A survey of current in vivo radiotherapy dosimetry practice

被引:13
作者
Edwards, CR
Grieveson, MH
Mountford, PJ
Rolfe, P
机构
[1] Dept. of Biomed. Eng. and Med. Phys., North Staffordshire Hospital, Staffordshire ST4 7LN, Princes Road
关键词
D O I
10.1259/bjr.70.831.9166056
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A questionnaire was sent out to 57 radiotherapy physics departments in the United Kingdom to determine the type of dosemeters used for in vivo measurements inside and outside X-ray treatment fields, and whether any correction is made for energy dependence when the dose to critical organs outside the main beam is estimated. 44 responses were received. 11 centres used a semi-conductor for central axis dosimetry compared with only two centres which used thermoluminescent dosimetry (TLD). 37 centres carried out dosimetry measurements outside the main beam; 25 centres used TLD and 12 centres used a semi-conductor detector. Of the 16 centres measuring the dose at both sites, 11 used a semi-conductor for the central axis measurement, but only four of those 11 changed to TLD for critical organ dosimetry despite the latter's lower variation in energy response. None of the centres stated that they made a correction for the variation in detector energy response when making measurements outside the main beam, indicating a need for a more detailed evaluation of the energy response of these detectors and the energy spectra outside the main beam.
引用
收藏
页码:299 / 302
页数:4
相关论文
共 11 条
[1]   A COMPARISON OF SEMICONDUCTOR AND THERMOLUMINESCENT DOSIMETERS FOR INVIVO DOSIMETRY [J].
AUKETT, RJ .
BRITISH JOURNAL OF RADIOLOGY, 1991, 64 (766) :947-952
[2]  
CAMERON J, 1991, ENVIRON HEALTH PERSP, V91, P45
[3]   EVALUATION OF TEMPERATURE EFFECTS IN P-TYPE SILICON DETECTORS [J].
GRUSELL, E ;
RIKNER, G .
PHYSICS IN MEDICINE AND BIOLOGY, 1986, 31 (05) :527-534
[5]   CLINICAL THERMOLUMINESCENCE DOSIMETRY - HOW DO EXPECTATIONS AND RESULTS COMPARE [J].
KRON, T ;
SCHNEIDER, M ;
MURRAY, A ;
MAMEGHAN, H .
RADIOTHERAPY AND ONCOLOGY, 1993, 26 (02) :151-161
[6]   QUALITY ASSURANCE IN RADIOTHERAPY BY INVIVO DOSIMETRY .1. ENTRANCE DOSE MEASUREMENTS, A RELIABLE PROCEDURE [J].
LEUNENS, G ;
VANDAM, J ;
DUTREIX, A ;
VANDERSCHUEREN, E .
RADIOTHERAPY AND ONCOLOGY, 1990, 17 (02) :141-151
[7]   QUALITY ASSURANCE IN RADIOTHERAPY BY INVIVO DOSIMETRY .2. DETERMINATION OF THE TARGET ABSORBED DOSE [J].
LEUNENS, G ;
VANDAM, J ;
DUTREIX, A ;
VANDERSCHUEREN, E .
RADIOTHERAPY AND ONCOLOGY, 1990, 19 (01) :73-87
[8]  
MARSHALL TO, 1984, PRACTICAL ASPECTS TH, P19
[9]  
MAYLES WPM, 1993, RADIOTHERAPY PHYSICS, P227
[10]   EFFECTS OF RADIATION-DAMAGE ON P-TYPE SILICON DETECTORS [J].
RIKNER, G ;
GRUSELL, E .
PHYSICS IN MEDICINE AND BIOLOGY, 1983, 28 (11) :1261-1267