摘要
This study evaluated reticulocyte counting with the automated reticulocyte function of the Coulter STKS Haematology Analyser. This is an upgrade option for Coulter STKS and MAXIM haematology analysers. Reticulocyte counts obtained with the automated reticulocyte counting function were compared with those obtained by visual counting. Reticulocyte counting with both methods gave excellent comparability with a correlation coefficient of 0.98. Results were consistent with the well documented imprecision of the manual method with a coefficient of variation (CV) of 16-22%. In contrast, the automated reticulocyte counting function was more precise with a CV of 12.3%. In both cases, counts were stable after storage for 24 h at room temperature and 4 degrees C. Our results suggest that the use of this upgrade will be beneficial for many laboratories.
引用
收藏
页码:89 / 92
页数:4
相关论文
共 16 条
[1]   COMPARISON OF A MODIFIED THIAZOLE ORANGE TECHNIQUE WITH A FULLY AUTOMATED ANALYZER FOR RETICULOCYTE COUNTING [J].
BOWEN, D ;
BENTLEY, N ;
HOY, T ;
CAVILL, I .
JOURNAL OF CLINICAL PATHOLOGY, 1991, 44 (02) :130-133
[2]   AUTOMATED RETICULOCYTE COUNTING AND MEASUREMENT OF RETICULOCYTE CELLULAR INDEXES - EVALUATION OF THE MILES-H-ASTERISK-3-BLOOD-ANALYZER [J].
BRUGNARA, C ;
HIPP, MJ ;
IRVING, PJ ;
LATHROP, H ;
LEE, PA ;
MINCHELLO, EM ;
WINKELMAN, J .
AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1994, 102 (05) :623-632
[3]   COUNTING RETICULOCYTES BY FLOW-CYTOMETRY - USE OF THIAZOLE ORANGE [J].
CARTER, JM ;
MCSWEENEY, PA ;
WAKEM, PJ ;
NEMET, AM .
CLINICAL AND LABORATORY HAEMATOLOGY, 1989, 11 (03) :267-271
[4]   ENUMERATION OF RETICULOCYTES USING FLUORESCENCE-ACTIVATED FLOW-CYTOMETRY [J].
CORASH, L ;
RHEINSCHMIDT, M ;
LIEU, S ;
MEERS, P ;
BREW, E .
PATHOLOGY AND IMMUNOPATHOLOGY RESEARCH, 1988, 7 (05) :381-394
[5]  
COULTER, EVALUATION COULTER R, P1
[6]   FLOW CYTOMETRIC RETICULOCYTE ANALYSIS - MULTIINSTITUTIONAL INTERLABORATORY CORRELATION STUDY [J].
DAVIS, BH ;
BIGELOW, NC ;
KOEPKE, JA ;
BOROWITZ, MJ ;
HOUWEN, B ;
JACOBBERGER, JW ;
PIERRE, RV ;
CORASH, L ;
AULT, KA ;
BATJER, JD .
AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1994, 102 (04) :468-477
[7]   EVALUATION OF RETICULOCYTE COUNTS BY FLOW-CYTOMETRY IN A ROUTINE LABORATORY [J].
FERGUSON, DJ ;
LEE, SF ;
GORDON, PA .
AMERICAN JOURNAL OF HEMATOLOGY, 1990, 33 (01) :13-17
[8]  
HALL R, 1991, MED LAB HAEMATOLOGY
[9]  
HAMILTON JL, 1995, CLIN LAB HAEMATOL, V17, P145
[10]   EVALUATION OF AUTOMATED RETICULOCYTE COUNTS AND THEIR RELIABILITY IN THE PRESENCE OF HOWELL-JOLLY BODIES [J].
LOFSNESS, KG ;
KOHNKE, ML ;
GEIER, NA .
AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1994, 101 (01) :85-90