Performance of a miniature magnetic position sensor for three-dimensional ultrasound imaging

被引:83
作者
Leotta, DF
Detmer, PR
Martin, RW
机构
[1] UNIV WASHINGTON,DEPT SURG,SEATTLE,WA 98195
[2] UNIV WASHINGTON,DEPT ANESTHESIOL,SEATTLE,WA 98195
关键词
ultrasound; imaging; three-dimensional; magnetometer; position; orientation;
D O I
10.1016/S0301-5629(97)00043-4
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
A miniature magnetic position sensor used for three-dimensional ultrasound imaging was tested for precision and accuracy in vitro. The sensor alone was able to locate points with root-mean-square (rms) uncertainty of 1.7 mm and accuracy of 0.05 +/- 0.62 mm over its specified operating range of 50 cm, With an ultrasound imaging system, a point was located from arbitrary viewing windows with 2.4-mm rms uncertainty and 0.06 +/- 0.68 mm accuracy, If viewing windows were Limited to those representative of a typical ultrasound examination, the system could achieve rms uncertainty in point location of < I mm, Performance was not affected by operation of the imaging system when the sensor was mounted on an ultrasound scanhead. Sensitivity to metals in the operating environment was also measured, (C) 1997 World Federation for Ultrasound in Medicine & Biology.
引用
收藏
页码:597 / 609
页数:13
相关论文
共 20 条
[1]   QUANTIFICATION OF ATHEROSCLEROTIC PLAQUES IN CAROTID ARTERIES BY 3-DIMENSIONAL ULTRASOUND [J].
DELCKER, A ;
DIENER, HC .
BRITISH JOURNAL OF RADIOLOGY, 1994, 67 (799) :672-678
[2]   3D ULTRASONIC IMAGE FEATURE LOCALIZATION BASED ON MAGNETIC SCANHEAD TRACKING - IN-VITRO CALIBRATION AND VALIDATION [J].
DETMER, PR ;
BASHEIN, G ;
HODGES, T ;
BEACH, KW ;
FILER, EP ;
BURNS, DH ;
STRANDNESS, DE .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1994, 20 (09) :923-936
[3]   IN-VITRO EVALUATION OF 3-DIMENSIONAL ULTRASONOGRAPHY IN VOLUME ESTIMATION OF ABDOMINAL ORGANS [J].
GILJA, OH ;
THUNE, N ;
MATRE, K ;
HAUSKEN, T ;
ODEGAARD, S ;
BERSTAD, A .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1994, 20 (02) :157-165
[4]   IN-VIVO COMPARISON OF 3D ULTRASONOGRAPHY AND MAGNETIC-RESONANCE-IMAGING IN VOLUME ESTIMATION OF HUMAN KIDNEYS [J].
GILJA, OH ;
SMIEVOLL, AI ;
THUNE, N ;
MATRE, K ;
HAUSKEN, T ;
ODEGAARD, S ;
BERSTAD, A .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1995, 21 (01) :25-32
[5]   A NEW INTEGRATED SYSTEM FOR 3-DIMENSIONAL ECHOCARDIOGRAPHIC RECONSTRUCTION - DEVELOPMENT AND VALIDATION FOR VENTRICULAR VOLUME WITH APPLICATION IN HUMAN-SUBJECTS [J].
HANDSCHUMACHER, MD ;
LETHOR, JP ;
SIU, SC ;
MELE, D ;
RIVERA, JM ;
PICARD, MH ;
WEYMAN, AE ;
LEVINE, RA .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1993, 21 (03) :743-753
[6]   ULTRASONIC 3-DIMENSIONAL RECONSTRUCTION - IN-VITRO AND IN-VIVO VOLUME AND AREA MEASUREMENT [J].
HODGES, TC ;
DETMER, PR ;
BURNS, DH ;
BEACH, KW ;
STRANDNESS, DE .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1994, 20 (08) :719-729
[7]   QUANTITATIVE 3-DIMENSIONAL RECONSTRUCTION OF ANEURYSMAL LEFT-VENTRICLES - IN-VITRO AND IN-VIVO VALIDATION [J].
JIANG, L ;
DEPRADA, JAV ;
HANDSCHUMACHER, MD ;
VUILLE, C ;
GUERERRO, JL ;
PICARD, MH ;
JOZIATIS, JT ;
FALLON, JT ;
WEYMAN, AE ;
LEVINE, RA .
CIRCULATION, 1995, 91 (01) :222-230
[8]   3-DIMENSIONAL US OF THE FETUS - WORK-IN-PROGRESS [J].
KELLY, IMG ;
GARDENER, JE ;
BRETT, AD ;
RICHARDS, R ;
LEES, WR .
RADIOLOGY, 1994, 192 (01) :253-259
[9]   3-DIMENSIONAL SPATIAL REGISTRATION AND INTERACTIVE DISPLAY OF POSITION AND ORIENTATION OF REAL-TIME ULTRASOUND IMAGES [J].
KING, DL ;
KING, DL ;
SHAO, MYC .
JOURNAL OF ULTRASOUND IN MEDICINE, 1990, 9 (09) :525-532
[10]  
KING DL, 1991, J ULTRAS MED, V10, P77