Fretting Crevice Corrosion of Stainless Steel Stem-CoCr Femoral Head Connections: Comparisons of Materials, Initial Moisture, and Offset Length

被引:86
作者
Gilbert, Jeremy L. [1 ]
Mehta, Manav [1 ]
Pinder, Bryan [1 ]
机构
[1] Syracuse Univ, Dept Biomed & Chem Engn, Syracuse, NY 13244 USA
关键词
modular taper; fretting corrosion; stainless steel; hip prostheses; MODULAR HIP PROSTHESES; TITANIUM-ALLOY STEMS; INTERFACE; PRODUCTS; COMPONENTS; IMPLANTS; TAPERS;
D O I
10.1002/jbm.b.31164
中图分类号
R318 [生物医学工程];
学科分类号
100103 [病原生物学];
摘要
Modular tapers continue to be used in a wide variety of orthopedic implants. In this study, stainless steel (ASTM F-1568) femoral hip stems combined with Co-Cr-Mo alloy heads (SS/CoCr) were tested in an in vitro fretting corrosion test set-up to assess the propensity for mechanically assisted corrosion. Three different aspects of the modular design were evaluated in this study: (1) material combination compared to CoCr/CoCr, (2) wet versus dry assembly for SS/CoCr couples, and (3) 0- and 6-mm head offset for SS/CoCr couples. Fretting corrosion tests over a range of cyclic loads up to 3300 N were performed, and continuous cyclic loading at 3300 N for 1 M cycles were performed on each group (n = 5). Fretting micromotion was measured as a function of cyclic load on select couples to detect the nature and extent of motion present. The results showed that SS/CoCr couples were more susceptible to fretting corrosion than CoCr/CoCr couples, that dry assembly does not prevent fretting corrosion from taking place but raises the onset load, and that 6-mm offset heads had higher visual evidence of fretting damage but showed mixed statistical results in terms of onset loads and OCP shifts and currents compared to the 0-mm offset samples. Current and voltage excursions over 1million cycles tended to diminish towards their unloaded control levels but did not fully recover until cyclic loading ceased. Micromotion measurements indicated fretting motions in the range of 10-25 pm where 0-mm heads tended to piston oil the trunion,while 6 mm heads tended to rock. (c) 2008 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Bionnater 88B: 162-173 2009
引用
收藏
页码:162 / 173
页数:12
相关论文
共 19 条
[1]
Effects of neck extension, coverage, and frequency on the fretting corrosion of modular THR bore and cone interface [J].
Brown, SA ;
Abera, A ;
DOnofrio, M ;
Flemming, C .
MODULARITY OF ORTHOPEDIC IMPLANTS, 1997, 1301 :189-198
[2]
FRETTING CORROSION ACCELERATES CREVICE CORROSION OF MODULAR HIP TAPERS [J].
BROWN, SA ;
FLEMMING, CAC ;
KAWALEC, JS ;
PLACKO, HE ;
VASSAUX, C ;
MERRITT, K ;
PAYER, JH ;
KRAAY, MJ .
JOURNAL OF APPLIED BIOMATERIALS, 1995, 6 (01) :19-26
[3]
Assessment of wear on the cones of modular stainless steel Exeter hip stems [J].
Chaplin, RPS ;
Lee, AJC ;
Hooper, RM .
JOURNAL OF MATERIALS SCIENCE-MATERIALS IN MEDICINE, 2004, 15 (09) :977-990
[4]
COLLIER JP, 1991, CLIN ORTHOP RELAT R, P305
[5]
CORROSION BETWEEN THE COMPONENTS OF MODULAR FEMORAL HIP PROSTHESES [J].
COLLIER, JP ;
SURPRENANT, VA ;
JENSEN, RE ;
MAYOR, MB ;
SURPRENANT, HP .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1992, 74 (04) :511-517
[6]
Gilbert J., 2006, ASM INT HDB CORROSIO, P826
[7]
Gilbert J.L., 1994, TOTAL HIP REVISION S, P41
[8]
IN-VIVO CORROSION OF MODULAR HIP-PROSTHESIS COMPONENTS IN MIXED AND SIMILAR METAL COMBINATIONS - THE EFFECT OF CREVICE, STRESS, MOTION, AND ALLOY COUPLING [J].
GILBERT, JL ;
BUCKLEY, CA ;
JACOBS, JJ .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1993, 27 (12) :1533-1544
[9]
In vitro corrosion testing of modular hip tapers [J].
Goldberg, JR ;
Gilbert, JL .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2003, 64B (02) :78-93
[10]
Corrosion testing of modular hip implants [J].
Goldberg, JR ;
Buckley, CA ;
Jacobs, JJ ;
Gilbert, JL .
MODULARITY OF ORTHOPEDIC IMPLANTS, 1997, 1301 :157-176