Beam calibration without a phantom for creating a 3-D freehand ultrasound system

被引:73
作者
Muratore, DM
Galloway, RL
机构
[1] Vanderbilt Univ, Dept Biomed Engn, Nashville, TN 37235 USA
[2] Vanderbilt Univ, Dept Neurol Surg, Nashville, TN 37235 USA
关键词
three-dimensional ultrasound; beam calibration; registration; freehand scanning; image-guided surgery;
D O I
10.1016/S0301-5629(01)00469-0
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
To create a freehand three-dimensional (3-D) ultrasound (US) system for image-guided surgical procedures, an US beam calibration process must be performed. The calibration method presented in this work does not use a phantom to define in 3-D space the pixel locations in the beam. Rather, the described method is based on the spatial relationship between an optically tracked pointer and a similarly tracked US transducer. The pointer tip was placed into the US beam, and US images, physical coordinates of the pointer and the transducer location were simultaneously recorded. US image coordinates of the pointer were mapped to the physical points using two different registration methods. Two sensitivity studies were performed to determine the location and number of points needed to calibrate the beam accurately. Results showed that the beam is most efficiently calibrated with approximately 20 points collected from throughout the beam. This method of beam calibration proved to be highly accurate, yielding registration errors of approximately 0.4 mm. (C) 2001 World Federation for Ultrasound in Medicine & Biology.
引用
收藏
页码:1557 / 1566
页数:10
相关论文
共 10 条
[1]   Dynamic three-dimensional freehand echocardiography using raw digital ultrasound data [J].
Berg, S ;
Torp, H ;
Martens, D ;
Steen, E ;
Samstad, S ;
Hoivik, I ;
Olstad, B .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1999, 25 (05) :745-753
[2]  
Blackall JM, 2000, LECT NOTES COMPUT SC, V1935, P462
[3]  
DUERK JL, 1999, TECHNICAL REQUIREMEN, P33
[4]   Performance of a miniature magnetic position sensor for three-dimensional ultrasound imaging [J].
Leotta, DF ;
Detmer, PR ;
Martin, RW .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1997, 23 (04) :597-609
[5]   Registration of head volume images using implantable fiducial markers [J].
Maurer, CR ;
Fitzpatrick, JM ;
Wang, MY ;
Galloway, RL ;
Maciunas, RJ ;
Allen, GS .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1997, 16 (04) :447-462
[6]   Reconstruction and visualization of irregularly sampled three- and four-dimensional ultrasound data for cerebrovascular applications [J].
Meairs, S ;
Beyer, J ;
Hennerici, M .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2000, 26 (02) :263-272
[7]   Rapid calibration for 3-D freehand ultrasound [J].
Prager, RW ;
Rohling, RN ;
Gee, AH ;
Berman, L .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1998, 24 (06) :855-869
[8]   Automatic registration of 3-D ultrasound images [J].
Rohling, RN ;
Gee, AH ;
Berman, L .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1998, 24 (06) :841-854
[9]   Image guidance of breast cancer surgery using 3-D ultrasound images and augmented reality visualization [J].
Sato, Y ;
Nakamoto, M ;
Tamaki, Y ;
Sasama, T ;
Sakita, I ;
Nakajima, Y ;
Monden, M ;
Tamura, S .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1998, 17 (05) :681-693
[10]  
TENDICK F, 1999, TECHNICAL REQUIREMEN, P25