Thermoluminescent dosimetry of the selectSeed 125I interstitial brachytherapy seed

被引:36
作者
Anagnostopoulos, G
Baltas, D [1 ]
Karaiskos, P
Sandilos, P
Papagiannis, P
Sakelliou, L
机构
[1] Klinikum Offenbach, Strahlenklin, Dept Med Phys & Engn, D-63069 Offenbach, Germany
[2] Natl Tech Univ Athens, Inst Commun & Comp Syst, GR-15773 Athens, Greece
[3] Hygeia Hosp, Dept Med Phys, Athens 15123, Greece
[4] Univ Athens, Dept Phys, Nucl & Particle Phys Sect, GR-15771 Athens, Greece
关键词
I-125; selectSeed; thermoluminescent dosimetry; Monte Carlo; brachytherapy;
D O I
10.1118/1.1469631
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
This work presents experimental dosimetry results for the new selectSeed I-125 prostate seed design for use with the seedSelectron afterloading device, in accordance with the AAPM advisory that all new low energy interstitial brachytherapy seeds should undergo one Monte Carlo (MC) and at least one experimental dosimetry characterization. TLD dosimetry was per-formed using 120 cylindrical LiF TLD type-100 rods calibrated using a 6 MV photon beam. They were irradiated in solid water phantoms for the experimental determination of the seed dose rate constant, radial dose functions and anisotropy functions. MC simulations were performed for the determination of the TLDs relative energy response that was found position independent and equal to 1.40+/-0.03, and for the calculation of the ratio of dose in liquid water to dose in solid water that was found to be well described by D-liquid water/D-solid (water) = 1.013*r+0.030 presenting only a minor dependence on polar angle. The selectSeed dose rate constant in liquid water was found equal to 0.938+/-0.065 cGy h(-1) U-1, which agrees within experimental uncertainties with corresponding MC results of Lambda(select Seed) = 0.954+/-0.005 cGy h(-1) U-1. The experimental radial dose and anisotropy function results were also found in good agreement with corresponding MC calculations. (C) 2002 American Association of Physicists in Medicine.
引用
收藏
页码:709 / 716
页数:8
相关论文
共 29 条
[1]   Application of the Monte Carlo integration (MCI) method for calculation of the anisotropy of 192Ir brachytherapy sources [J].
Baltas, D ;
Giannouli, S ;
Garbi, A ;
Diakonos, F ;
Geramani, K ;
Ioannidis, GT ;
Tsalpatouros, A ;
Uzunoglu, N ;
Kolotas, C ;
Zamboglou, N .
PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (06) :1783-1801
[2]   Spectra and air-kerma strength for encapsulated 192Ir sources [J].
Borg, J ;
Rogers, DWO .
MEDICAL PHYSICS, 1999, 26 (11) :2441-2444
[3]   TWO-DIMENSIONAL DOSE DISTRIBUTION AROUND A COMMERCIAL I125 SEED [J].
BURNS, GS ;
RAESIDE, DE .
MEDICAL PHYSICS, 1988, 15 (01) :56-60
[4]   Anisotropy functions for low energy interstitial brachytherapy sources:: an EGS4 Monte Carlo study [J].
Capote, R ;
Mainegra, E ;
López, E .
PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (01) :135-150
[5]   DOSE-RATE DETERMINATION FOR I-125 SEEDS [J].
CHIUTSAO, ST ;
ANDERSON, LL ;
OBRIEN, K ;
SANNA, R .
MEDICAL PHYSICS, 1990, 17 (05) :815-825
[6]   A SOLID WATER PHANTOM MATERIAL FOR RADIOTHERAPY X-RAY AND GAMMA-RAY BEAM CALIBRATIONS [J].
CONSTANTINOU, C ;
ATTIX, FH ;
PALIWAL, BR .
MEDICAL PHYSICS, 1982, 9 (03) :436-441
[7]  
DILLMAN LT, 10 NM MIRD
[8]  
Hubbel J H, 1995, 5632 NISTIR
[9]  
Hubbell J. H., 1975, Journal of Physical and Chemical Reference Data, V4, P471, DOI 10.1063/1.555523
[10]   RELATIVISTIC ATOMIC FORM-FACTORS AND PHOTON COHERENT SCATTERING CROSS-SECTIONS [J].
HUBBELL, JH ;
OVERBO, I .
JOURNAL OF PHYSICAL AND CHEMICAL REFERENCE DATA, 1979, 8 (01) :69-105