Quantification of material loss from the neck piece taper junctions of a bimodular primary hip prosthesis. A retrieval study from 27 failed Rejuvenate bimodular hip arthroplasties

被引:22
作者
Buente, D. [1 ]
Huber, G. [1 ]
Bishop, N. [1 ]
Morlock, M. [1 ]
机构
[1] TUHH Hamburg Univ Technol, D-21073 Hamburg, Germany
关键词
METAL-IONS; CORROSION; FAILURE; CHROMIUM; COBALT; REPLACEMENT; RELEASE; STEM;
D O I
10.1302/0301-620X.97B10.35342
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
100224 [整形外科学];
摘要
The early failure and revision of bimodular primary total hip arthroplasty prostheses requires the identification of the risk factors for material loss and wear at the taper junctions through taper wear analysis. Deviations in taper geometries between revised and pristine modular neck tapers were determined using high resolution tactile measurements. A new algorithm was developed and validated to allow the quantitative analysis of material loss, complementing the standard visual inspection currently used. The algorithm was applied to a sample of 27 retrievals (in situ from 2.9 to 38.1 months) of the withdrawn Rejuvenate modular prosthesis. The mean wear volumes on the flat distal neck piece taper was 3.35 mm(3) (0.55 to 7.57), mainly occurring in a characteristic pattern in areas with high mechanical loading. Wear volume tended to increase with time to revision (r(2) = 0.423, p = 0.001). Implant and patient specific data (offset, stem size, patient's mass, age and body mass index) did not correlate with the amount of material loss observed (p > 0.078). Bilaterally revised implants showed higher amounts of combined total material loss and similar wear patterns on both sides. The consistent wear pattern found in this study has not been reported previously, suggesting that the device design and materials are associated with the failure of this prosthesis.
引用
收藏
页码:1350 / 1357
页数:8
相关论文
共 26 条
[1]
Australian Orthopaedic Association, Australian Orthopaedic Association National Joint Replacement Registry
[2]
Wear patterns of taper connections in retrieved large diameter metal-on-metal bearings [J].
Bishop, Nicholas ;
Witt, Florian ;
Pourzal, Robin ;
Fischer, Alfons ;
Ruetschi, Marcel ;
Michel, Markus ;
Morlock, Michael .
JOURNAL OF ORTHOPAEDIC RESEARCH, 2013, 31 (07) :1116-1122
[3]
CORROSION AND WEAR AT THE MODULAR INTERFACE OF UNCEMENTED FEMORAL STEMS [J].
COOK, SD ;
BARRACK, RL ;
CLEMOW, AJT .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1994, 76B (01) :68-72
[4]
Adverse Local Tissue Reaction Arising from Corrosion at the Femoral Neck-Body Junction in a Dual-Taper Stem with a Cobalt-Chromium Modular Neck [J].
Cooper, H. John ;
Urban, Robert M. ;
Wixson, Richard L. ;
Meneghini, R. Michael ;
Jacobs, Joshua J. .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2013, 95A (10) :865-872
[5]
Delaunay B., 1934, Classe des sciences mathmatiques et naturelles, V6, P793
[6]
IN-VIVO CORROSION OF MODULAR HIP-PROSTHESIS COMPONENTS IN MIXED AND SIMILAR METAL COMBINATIONS - THE EFFECT OF CREVICE, STRESS, MOTION, AND ALLOY COUPLING [J].
GILBERT, JL ;
BUCKLEY, CA ;
JACOBS, JJ .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1993, 27 (12) :1533-1544
[7]
Corrosion at the neck-stem junction as a cause of metal ion release and pseudotumour formation [J].
Gill, I. P. S. ;
Webb, J. ;
Sloan, K. ;
Beaver, R. J. .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2012, 94B (07) :895-900
[8]
Risk factors for inflammatory pseudotumour formation following hip resurfacing [J].
Glyn-Jones, S. ;
Pandit, H. ;
Kwon, Y. -M. ;
Doll, H. ;
Gill, H. S. ;
Murray, D. W. .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2009, 91B (12) :1566-1574
[9]
Goldberg JR, 2002, CLIN ORTHOP RELAT R, P149
[10]
Modular titanium alloy neck adapter failures in hip replacement - failure mode analysis and influence of implant material [J].
Grupp, Thomas M. ;
Weik, Thomas ;
Bloemer, Wilhelm ;
Knaebel, Hanns-Peter .
BMC MUSCULOSKELETAL DISORDERS, 2010, 11