PTW-diamond detector: Dose rate and particle type dependence

被引:35
作者
Fidanzio, A [1 ]
Azario, L [1 ]
Miceli, R [1 ]
Russo, A [1 ]
Piermattei, A [1 ]
机构
[1] Univ Cattolica Sacro Cuore, Inst Fis, Rome, Italy
关键词
diamond detector; dose rate dependence; dosimetry;
D O I
10.1118/1.1318218
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
In this paper the suitability of a PTW natural diamond detector (DD) for relative and reference dosimetry of photon and electron beams, with dose per pulse between 0.068 mGy and 0.472 mGy, was studied and the results were compared with those obtained by a stereotactic silicon detector (SFD). The results show that, in the range of the examined dose per pulse the DD sensitivity changes up to 1.8% while the SFD sensitivity changes up to 4.5%. The fitting parameter, a, used to correct the dose per pulse dependence of solid state detectors, was Delta = 0.993+/-0.002 and Delta = 1.025+/-0.002 for the diamond detector and for the silicon diode, respectively. The a values were found to be independent of particle type of two conventional beams (a 10 MV x-ray beam and a 21 MeV electron beam). So if Delta is determined for a radiotherapy beam; it can be used to correct relative dosimetry for other conventional radiotherapy beams. Moreover the diamond detector shows a calibration factor which is independent of beam quality and particle type, so an empirical dosimetric formalism is proposed here to obtain the reference dosimetry. This formalism is based on a dose-to-water calibration factor and on an empirical coefficient, that takes into account the reading dependence on the dose per pulse. (C) 2000 American Association of Physicists in Medicine. [S0094-2405(00)00711-2].
引用
收藏
页码:2589 / 2593
页数:5
相关论文
共 12 条
[1]  
*ASS IT FIS BIOM, 1988, 6 AIFB, P2
[2]  
Fowler J.F., 1966, RAD DOSIMETRY
[3]  
Heydarian M, 1997, PHYS MEDICA, V13, P55
[4]   DOSE-RATE DEPENDENCE OF A PTW DIAMOND DETECTOR IN THE DOSIMETRY OF A 6 MV PHOTON-BEAM [J].
HOBAN, PW ;
HEYDARIAN, M ;
BECKHAM, WA ;
BEDDOE, AH .
PHYSICS IN MEDICINE AND BIOLOGY, 1994, 39 (08) :1219-1229
[5]  
*INT AT EN AG, 1987, TECHN REP SER INT AT, V277
[6]   Energy and dose rate dependence of a diamond detector in the dosimetry of 4-25 MV photon beams [J].
Laub, WU ;
Kaulich, TW ;
Nusslin, F .
MEDICAL PHYSICS, 1997, 24 (04) :535-536
[7]   An EGS4 Monte Carlo examination of the response of a PTW-diamond radiation detector in megavoltage electron beams [J].
Mobit, PN ;
Sandison, GA .
MEDICAL PHYSICS, 1999, 26 (05) :839-844
[8]  
Piermattei A, 1998, PHYS MEDICA, V14, P9
[9]   EVALUATION OF DIAMOND RADIATION DOSIMETERS [J].
PLANSKOY, B .
PHYSICS IN MEDICINE AND BIOLOGY, 1980, 25 (03) :519-532
[10]  
Rikner G, 1983, SILICON DIODES DETEC