First clinical trial of the "MiRo" capsule endoscope by using a novel transmission technology: electric-field propagation

被引:94
作者
Bang, Seungmin [1 ]
Park, Jeong Youp [1 ]
Jeong, Seok [2 ]
Kim, Young Ho [3 ]
Shim, Han Bo [3 ]
Kim, Tae Song [4 ]
Lee, Don Haeng [2 ]
Song, Si Young [1 ]
机构
[1] Yonsei Univ, Coll Med, Div Gastroenterol, Dept Internal Med,Inst Gastroenterol, Seoul 120752, South Korea
[2] Inha Univ, Dept Internal Med, Inchon, South Korea
[3] IntroMed Co Ltd, Seoul, South Korea
[4] Korea Inst Sci & Technol, Intelligent Microsyst Ctr, Seoul, South Korea
关键词
PILLCAM COLON CAPSULE; PUSH-ENTEROSCOPY; COLONOSCOPY;
D O I
10.1016/j.gie.2008.04.033
中图分类号
R57 [消化系及腹部疾病];
学科分类号
摘要
Background: We developed a capsule endoscope (CE), "MiRo," with the novel transmission technology of electric-field propagation. The technology uses the human body as a conductive medium for data transmission. Specifications of the prototype include the ability to receive real-time images; size, 10.8 x 24 mm; weight, 3.3 g; field of view, 150 degrees; resolution of power, 320 X 320 pixels; and transmittal speed, 2 frames per second. Objective: To evaluate the clinical safety and diagnostic feasibility of the prototype MiRo, we conducted a multicenter clinical trial. Design and Patients: All volunteers underwent baseline examinations, including EGD and electrocardiography for the screening of GI obstructive and cardiovascular diseases, before the trial. in the first 10 cases, 24-hour Holter monitoring was also performed. To evaluate the diagnostic feasibility, transmission rate of the captured images, inspection rate of the entire small bowel, and quality of transmitted images (graded as outstanding, excellent, good/average, below average, and poor) were analyzed. Results: Of the 49 healthy volunteers, 45 were included in the trial, and 4 were excluded because of baseline abnormalities. No adverse effects were noted. All CEs were expelled within 2 clays, and the entire small bowel could be explored in all cases. The transmission rates of the captured image in the stomach, small bowel, and colon were 99.5%, 99.6%, and 97.2%, respectively The mean total duration of image transmission was 9 hours, 51 minutes, and the mean transit time of the entire small bowel was 4 hours, 33 minutes. Image quality was graded as good or better in 41 cases (91.1%). Details of the villi and vascular structures of the entire small bowel were clearly visualized in 31 cases (68.9%). Conclusions: MiRo is safe and effective for exploring the entire small bowel, with good image quality and real-time feasibility This novel transmission technology may have applications beyond the field of capsule endoscopy (Gastrointest Endosc 2009;69:253-9.)
引用
收藏
页码:253 / 259
页数:7
相关论文
共 12 条
[1]   A randomized trial comparing wireless capsule endoscopy with push enteroscopy for the detection of small-bowel lesions [J].
Appleyard, M ;
Fireman, Z ;
Glukhovsky, A ;
Jacob, H ;
Shreiver, R ;
Kadirkamanathan, S ;
Lavy, A ;
Lewkowicz, S ;
Scapa, E ;
Shofti, R ;
Swain, P ;
Zaretsky, A .
GASTROENTEROLOGY, 2000, 119 (06) :1431-1438
[2]   Evaluation of the PillCam Colon capsule in the detection of colonic pathology: results of the first multicenter, prospective, comparative study [J].
Eliakim, R. ;
Fireman, Z. ;
Gralnek, I. M. ;
Yassin, K. ;
Waterman, M. ;
Kopelman, Y. ;
Lachter, J. ;
Koslowsky, B. ;
Adler, S. N. .
ENDOSCOPY, 2006, 38 (10) :963-970
[3]  
Haykin S., 2007, INTRO ANALOG DIGITAL, VSecond
[4]   Wireless capsule endoscopy [J].
Iddan, G ;
Meron, G ;
Glukhovsky, A ;
Swain, P .
NATURE, 2000, 405 (6785) :417-417
[5]  
Kim T.-S., 2007, U.S. Patent, Patent No. [7307544B2, 7307544, US 7,307,544 B2]
[6]  
KIM TS, 2007, C P IEEE ENG MED BIO, P2823
[7]   The development of the swallowable video capsule (M2A) [J].
Meron, GD .
GASTROINTESTINAL ENDOSCOPY, 2000, 52 (06) :817-819
[8]   Wireless capsule endoscopy: a comparison with push enteroscopy in patients with gastroscopy and colonoscopy negative gastrointestinal bleeding [J].
Mylonaki, M ;
Fritscher-Ravens, A ;
Swain, P .
GUT, 2003, 52 (08) :1122-1126
[9]  
Saurin JC, 2003, ENDOSCOPY, V35, P576
[10]   PillCam colon capsule endoscopy compared with colonoscopy for colorectal tumor diagnosis:: a prospective pilot study [J].
Schoofs, N. ;
Deviere, J. ;
Van Gossum, A. .
ENDOSCOPY, 2006, 38 (10) :971-977