Fat mass measured by DXA varies with scan velocity

被引:21
作者
Black, E
Petersen, L
Kreutzer, M
Toubro, S
Sorensen, TIA
Pedersen, O
Astrup, A
机构
[1] Royal Vet & Agr Univ, Res Dept Human Nutr, Ctr Adv Food Studies, DK-1958 Frederiksberg C, Denmark
[2] Steno Diabet Ctr, DK-2820 Gentofte, Denmark
[3] Copenhagen Univ Hosp, Inst Prevent Med, Danish Epidemiol Sci Ctr, Copenhagen, Denmark
来源
OBESITY RESEARCH | 2002年 / 10卷 / 02期
关键词
body composition; DXA; human; scan velocity;
D O I
10.1038/oby.2002.12
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective: To study the influence of scan velocities of DXA on the measured size of fat mass, lean body mass, bone mineral content and density, and total body weight. Research Methods and Procedures: The subjects were 71 healthy white adults, 38 women and 33 men. The mean age was 41.7 +/- 13.5 years and body mass index was 28.6 +/- 5.6 kg/m(2). The subjects were scanned consecutively in slow, medium, and fast scan mode by a Lunar DPX-IQ DXA scanner. Results: Throughout the body mass index and sagittal height ranges, scanned lean body mass significantly decreased with higher scan velocity and lean body mass was 2.7% lower in fast than in medium mode (p < 0.0001). In contrast, fat mass, percentage of body fat, and bone mineral contents were higher with increasing scan velocity. Areas not analyzed by the scanner, so called blue spots, increased with scan velocity and sagittal height, and their presence significantly enhanced the error. Body weight estimated by DXA in slow mode was -0.8% lower than scale weight in the women (p < 0.001) and -0.2% in men (not significant), and the difference was greater with increasing scan velocity. Discussion: Scan velocity significantly influences the measured fat mass size, lean body mass, bone mineral content, and body weight. To obtain the most accurate results, slow mode is preferable and fast scans should be avoided. Future studies should report and take scan velocity into consideration.
引用
收藏
页码:69 / 77
页数:9
相关论文
共 15 条
[1]   Body composition by DEXA in older adults: Accuracy and influence of scan mode [J].
Clasey, JL ;
Hartman, ML ;
Kanaley, J ;
Wideman, L ;
Teates, CD ;
Bouchard, C ;
Weltman, A .
MEDICINE AND SCIENCE IN SPORTS AND EXERCISE, 1997, 29 (04) :560-567
[2]   Untitled [J].
Hanson, J ;
Felsenberg, D ;
Fuerst, T ;
Lu, Y ;
Genant, HK ;
Hangartner, T ;
Harrold, L ;
Johnston, CC ;
Kalender, WA ;
Mazess, R ;
Nord, R ;
Miller, PD ;
Morita, R ;
Nakamura, A ;
Tamegai, T ;
Reeve, J ;
Schiessl, H ;
TysarczykNiemeyer, G ;
Steiger, P ;
Stein, J ;
vonStetten, E .
BONE, 1997, 21 (04) :369-370
[3]   Universal standardization of bone density measurements: A method with optimal properties for calibration among several instruments [J].
Hui, SL ;
Gao, SJ ;
Zhou, XH ;
Johnston, CC ;
Lu, Y ;
Gluer, CC ;
Grampp, S ;
Genant, H .
JOURNAL OF BONE AND MINERAL RESEARCH, 1997, 12 (09) :1463-1470
[4]   Body composition analysis by dual energy X-ray absorptiometry in female diabetics differ between manufacturers [J].
Kistorp, CN ;
Svendsen, OL .
EUROPEAN JOURNAL OF CLINICAL NUTRITION, 1997, 51 (07) :449-454
[5]   Accuracy and precision of 62 bone densitometers using a European Spine Phantom [J].
Kolta, S ;
Ravaud, P ;
Fechtenbaum, J ;
Dougados, M ;
Roux, C .
OSTEOPOROSIS INTERNATIONAL, 1999, 10 (01) :14-19
[6]   Differences in whole body measurements by DXA-scanning using two Lunar DPX-L machines [J].
Lantz, H ;
Samuelson, G ;
Bratteby, LE ;
Mallmin, H ;
Sjöström, L .
INTERNATIONAL JOURNAL OF OBESITY, 1999, 23 (07) :764-770
[7]   Dual-energy X-ray absorptiometry and body composition [J].
Laskey, MA .
NUTRITION, 1996, 12 (01) :45-51
[8]   Comparison of adult and paediatric spine and whole body software for the Lunar dual energy X-ray absorptiometer [J].
Laskey, MA ;
Prentice, A .
BRITISH JOURNAL OF RADIOLOGY, 1999, 72 (862) :967-976
[9]  
*LUN CORP, 1997, DPX IQ REF MAN DOC V, pD6
[10]  
*LUN CORP, 1997, DPX IQ REF MAN DOC V, P7