Development of continuous on-line purge and trap analysis

被引:15
作者
Saridara, C
Brukh, R
Mitra, S
机构
[1] New Jersey Inst Technol, Dept Chem & Environm Sci, Newark, NJ 07102 USA
[2] Chulalongkorn Univ, Int Postgrad Programs Environm Management, Bangkok, Thailand
关键词
D O I
10.1002/jssc.200401897
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An on-line purge and trap system for continuous monitoring of Volatile Organic Compounds (VOC) is presented. The purge chamber was designed for continuous extraction of VOC from water with nitrogen. The analytes were preconcentrated on a microtrap prior to analysis by GC with flame ionization detection. The microtrap served as a fast injection device for carrying out analysis at high frequency. Continuous monitoring was accomplished by performing injections at fixed intervals. This system showed high sensitivity, high precision, detection limits at the ppb level, and stable response over long periods of continuous operation. Factors affecting system performance were studied. A predictive model based on gas-liquid partitioning is also presented.
引用
收藏
页码:446 / 452
页数:7
相关论文
共 30 条
[1]   Real-time analysis of methanol in air and water by membrane introduction mass spectrometry [J].
Allen, TM ;
Falconer, TM ;
Cisper, ME ;
Borgerding, AJ ;
Wilkerson, CW .
ANALYTICAL CHEMISTRY, 2001, 73 (20) :4830-4835
[2]  
BELLAR TA, 1974, J AM WATER WORKS ASS, V66, P739
[3]   MEMBRANE INTERFACE FOR SELECTIVE INTRODUCTION OF VOLATILE COMPOUNDS DIRECTLY INTO THE IONIZATION-CHAMBER OF A MASS-SPECTROMETER [J].
BIER, ME ;
COOKS, RG .
ANALYTICAL CHEMISTRY, 1987, 59 (04) :597-601
[4]   Process modeling and on-line monitoring of benzene and other species during the two-stage combustion of ethylene in air [J].
Brukh, R ;
Salem, T ;
Slanvetpan, T ;
Barat, R ;
Mitra, S .
ADVANCES IN ENVIRONMENTAL RESEARCH, 2002, 6 (03) :359-367
[5]  
Feng CH, 2000, J MICROCOLUMN SEP, V12, P267, DOI 10.1002/(SICI)1520-667X(2000)12:4<267::AID-MCS11>3.0.CO
[6]  
2-B
[7]   Two-stage microtrap as an injection device for continuous on-line gas chromatographic monitoring [J].
Feng, CH ;
Mitra, S .
JOURNAL OF CHROMATOGRAPHY A, 1998, 805 (1-2) :169-176
[8]   THE MICRO-TRAP - AN ALTERNATIVE TO CRYOFOCUSING IN CAPILLARY GAS-CHROMATOGRAPHY [J].
FRANK, W ;
FRANK, H .
CHROMATOGRAPHIA, 1990, 29 (11-12) :571-574
[9]   Development of pulse introduction membrane extraction for analysis of volatile organic compounds in individual aqueous samples, and for continuous on-line monitoring [J].
Guo, XM ;
Mitra, S .
JOURNAL OF CHROMATOGRAPHY A, 1998, 826 (01) :39-47
[10]   Determination of volatile organic compounds in water by solid phase microextraction and infrared spectroscopy [J].
Heglund, DL ;
Tilotta, DC .
ENVIRONMENTAL SCIENCE & TECHNOLOGY, 1996, 30 (04) :1212-1219