Interference of the electrospray voltage on chromatographic separations using porous graphitic carbon columns

被引:27
作者
Törnkvist, A
Nilsson, S
Amirkhani, A
Nyholm, LM
Nyholm, L
机构
[1] Univ Uppsala, Dept Analyt Chem, SE-75124 Uppsala, Sweden
[2] Amersham Biosci, SE-75184 Uppsala, Sweden
来源
JOURNAL OF MASS SPECTROMETRY | 2004年 / 39卷 / 02期
关键词
catecholamine; dopamine; electrochemistry; porous graphitic carbon; electrospray; mass spectrometry; liquid chromatography;
D O I
10.1002/jms.589
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The electrospray ionization (ESI) voltage is shown to interfere with liquid chromatographic separations performed with packed porous graphitic carbon (PGC) capillary columns. This interference is ascribed to the presence of an electric field over the conductive column in the absence of an earth point between the column and the ESI emitter. The current evolved alters the chromatographic behavior of the catecholamine metabolite 3-O-methyl-DOPA significantly, as both peak splitting and a dramatic decrease in the retention time were observed. Furthermore, the response from the mass spectrometer was decreased by 33% at the same time. A related compound, tyrosine, exhibited decreased retention times but no peak splitting, whereas no shifts in the retention times (or peak splitting) were seen for the less retained dopamine and noradrenaline. When the current through the PGC column was eliminated by the use of an earth point between the column and the ESI emitter, the chromatographic behavior of the column was found to return slowly to normal after hours of equilibration with 60:40 (v/v) methanol -ammonium formate buffer of pH 2.9. The behavior of the PGC column with and without the earth point was found to be highly reproducible during a period of I month. We propose that the effect of the ESI voltage on the chromatographic behavior of the PGC column is due to associated redox reactions affecting both the PGC particles and the analytes. It is concluded that (for analytical reasons), care should be taken to ensure that no current is flowing through the chromatographic system when interfacing PGC columns, and conducting parts in general, to ESI mass spectrometry. Copyright (C) 2004 John Wiley Sons, Ltd.
引用
收藏
页码:216 / 222
页数:7
相关论文
共 14 条
[1]   A wireless electrochemiluminescence detector applied to direct and indirect detection for electrophoresis on a microfabricated glass device [J].
Arora, A ;
Eijkel, JCT ;
Morf, WE ;
Manz, A .
ANALYTICAL CHEMISTRY, 2001, 73 (14) :3282-3288
[2]   MECHANISM OF ELECTROSPRAY MASS-SPECTROMETRY - ELECTROSPRAY AS AN ELECTROLYSIS CELL [J].
BLADES, AT ;
IKONOMOU, MG ;
KEBARLE, P .
ANALYTICAL CHEMISTRY, 1991, 63 (19) :2109-2114
[3]   Electrochemically modulated liquid chromatography coupled on-line with electrospray mass spectrometry [J].
Deng, HT ;
Van Berkel, OJ .
ANALYTICAL CHEMISTRY, 2000, 72 (11) :2641-2647
[4]   Electrochemically modulated liquid chromatography: an electrochemical strategy for manipulating chromatographic retention [J].
Harnisch, JA ;
Porter, MD .
ANALYST, 2001, 126 (11) :1841-1849
[5]  
Kebarle P. A., 1993, ANAL CHEM, V65, P972
[6]  
Kinoshita Kim., 1988, CARBON ELECTROCHEMIC
[7]   Electrochemically induced pH changes resulting in protein unfolding in the ion source of an electrospray mass spectrometer [J].
Konermann, L ;
Silva, EA ;
Sogbein, OF .
ANALYTICAL CHEMISTRY, 2001, 73 (20) :4836-4844
[8]   On-column electrochemical reactions accompanying the electrospray process [J].
Liu, SY ;
Griffiths, WJ ;
Sjövall, J .
ANALYTICAL CHEMISTRY, 2003, 75 (04) :1022-1030
[9]   BACKGROUND-NOISE IN CAPILLARY ELECTROPHORETIC AMPEROMETRIC DETECTION [J].
LU, WZ ;
CASSIDY, RM .
ANALYTICAL CHEMISTRY, 1994, 66 (02) :200-204
[10]   Evaluations of the stability of sheathless electrospray ionization mass spectrometry emitters using electrochemical techniques [J].
Nilsson, S ;
Svedberg, M ;
Pettersson, J ;
Björefors, F ;
Markides, K ;
Nyholm, L .
ANALYTICAL CHEMISTRY, 2001, 73 (19) :4607-4616