An investigation of the operating characteristics of two PTW diamond detectors in photon and electron beams

被引:64
作者
De Angelis, C
Onori, S
Pacilio, M
Cirrone, GAP
Cuttone, G
Raffaele, L
Bucciolini, M
Mazzocchi, S
机构
[1] Ist Super Sanita, Phys Lab, I-00161 Rome, Italy
[2] Ist Nazl Fis Nucl, Rome, Italy
[3] Ist Nazl Fis Nucl, Lab Nazl Sud, I-95129 Catania, Italy
[4] Univ Catania, Inst Radiol, Catania, Italy
[5] Univ Florence, Dipartimento Fisiopatol Clin, Florence, Italy
[6] Ist Nazl Fis Nucl, I-50125 Florence, Italy
关键词
diamond detector; solid state dosimetry;
D O I
10.1118/1.1446101
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The dosimetric properties of two PTW Riga diamond detectors type 60003 were studied in high-energy photon and electron therapy beam. Properties under study were current-voltage characteristic, polarization effect, time stability of response, dose response, dose-rate dependence, temperature stability, and beam quality dependence of the sensitivity factor. Differences were shown between the two detectors for most of the previous properties. Also, the observed behavior was, to some extent, different from what was reported in the PTW technical specifications. The necessity to characterize each diamond detector individually was addressed. (C) 2002 American Association of Physicists in Medicine.
引用
收藏
页码:248 / 254
页数:7
相关论文
共 19 条
[1]  
[Anonymous], 49 ICRU
[2]   PTW-diamond detector: Dose rate and particle type dependence [J].
Fidanzio, A ;
Azario, L ;
Miceli, R ;
Russo, A ;
Piermattei, A .
MEDICAL PHYSICS, 2000, 27 (11) :2589-2593
[3]  
Fowler J.F., 1966, RAD DOSIMETRY
[4]   EVALUATION OF A PTW DIAMOND DETECTOR FOR ELECTRON-BEAM MEASUREMENTS [J].
HEYDARIAN, M ;
HOBAN, PW ;
BECKHAM, WA ;
BORCHARDT, IM ;
BEDDOE, AH .
PHYSICS IN MEDICINE AND BIOLOGY, 1993, 38 (08) :1035-1042
[5]  
Heydarian M, 1997, PHYS MEDICA, V13, P55
[6]   DOSE-RATE DEPENDENCE OF A PTW DIAMOND DETECTOR IN THE DOSIMETRY OF A 6 MV PHOTON-BEAM [J].
HOBAN, PW ;
HEYDARIAN, M ;
BECKHAM, WA ;
BEDDOE, AH .
PHYSICS IN MEDICINE AND BIOLOGY, 1994, 39 (08) :1219-1229
[7]  
*IAEA, 1987, 277 IAEA
[8]  
*IAEA, 1997, 381 IAEA
[9]  
JOHNS HE, 1983, PHYSICS RADIOLOGY
[10]  
KHRUNOV VS, 1990, RADIAT PROT DOSIM, V33, P155