Development of a surface acoustic wave immunosensor

被引:61
作者
Welsch, W [1 ]
Klein, C [1 ]
vonSchickfus, M [1 ]
Hunklinger, S [1 ]
机构
[1] UNIV HEIDELBERG, INST ANGEW PHYS, D-69120 HEIDELBERG, GERMANY
关键词
D O I
10.1021/ac960198z
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
In this paper the first application of horizontal polarized surface acoustic waves (HP-SAW) in LiTaO3 for an immunosensor is presented, A SAW device with two paths, a sensitive path and a reference path operating in liquid at 345 MHz, is used as the sensor, Antigens and antibodies were both immunoglobulins. The antibodies were immobilized via protein A on the sensitive path with a mass density of 0.4 ng/mm(2), as shown by ELISA measurements, A theoretical detection limit of similar to 33 pg and a high sensitivity of 110 kHz/(ng/mm(2)) could be derived from the experiments, High signal enhancement factors of 4.5 and 12 are achievable by gold colloid labeled antigens with colloid diameters of 5 and 10 nm, respectively. Specificity of the antibody was demonstrated by a preabsorption experiment. As a further antibody/antigen system the binding of the anti-human serum albumin (HSA) to HSA was examined.
引用
收藏
页码:2000 / 2004
页数:5
相关论文
共 19 条
[11]   PIEZOELECTRIC CRYSTAL BIOSENSOR MODIFIED WITH PROTEIN-A FOR DETERMINATION OF IMMUNOGLOBULINS [J].
MURAMATSU, H ;
DICKS, JM ;
TAMIYA, E ;
KARUBE, I .
ANALYTICAL CHEMISTRY, 1987, 59 (23) :2760-2763
[12]   DETECTION OF HUMAN TRANSFERRIN BY THE PIEZOELECTRIC CRYSTAL [J].
PRUSAKSOCHACZEWSKI, E ;
LUONG, JHT .
ANALYTICAL LETTERS, 1990, 23 (02) :183-194
[13]   A NEW APPROACH TO THE DEVELOPMENT OF A REUSABLE PIEZOELECTRIC CRYSTAL BIOSENSOR [J].
PRUSAKSOCHACZEWSKI, E ;
LUONG, JHT .
ANALYTICAL LETTERS, 1990, 23 (03) :401-409
[14]   MICROGRAVIMETRIC IMMUNOASSAY WITH PIEZOELECTRIC-CRYSTALS [J].
ROEDERER, JE ;
BASTIAANS, GJ .
ANALYTICAL CHEMISTRY, 1983, 55 (14) :2333-2336
[15]   SELECTIVE AND REPEATABLE DETECTION OF HUMAN SERUM-ALBUMIN BY USING PIEZOELECTRIC IMMUNOSENSOR [J].
SAKAI, G ;
SAIKI, T ;
UDA, T ;
MIURA, N ;
YAMAZOE, N .
SENSORS AND ACTUATORS B-CHEMICAL, 1995, 24 (1-3) :134-137
[16]  
SCHRAMM W, 1987, CLIN CHEM, V33, P1338
[17]   DESIGN OF SAW SENSOR IN LIQUID [J].
SHIOKAWA, S ;
MORIIZUMI, T .
JAPANESE JOURNAL OF APPLIED PHYSICS PART 1-REGULAR PAPERS SHORT NOTES & REVIEW PAPERS, 1988, 27 :142-144
[18]   DETERMINATION OF IMMUNOGLOBULIN-M CONCENTRATION BY PIEZOELECTRIC CRYSTAL IMMUNOBIOSENSOR COATED WITH PROTAMINE [J].
SURI, CR ;
RAJE, M ;
MISHRA, GC .
BIOSENSORS & BIOELECTRONICS, 1994, 9 (07) :535-542
[19]   ATRAZINE MEASUREMENTS USING SURFACE TRANSVERSE-WAVE DEVICES [J].
TOMMOY, M ;
BAER, RL ;
SPIRASOLOMON, D ;
DOHERTY, TP .