Low kilovoltage cardiac dual-source CT: attenuation, noise, and radiation dose

被引:363
作者
Leschka, Sebastian [1 ,2 ]
Stolzmann, Paul [1 ]
Schmid, Florian T. [2 ]
Scheffel, Hans [1 ]
Stinn, Bjoern [2 ]
Marincek, Borut [1 ]
Alkadhi, Hatem [1 ]
Wildermuth, Simon [2 ]
机构
[1] Univ Zurich Hosp, Inst Diagnost Radiol, CH-8091 Zurich, Switzerland
[2] Kantonsspital, Inst Radiol, St Gallen, Switzerland
关键词
dual-source CT coronary angiography; low kilovoltage; radiation exposure; contrast-to-noise ratio;
D O I
10.1007/s00330-008-0966-1
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 [临床医学]; 100207 [影像医学与核医学]; 1009 [特种医学];
摘要
The purpose of this study was to investigate the effect of low kilovoltage dual-source computed tomography coronary angiography (CTCA) on qualitative and quantitative image quality parameters and radiation dose. Dual-source CTCA with retrospective ECG gating was performed in 80 consecutive patients of normal weight. Forty were examined with a standard protocol (120 kV/330mAs), 20 were examined at 100 kV/330mAs, and 20 at 100 kV/220mAs. Two blinded observers independently assessed image quality of each coronary segment and measured the image parameters noise, attenuation, and contrast-to-noise ratio (CNR). The effective radiation dose was calculated using CT dose volume index and the dose-length product. Diagnostic image quality was obtained in 99% of all coronary segments (1,127/1,140) without significant differences among the protocols. Image noise, attenuation, and CNR were significantly higher for 100 kV/330mAs (26+/-3 HU, 549+/-62 HU, 25.5+/-3.2; each P < 0.01) and 100 kV/220mAs (27+/-2 HU, 560+/-43 HU, 25.0+/-2.2; each P < 0.01) when compared to the 120-kV protocol (21+/-2 HU, 317+/-28 HU, 20.6+/-1.7). There was no significant difference between the two 100-kV protocols. Estimated effective radiation dose of the 120-kV protocol (8.9+/-1.2 mSv) was significantly higher than the 100 kV/330mAs (6.7+/-0.8 mSv, P < 0.01) or 100 kV/220mAs (4.4+/-0.6 mSv, P < 0.001) protocols. Dual-source CTCA with 100 kV is feasible in patients of normal weight, results in a diagnostic image quality with a higher CNR, and at the same time significantly reduces the radiation dose.
引用
收藏
页码:1809 / 1817
页数:9
相关论文
共 32 条
[1]
MDCT of the coronary arteries: Feasibility of low-dose CT with ECG-pulsed tube current modulation to reduce radiation dose [J].
Abada, Hicham T. ;
Larchez, Christophe ;
Daoud, Beatrice ;
Sigal-Cinqualbre, Anne ;
Paul, Jean-Francois .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2006, 186 (06) :S387-S390
[2]
Austen W G, 1975, Circulation, V51, P5
[3]
Dose reduction in pediatric CT: A rational approach [J].
Boone, JM ;
Geraghty, EM ;
Seibert, JA ;
Wootton-Gorges, SL .
RADIOLOGY, 2003, 228 (02) :352-360
[4]
QUANTITATIVE THEORY OF HOUNSFIELD UNIT AND ITS APPLICATION TO DUAL ENERGY SCANNING [J].
BROOKS, RA .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1977, 1 (04) :487-493
[5]
Assessment of coronary artery disease by cardiac computed tomography - A scientific statement from the American Heart Association committee on cardiovascular imaging and intervention, council on cardiovascular radiology and intervention, and Committee on Cardiac Imaging, Council on Clinical Cardiology [J].
Budoff, Matthew J. ;
Achenbach, Stephan ;
Blumenthal, Roger S. ;
Carr, J. Jeffrey ;
Goldin, Jonathan G. ;
Greenland, Philip ;
Guerci, Alan D. ;
Lima, Joao A. C. ;
Rader, Daniel J. ;
Rubin, Geoffrey D. ;
Shaw, Leslee J. ;
Wiegers, Susan E. .
CIRCULATION, 2006, 114 (16) :1761-1791
[6]
Use of automatic exposure control in multislice computed tomography of the coronaries:: comparison of 16-slice and 64-slice scanner data with conventional coronary angiography [J].
Deetjen, Anja ;
Moellmann, Susanne ;
Conradi, Guido ;
Rolf, Andreas ;
Schmermund, Axel ;
Hamm, Christian W. ;
Dill, Thorsten .
HEART, 2007, 93 (09) :1040-1043
[7]
Quantitative parameters of image quality in 64-slice computed tomography angiography of the coronary arteries [J].
Ferencik, M ;
Nomura, CH ;
Maurovich-Horvat, P ;
Hoffmann, U ;
Pena, AJ ;
Cury, RC ;
Abbara, S ;
Nieman, K ;
Fatima, U ;
Achenbach, S ;
Brady, TJ .
EUROPEAN JOURNAL OF RADIOLOGY, 2006, 57 (03) :373-379
[8]
First performance evaluation of a dual-source CT (DSCT) system [J].
Flohr, TG ;
McCollough, CH ;
Bruder, H ;
Petersilka, M ;
Gruber, K ;
Süss, C ;
Grasruck, M ;
Stierstorfer, K ;
Krauss, B ;
Raupach, R ;
Primak, AN ;
Küttner, A ;
Achenbach, S ;
Becker, C ;
Kopp, A ;
Ohnesorge, BM .
EUROPEAN RADIOLOGY, 2006, 16 (02) :256-268
[9]
Techniques and parameters for estimating radiation exposure and dose in cardiac computed tomography [J].
Gerber, TC ;
Kuzo, RS ;
Morin, RL .
INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2005, 21 (01) :165-176
[10]
Radiation dose estimates from cardiac multislice computed tomography in daily practice -: Impact of different scanning protocols on effective dose estimates [J].
Hausleiter, J ;
Meyer, T ;
Hadamitzky, M ;
Huber, E ;
Zankl, M ;
Martinoff, S ;
Kastrati, A ;
Schömig, A .
CIRCULATION, 2006, 113 (10) :1305-1310