Simultaneous voltammetric measurement of ascorbic acid, epinephrine and uric acid at a glassy carbon electrode modified with caffeic acid

被引:230
作者
Ren, W [1 ]
Luo, HQ [1 ]
Li, NB [1 ]
机构
[1] SW China Normal Univ, Sch Chem & Chem Engn, Chongqing 400715, Peoples R China
关键词
poly(caffeic acid) modified electrode; ascorbic acid; epinephrine; uric acid;
D O I
10.1016/j.bios.2005.04.002
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
A stable electroactive thin film of poly(caffeic acid) has been deposited on the surface of a glassy carbon electrode by potentiostatic technique in an aqueous solution containing caffeic acid. Poly(caffeic acid) was used as a modified electrode for the detection of ascorbic acid (AA), epinephrine (EP), uric acid (UA) and their mixture by cyclic voltammetry. This modified electrode exhibits potent and persistent electronmediating behavior followed by well-separated oxidation peaks towards AA, EP and UA with activation overpotential. For the ternary mixture containing AA, EP and UA, the three compounds can well separate from each other at the scan rate of 20 mV s(-1) with a potential difference of 156, 132 and 288 mV between AA and EP, EP and UA and AA and UA, respectively, which was large enough to determine AA, EP and UA individually and simultaneously. The catalytic peak current obtained, was linearly dependent on the AA, EP and UA concentrations in the range of 2.0 x 10(-5) to 1.0 X 10(-3) Mol l(-1), 2.0 x 10(-6) to 8.0 X 10(-5) mol l(-1) and 5.0 x 10(-6) to 3.0 x 10(-4) Mol l(-1), and the detection limits for AA, EP and UA were 7.0 x 10(-6) 2.0 x 10(-7) and 6.0 x 10(-7) mol l(-1), respectively. The modified electrode shows good sensitivity, selectivity and stability, and has been applied to the determination of EP in practical injection samples and that of EP, UA and AA simultaneously with satisfactory results. (c) 2005 Elsevier B.V. All rights reserved.
引用
收藏
页码:1086 / 1092
页数:7
相关论文
共 28 条
[1]   Capability of a carbon-polyvinylchloride composite electrode for the detection of dopamine, ascorbic acid and uric acid [J].
Aguilar, R ;
Dávila, MM ;
Elizalde, MP ;
Mattusch, J ;
Wennrich, R .
ELECTROCHIMICA ACTA, 2004, 49 (06) :851-859
[2]   Ascorbic acid: much more than just an antioxidant [J].
Arrigoni, O ;
De Tullio, MC .
BIOCHIMICA ET BIOPHYSICA ACTA-GENERAL SUBJECTS, 2002, 1569 (1-3) :1-9
[3]   FLOW-INJECTION CHEMILUMINOMETRIC DETERMINATION OF EPINEPHRINE, NOREPINEPHRINE, DOPAMINE AND L-DOPA [J].
DEFTEREOS, NT ;
CALOKERINOS, AC ;
EFSTATHIOU, CE .
ANALYST, 1993, 118 (06) :627-632
[4]  
DON SJ, 1995, CHEM MODIFIED ELECT
[5]   Selective voltammetric detection of epinephrine in the presence of ascorbic acid at poly(4-methoxyphenol)-modified electrode [J].
Erdogdu, G .
JOURNAL OF ANALYTICAL CHEMISTRY, 2002, 57 (07) :620-623
[6]   Cyclic voltammetry determination of epinephrine with a carbon fiber ultramicroelectrode [J].
Hernandez, P ;
Sanchez, I ;
Paton, F ;
Hernandez, L .
TALANTA, 1998, 46 (05) :985-991
[7]   Simultaneous measurement of epinephrine and ascorbic acid at the carbon nanotube electrode [J].
Hu, CG ;
Wang, WL ;
Liao, KJ ;
Wang, YT .
SCIENCE AND TECHNOLOGY OF NANOMATERIALS - ICMAT 2003, 2005, 23 :305-308
[8]   ACTION OF BIOLOGICALLY-RELEVANT OXIDIZING SPECIES UPON URIC-ACID - IDENTIFICATION OF URIC-ACID OXIDATION-PRODUCTS [J].
KAUR, H ;
HALLIWELL, B .
CHEMICO-BIOLOGICAL INTERACTIONS, 1990, 73 (2-3) :235-247
[9]   Modified microelectrodes and multivariate calibration for flow injection amperometric simultaneous determination of ascorbic acid, dopamine, epinephrine and dipyrone [J].
Matos, RC ;
Angnes, L ;
Araújo, MCU ;
Saldanha, TCB .
ANALYST, 2000, 125 (11) :2011-2015
[10]   Flow injection analysis-amperometric determination of ascorbic and uric acids in urine using arrays of gold microelectrodes modified by electrodeposition of palladium [J].
Matos, RC ;
Augelli, MA ;
Lago, CL ;
Angnes, L .
ANALYTICA CHIMICA ACTA, 2000, 404 (01) :151-157