
*ALLT, 1999, 450 ALLT
[2]  
*BODM IND, CHROM PROD GUID
[3]   Visualization of viscous fingering in chromatographic columns [J].
Broyles, BS ;
Shalliker', RA ;
Cherrak, DE ;
Guiochon, G .
JOURNAL OF CHROMATOGRAPHY A, 1998, 822 (02) :173-187
[4]  
BROYLES BS, UNPUB
[5]   SAMPLING AND EXTRA-COLUMN EFFECTS IN HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY - INFLUENCE OF PEAK SKEW ON PLATE COUNT CALCULATIONS [J].
KIRKLAND, JJ ;
YAU, WW ;
STOKLOSA, HJ ;
DILKS, CH .
JOURNAL OF CHROMATOGRAPHIC SCIENCE, 1977, 15 (08) :303-316
[6]   Visualization of solute migration in liquid chromatography columns [J].
Shalliker, RA ;
Broyles, BS ;
Guiochon, G .
JOURNAL OF CHROMATOGRAPHY A, 1998, 826 (01) :1-13
[7]  
Shalliker RA, 1998, AM LAB, V30, P124
[8]  
*UPCH SCI, 1999, 1999 CAT CHROM FLUID