Low-cost microelectrode array with integrated heater for extracellular recording of cardiomyocyte cultures using commercial flexible printed circuit technology

被引:31
作者
Giovangrandi, L [1 ]
Gilchrist, KH [1 ]
Whittington, RH [1 ]
Kovacs, GTA [1 ]
机构
[1] Stanford Univ, Dept Elect Engn, Stanford, CA 94305 USA
关键词
microelectrode array; flexible printed-circuit board; temperature control; cardiac myocytes; HL-1;
D O I
10.1016/j.snb.2005.03.059
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
This article reports the use of commercial, flexible printed circuit technology for the fabrication of low-cost microelectrode arrays (MEAs) for recording extracellular electrical signals from cardiomyocyte cultures. A 36-electrode array has been designed and manufactured using standard, two-layer, polyimide-based flexible circuit technology, with electrode diameters of 75 and 100 mu m. Copper structures defined on the backside of the array have been used for low-power thermal regulation of the culture. Electrical characterization of the gold-plated electrodes showed impedances below 250 k Omega at 1 kHz. Functional testing was conducted using HL-1 cardiac myocytes. The arrays proved biocompatible, and supported the formation of functional syncytia, as demonstrated by electrical recordings of depolarization waves across the array. A comparison with conventional, glass-based MEAs is presented, which reveals differences in signal strength (smaller for larger electrode) and variability (less for larger electrodes), but no effect of the substrate types on culture parameters such as beat rate or conduction velocity. The performance of the on-chip heating was evaluated, with typical temperature settling times (to 0.1 degrees C) below 10 s, for a power consumption around 1 W (at 37 degrees C). Accuracy and stability are discussed. HL-1 cell responses to various temperature profiles enabled by the on-chip heating are presented, showing a remarkable correlation between temperature and beat rate. (c) 2005 Elsevier B.V. All rights reserved.
引用
收藏
页码:545 / 554
页数:10
相关论文
共 21 条
[1]   Estimation of conduction velocity vector fields from epicardial mapping data [J].
Bayly, PV ;
KenKnight, BH ;
Rogers, JM ;
Hillsley, RE ;
Ideker, RE ;
Smith, WM .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1998, 45 (05) :563-571
[2]   A FLEXIBLE PERFORATED MICROELECTRODE ARRAY FOR EXTENDED NEURAL RECORDINGS [J].
BOPPART, SA ;
WHEELER, BC ;
WALLACE, CS .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1992, 39 (01) :37-42
[3]   Microelectrode arrays for stimulation of neural slice preparations [J].
Borkholder, DA ;
Bao, J ;
Maluf, NI ;
Perl, ER ;
Kovacs, GTA .
JOURNAL OF NEUROSCIENCE METHODS, 1997, 77 (01) :61-66
[4]  
BORKHOLDER DA, 1996, SOLID STATE SENSOR A
[5]   HL-1 cells: A cardiac muscle cell line that contracts and retains phenotypic characteristics of the adult cardiomyocyte [J].
Claycomb, WC ;
Lanson, NA ;
Stallworth, BS ;
Egeland, DB ;
Delcarpio, JB ;
Bahinski, A ;
Izzo, NJ .
PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 1998, 95 (06) :2979-2984
[6]   Portable cell-based biosensor system using integrated CMOS cell-cartridges [J].
DeBusschere, BD ;
Kovacs, GTA .
BIOSENSORS & BIOELECTRONICS, 2001, 16 (7-8) :543-556
[7]   A simulation study evaluating the performance of high-density electrode arrays on myocardial tissue [J].
Eason, JC ;
Eason, C ;
Malkin, RA .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2000, 47 (07) :893-901
[8]   Copper toxicity, oxidative stress, and antioxidant nutrients [J].
Gaetke, LM ;
Chow, CK .
TOXICOLOGY, 2003, 189 (1-2) :147-163
[9]   Sensitivity of cell-based biosensors to environmental variables [J].
Gilchrist, KH ;
Giovangrandi, L ;
Whittington, RH ;
Kovacs, GTA .
BIOSENSORS & BIOELECTRONICS, 2005, 20 (07) :1397-1406
[10]   General purpose, field-portable cell-based biosensor platform [J].
Gilchrist, KH ;
Barker, VN ;
Fletcher, LE ;
DeBusschere, BD ;
Ghanouni, P ;
Giovangrandi, L ;
Kovacs, GTA .
BIOSENSORS & BIOELECTRONICS, 2001, 16 (7-8) :557-564