Thermoluminescent dosimetry of the selectSeed 125I interstitial brachytherapy seed

被引:36
作者
Anagnostopoulos, G
Baltas, D [1 ]
Karaiskos, P
Sandilos, P
Papagiannis, P
Sakelliou, L
机构
[1] Klinikum Offenbach, Strahlenklin, Dept Med Phys & Engn, D-63069 Offenbach, Germany
[2] Natl Tech Univ Athens, Inst Commun & Comp Syst, GR-15773 Athens, Greece
[3] Hygeia Hosp, Dept Med Phys, Athens 15123, Greece
[4] Univ Athens, Dept Phys, Nucl & Particle Phys Sect, GR-15771 Athens, Greece
关键词
I-125; selectSeed; thermoluminescent dosimetry; Monte Carlo; brachytherapy;
D O I
10.1118/1.1469631
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
This work presents experimental dosimetry results for the new selectSeed I-125 prostate seed design for use with the seedSelectron afterloading device, in accordance with the AAPM advisory that all new low energy interstitial brachytherapy seeds should undergo one Monte Carlo (MC) and at least one experimental dosimetry characterization. TLD dosimetry was per-formed using 120 cylindrical LiF TLD type-100 rods calibrated using a 6 MV photon beam. They were irradiated in solid water phantoms for the experimental determination of the seed dose rate constant, radial dose functions and anisotropy functions. MC simulations were performed for the determination of the TLDs relative energy response that was found position independent and equal to 1.40+/-0.03, and for the calculation of the ratio of dose in liquid water to dose in solid water that was found to be well described by D-liquid water/D-solid (water) = 1.013*r+0.030 presenting only a minor dependence on polar angle. The selectSeed dose rate constant in liquid water was found equal to 0.938+/-0.065 cGy h(-1) U-1, which agrees within experimental uncertainties with corresponding MC results of Lambda(select Seed) = 0.954+/-0.005 cGy h(-1) U-1. The experimental radial dose and anisotropy function results were also found in good agreement with corresponding MC calculations. (C) 2002 American Association of Physicists in Medicine.
引用
收藏
页码:709 / 716
页数:8
相关论文
共 29 条
[21]   In-phantom response of LiF TLD-100 for dosimetry of 192Ir HDR source [J].
Pradhan, AS ;
Quast, U .
MEDICAL PHYSICS, 2000, 27 (05) :1025-1029
[22]   Dosimetric study of the new InterSource125 iodine seed [J].
Reniers, B ;
Vynckier, S ;
Scalliet, P .
MEDICAL PHYSICS, 2001, 28 (11) :2285-2288
[23]  
SCOFIELD JH, 1973, UCRL51326 LAWR LIV L
[24]   Report on the dosimetry of a new design 125Iodine brachytherapy source [J].
Wallace, RE ;
Fan, JJ .
MEDICAL PHYSICS, 1999, 26 (09) :1925-1931
[25]   Dosimetric prerequisites for routine clinical use of new low energy photon interstitial brachytherapy sources [J].
Williamson, J ;
Coursey, BM ;
DeWerd, LA ;
Hanson, WF ;
Nath, R .
MEDICAL PHYSICS, 1998, 25 (12) :2269-2270
[26]  
Williamson J S, 1988, Reg Immunol, V1, P15
[27]   Guidance to users of Nycomed Amersham and North American Scientific, Inc., I-125 Interstitial Sources: Dosimetry and calibration changes: Recommendations of the American Association of Physicists in Medicine Radiation Therapy Committee Ad Hoc Subcommittee on Low-Energy Seed Dosimetry [J].
Williamson, JF ;
Coursey, BM ;
DeWerd, LA ;
Hanson, WF ;
Nath, R ;
Ibbott, G .
MEDICAL PHYSICS, 1999, 26 (04) :570-573
[28]   COMPARISON OF MEASURED AND CALCULATED DOSE-RATES IN WATER NEAR I-125 AND IR-192 SEEDS [J].
WILLIAMSON, JF .
MEDICAL PHYSICS, 1991, 18 (04) :776-786
[29]  
WILLIAMSON JF, 1994, BRACHYTHERAPY PHYSIC, V95