Real-time-MR guidance for placement of a self-made fully MR-compatible atrial septal occluder:: in vitro test

被引:12
作者
Buecker, A
Spuentrup, E
Grabitz, R
Freudenthal, F
Schaeffter, T
van Vaals, JJ
Günther, W
机构
[1] Univ Technol Aachen, Clin Diagnost Radiol, D-52057 Aachen, Germany
[2] Univ Kiel, Dept Pediat Cardiol, D-24098 Kiel, Germany
[3] Philips Res Labs, Hamburg, Germany
[4] Philips Med Syst, Best, Netherlands
来源
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN | 2002年 / 174卷 / 03期
关键词
magnetic resonance imaging; real-time MR imaging; interventional radiology; atrial septal defect;
D O I
10.1055/s-2002-20601
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: This in vitro study investigated the feasibility to visualize the placement of three different atrial septal occluder systems using real-time MR control. Methods: The experiments were performed on an interventional 1.5T high field whole body system. Real-time MR imaging was achieved by radial or spiral k-space filling in conjunction with the sliding window reconstruction technique yielding an imaging speed of 15 frames per second. The CardioSeal, Amplatzer Septal Occluder and a specially designed MR-compatible closure device were tested in a water bath. A punctured plastic wall served as model for the atrial septal defect. Results: The delivery systems of the CardioSeal and Amplatzer Occluder were ferromagnetic and caused substantial artifacts, making the device placement impossible, even if the magnetic forces would have been acceptable. The self-made prototype caused only minor susceptibility artifacts allowing its visualization on the MR images. The MR imaging techniques applied enabled real-time control of the occluder including steering through the artificial septal foramen and visualization of the occluder deployment. Conclusion: Real-time MR imaging allows for guidance and placement of an MR-compatible septal occluder in vitro suggesting the feasibility to perform atrial septal occlusion under MR-guidance in vivo as well.
引用
收藏
页码:283 / 285
页数:3
相关论文
共 14 条
[1]   Visualization of dedicated catheters using fast scanning techniques with potential for MR-guided vascular interventions [J].
Bakker, CJ ;
Hoogeveen, RM ;
Weber, J ;
vanVaals, JJ ;
Viergever, MA ;
Mali, WP .
MAGNETIC RESONANCE IN MEDICINE, 1996, 36 (06) :816-820
[2]   Real-time MRI with radial scanning technique for control of angiographic interventions. [J].
Bucker, A ;
Adam, G ;
Neuerburg, JM ;
Glowinski, A ;
Nolte-Ernsting, CCA ;
Hunter, D ;
Grosskortenhaus, S ;
Leenen, M ;
Rasche, V ;
Proksa, R ;
van Vaals, JJ ;
Hurtak, W ;
Gunther, RW .
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 1998, 169 (05) :542-546
[3]   Real-time MR guidance for inferior vena cava filter placement in an animal model [J].
Bücker, A ;
Neuerburg, JM ;
Adam, GB ;
Glowinski, A ;
Schaeffter, T ;
Rasche, V ;
van Vaals, JJ ;
Günther, RW .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2001, 12 (06) :753-756
[4]  
EGGERS H, 1999, P 7 ANN M ISMRM PHIL, P95
[5]  
Konings MK, 2000, J MAGN RESON IMAGING, V12, P79, DOI 10.1002/1522-2586(200007)12:1<79::AID-JMRI9>3.3.CO
[6]  
2-K
[7]  
Ladd ME, 2000, MAGN RESON MED, V43, P615, DOI 10.1002/(SICI)1522-2594(200004)43:4<615::AID-MRM19>3.0.CO
[8]  
2-B
[9]   Instrument visualization in a magnetic resonance imaging environment [J].
Ladd, ME ;
Debatin, JF .
SEMINARS IN INTERVENTIONAL RADIOLOGY, 1999, 16 (01) :13-21
[10]   Visualization and temporal/spatial characterization of cardiac radiofrequency ablation lesions using magnetic resonance imaging [J].
Lardo, AC ;
McVeigh, ER ;
Jumrussirikul, P ;
Berger, RD ;
Calkins, H ;
Lima, J ;
Halperin, HR .
CIRCULATION, 2000, 102 (06) :698-705