A linear array silicon pixel detector: images of a mammographic test object and evaluation of delivered doses

被引:16
作者
Arfelli, F
Bonvicini, V
Bravin, A
Cantatore, G
Castelli, E
DallaPalma, L
DiMichiel, M
Longo, R
Olivo, A
Pani, S
Pontoni, D
Poropat, P
Prest, M
Rashevsky, A
Tromba, G
Vacchi, A
机构
[1] UNIV TRIESTE,DIPARTMENTO FIS,I-34127 TRIESTE,ITALY
[2] IST NAZL FIS NUCL,I-34127 TRIESTE,ITALY
[3] SOC SINCROTRONE,I-34012 TRIESTE,ITALY
[4] UNIV TRIESTE,OSPED CATTINARA,IST RADIOL,I-34100 TRIESTE,ITALY
关键词
D O I
10.1088/0031-9155/42/8/007
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We present images of a mammographic test object obtained using a linear array silicon pixel detector capable of single-photon counting. The detector pixel size was 200 x 300 mu m(2) and images were acquired by scanning the test object between the laminar detector and the x-ray source with a scanning step of 100 mu m. A molybdenum anode tube was used with two different filtrations: 2 mm aluminium and 25 mu m molybdenum. Conventional film-screen images were also obtained in order to compare spatial and contrast resolution. In our digital images it is possible to recognize low-contrast details having dimensions smaller than or equal to the dimensions of details visible by means of a clinical mammographic unit. The detection of microcalcifications smaller than 150 mu m was possible only when using the Mo filtration. However a copper wire of 50 mu m diameter was detectable when embedded in a simulated tissue. We discuss in detail the mean glandular doses (MGDs) delivered during the image acquisition. The MGDs necessary to obtain good-quality images are always smaller than at a conventional mammographic unit. Since MGDs depend on the x-ray spectrum, the dose reduction becomes larger when the applied spectrum is harder than in film-screen acquisition (Al filtration and 35 kVp).
引用
收藏
页码:1565 / 1573
页数:9
相关论文
共 20 条
[11]   A NEW PHANTOM FOR MAMMOGRAPHY [J].
LAW, J .
BRITISH JOURNAL OF RADIOLOGY, 1991, 64 (758) :116-120
[12]   DYNAMIC-RANGE REQUIREMENTS IN DIGITAL MAMMOGRAPHY [J].
MAIDMENT, ADA ;
FAHRIG, R ;
YAFFE, MJ .
MEDICAL PHYSICS, 1993, 20 (06) :1621-1633
[13]  
NCRP (National Council on Radiation Protection and Measurements), 1986, 85 NCRP
[14]  
PANI S, 1996, THESIS U TRIESTE
[15]   A PROTOTYPE AMORPHOUS SELENIUM IMAGING PLATE SYSTEM FOR DIGITAL RADIOGRAPHY [J].
PAPIN, PJ ;
HUANG, HK .
MEDICAL PHYSICS, 1987, 14 (03) :322-329
[16]  
Rose A, 1973, VISION HUMAN ELECT, P21
[17]   MAMMOGRAPHIC PARENCHYMAL PATTERNS - RISK INDICATOR FOR BREAST-CANCER [J].
TABAR, L ;
DEAN, PB .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1982, 247 (02) :185-189
[18]   Digital mammography: Performance considerations and current detector designs [J].
Williams, MB ;
Fajardo, LL .
ACADEMIC RADIOLOGY, 1996, 3 (05) :429-437
[19]  
YAFFE MJ, 1993, MED PROG TECHNOL, V19, P13
[20]  
YAFFE MJ, 1992, SYLLABUS CATEGORICAL, P69