Real-time MR guidance for inferior vena cava filter placement in an animal model

被引:36
作者
Bücker, A
Neuerburg, JM
Adam, GB
Glowinski, A
Schaeffter, T
Rasche, V
van Vaals, JJ
Günther, RW
机构
[1] Univ Technol Aachen, Clin Diagnost Radiol, Aachen, Germany
[2] Philips Res Labs, Hamburg, Germany
[3] Philips Med Syst, Best, Netherlands
关键词
magnetic resonance (MR); guidance; venae cavae; filters;
D O I
10.1016/S1051-0443(07)61448-1
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
It was the aim of this study to examine the feasibility of real-time magnetic resonance (MR) imaging for MR-guided placement of inferior vena cava (IVC) filters, which were placed in five pigs via a femoral approach. The introducer sheath and dilator were marked with Dysprosium rings. The procedures were performed under MR guidance with use of a 1.5-T ACS-NT imager. Radial filling of k-space in conjunction with the sliding window reconstruction technique achieved real-time MR imaging with a frame rate of 20 images/sec. Simultaneous real-time visualization of the vascular anatomy and interventional instruments was achieved under real-time conditions and allowed correct placement of IVC filters in all five cases as confirmed by radiographic angiography.
引用
收藏
页码:753 / 756
页数:4
相关论文
共 18 条
[1]  
Bartels LW, 2000, J MAGN RESON IMAGING, V12, P599, DOI 10.1002/1522-2586(200010)12:4<599::AID-JMRI13>3.3.CO
[2]  
2-Z
[3]   Stent placement under real-time MR control:: First experience in an animal model. [J].
Bücker, A ;
Neuerburg, JM ;
Adam, G ;
Schürmann, K ;
Rasche, V ;
van Vaals, JJ ;
Molgaard-Nielsen, A ;
Günther, RW .
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 1998, 169 (06) :655-657
[4]   Real-time MRI with radial scanning technique for control of angiographic interventions. [J].
Bucker, A ;
Adam, G ;
Neuerburg, JM ;
Glowinski, A ;
Nolte-Ernsting, CCA ;
Hunter, D ;
Grosskortenhaus, S ;
Leenen, M ;
Rasche, V ;
Proksa, R ;
van Vaals, JJ ;
Hurtak, W ;
Gunther, RW .
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 1998, 169 (05) :542-546
[5]   MR-guided placement of a temporary vena cava filter: Technique and feasibility [J].
Frahm, C ;
Gehl, HB ;
Lorch, H ;
Zwaan, M ;
Drobnitzky, M ;
Laub, GA ;
Weiss, HD .
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1998, 8 (01) :105-109
[6]   PROJECTION RECONSTRUCTION TECHNIQUES FOR REDUCTION OF MOTION EFFECTS IN MRI [J].
GLOVER, GH ;
PAULY, JM .
MAGNETIC RESONANCE IN MEDICINE, 1992, 28 (02) :275-289
[7]   MR-guided transjugular portosystemic shunt placement in a swine model [J].
Kee, ST ;
Rhee, JS ;
Butts, K ;
Daniel, B ;
Pauly, J ;
Kerr, A ;
O'Sullivan, GJ ;
Sze, DY ;
Razavi, MK ;
Semba, CP ;
Herfkens, RJ ;
Dake, MD .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 1999, 10 (05) :529-535
[8]   Current practice of temporary vena cava filter insertion:: A multicenter registry [J].
Lorch, H ;
Welger, D ;
Wagner, V ;
Hillner, B ;
Strecker, EP ;
Herrmann, H ;
Voshage, G ;
Zur, C ;
Schwarzbach, C ;
Schröder, J ;
Gullotta, U ;
Pleissner, J ;
Huttner, S ;
Siering, U ;
Märcklin, C ;
Chavan, A ;
Gläser, F ;
Apitzsch, DE ;
Moubayed, K ;
Leonhardi, J ;
Schuchard, UM ;
Weiss, HD ;
Zwaan, M .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2000, 11 (01) :83-88
[9]  
MACK C, 2000, ROFO FORTSCHR RONTG, V172, P92
[10]   Gunther tulip filter: Preliminary clinical experience with retrieval [J].
Millward, SF ;
Bhargava, A ;
Aquino, J ;
Peterson, RA ;
Veinot, JP ;
Bormanis, J ;
Wells, PS .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2000, 11 (01) :75-82