A computer simulation study comparing lesion detection accuracy with digital mammography, breast tomosynthesis, and cone-beam CT breast imaging

被引:150
作者
Gong, X
Glick, SJ
Liu, B
Vedula, AA
Thacker, S
机构
[1] Rush Univ, Med Ctr, Chicago, IL 60612 USA
[2] Rush Univ, Med Ctr, Dept Radiat Oncol, Chicago, IL 60612 USA
[3] Univ Massachusetts, Sch Med, Dept Radiol, Worcester, MA 01655 USA
[4] Massachusetts Gen Hosp, Boston, MA 02114 USA
[5] Harvard Univ, Sch Med, Boston, MA 02114 USA
关键词
mammography; tomosynthysis; breast imaging; cone-beam CT; flat-panel imager; ROC observer study;
D O I
10.1118/1.2174127
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Although conventional mammography is currently the best modality to detect early breast cancer, it is limited in that the recorded image represents the superposition of a three-dimensional (3D) object onto a 2D plane. Recently, two promising approaches for 3D volumetric breast imaging have been proposed, breast tomosynthesis (BT) and CT breast imaging (CTBI). To investigate possible improvements in lesion detection accuracy with either breast tomosynthesis or CT breast imaging as compared to digital mammography (DM), a computer simulation study was conducted using simulated lesions embedded into a structured 3D breast model. The computer simulation realistically modeled x-ray transport through a breast model, as well as the signal and noise propagation through a CsI based flat-panel imager. Polyenergetic x-ray spectra of Mo/Mo 28 kVp for digital mammography, Mo/Rh 28 kVp for BT, and W/Ce 50 kVp for CTBI were modeled. For the CTBI simulation, the intensity of the x-ray spectra for each projection view was determined so as to provide a total average glandular dose of 4 mGy, which is approximately equivalent to that given in conventional two-view screening mammography. The same total dose was modeled for both the DM and BT simulations. Irregular lesions were simulated by using a stochastic growth algorithm providing lesions with an effective diameter of 5 mm. Breast tissue was simulated by generating an ensemble of backgrounds with a power law spectrum, with the composition of 50% fibroglandular and 50% adipose tissue. To evaluate lesion detection accuracy, a receiver operating characteristic (ROC) study was performed with five observers reading an ensemble of images for each case. The average area under the ROC curves (A) was 0.76 for DM, 0.93 for BT, and 0.94 for CTBI. Results indicated that for the same dose, a 5 mm lesion embedded in a structured breast phantom was detected by the two volumetric breast imaging systems, BT and CTBI, with statistically significant higher confidence than with planar digital mammography, while the difference in lesion detection between BT and CTBI was not statistically significant. (c) 2006 American Association of Physicists in Medicine.
引用
收藏
页码:1041 / 1052
页数:12
相关论文
共 56 条
[1]  
[Anonymous], P SPIE
[2]  
[Anonymous], 2003, BREAST IM REP DAT SY
[3]   Mammogram synthesis using a 3D simulation. 1. Breast tissue model and image acquisition simulation [J].
Bakic, PR ;
Albert, M ;
Brzakovic, D ;
Maidment, ADA .
MEDICAL PHYSICS, 2002, 29 (09) :2131-2139
[4]  
BARRETT HH, 2001, P SPIE, V4320
[5]   Statistical texture synthesis of mammographic images with clustered lumpy backgrounds [J].
Bochud, FO ;
Abbey, CK ;
Eckstein, MP .
OPTICS EXPRESS, 1999, 4 (01) :33-43
[6]   Glandular breast dose for monoenergetic and high-energy X-ray beams: Monte Carlo assessment [J].
Boone, JM .
RADIOLOGY, 1999, 213 (01) :23-37
[7]   A comprehensive analysis of DgNCT coefficients for pendant-geometry cone-beam breast computed tomography [J].
Boone, JM ;
Shah, N ;
Nelson, TR .
MEDICAL PHYSICS, 2004, 31 (02) :226-235
[8]   Molybdenum, rhodium, and tungsten anode spectral models using interpolating polynomials with application to mammography [J].
Boone, JM ;
Fewell, TR ;
Jennings, RJ .
MEDICAL PHYSICS, 1997, 24 (12) :1863-1874
[9]   Dedicated breast CT: Radiation dose and image quality evaluation [J].
Boone, JM ;
Nelson, TR ;
Lindfors, KK ;
Seibert, JA .
RADIOLOGY, 2001, 221 (03) :657-667
[10]   Accurate method for computer-generating tungsten anode x-ray spectra from 30 to 140 kV [J].
Boone, JM ;
Seibert, JA .
MEDICAL PHYSICS, 1997, 24 (11) :1661-1670