Glandular breast dose for monoenergetic and high-energy X-ray beams: Monte Carlo assessment

被引:245
作者
Boone, JM [1 ]
机构
[1] Univ Calif Davis, Med Ctr, Dept Radiol, Radiol Res Labs, Sacramento, CA 95817 USA
关键词
breast radiography; radiation dose; technology; utilization; physics;
D O I
10.1148/radiology.213.1.r99oc3923
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE: To extend the utility of normalized glandular dose (D-gN) calculations to higher x-ray energies (up to 120 keV) and to provide the tools for investigators to calculate D-gN values for arbitrary mammographic and x-ray spectra. MATERIALS AND METHODS: Validated Monte Carte methods were used to assess D-gN values. One million x-ray photons (1-120 keV, in 1-keV increments)were input to a semicircular breast geometry of thicknesses from 2 to 12 cm and breast compositions from 0% to 100% glandular. D-gN values for monoenergetic (1-120 keV) x-ray beams, polyenergetic (40-120 kV, tungsten anode) x-ray spectra, and polyenergetic mammographic spectra were computed. Skin thicknesses of 4-5 mm were used. RESULTS: The calculated D-gN Values were in agreement within approximately 1%-6% with previously published data, depending on breast composition. D-gN tables were constructed for a variety of x-ray tube anode-filter combinations, including molybdenum anode-molybdenum filter, molybdenum anode-rhodium filter, rhodium anode-rhodium filter, tungsten anode-rhodium filter; tungsten anode-palladium filter, and tungsten anode-silver filter. D-gN values also were graphed for monoenergetic beams to 120 keV and for general diagnostic x-ray beams to 120 kV. CONCLUSION: The tables and graphs may be useful for optimizing mammographic procedures. The higher energy data may be useful for investigations of the potential of dual-energy mammography or for calculation of dose in general diagnostic or computed tomographic procedures.
引用
收藏
页码:23 / 37
页数:15
相关论文
共 29 条
[1]   A linear array silicon pixel detector: images of a mammographic test object and evaluation of delivered doses [J].
Arfelli, F ;
Bonvicini, V ;
Bravin, A ;
Cantatore, G ;
Castelli, E ;
DallaPalma, L ;
DiMichiel, M ;
Longo, R ;
Olivo, A ;
Pani, S ;
Pontoni, D ;
Poropat, P ;
Prest, M ;
Rashevsky, A ;
Tromba, G ;
Vacchi, A .
PHYSICS IN MEDICINE AND BIOLOGY, 1997, 42 (08) :1565-1573
[2]   SCATTERED ENERGY DEPOSITION UNDER SHIELDING [J].
BOONE, JM ;
SEIBERT, JA .
INVESTIGATIVE RADIOLOGY, 1988, 23 (08) :627-631
[3]   MONTE-CARLO SIMULATION OF THE SCATTERED RADIATION DISTRIBUTION IN DIAGNOSTIC-RADIOLOGY [J].
BOONE, JM ;
SEIBERT, JA .
MEDICAL PHYSICS, 1988, 15 (05) :713-720
[4]   DUAL-ENERGY MAMMOGRAPHY - A DETECTOR ANALYSIS [J].
BOONE, JM ;
SHABER, GS ;
TECOTZKY, M .
MEDICAL PHYSICS, 1990, 17 (04) :665-675
[5]   Molybdenum, rhodium, and tungsten anode spectral models using interpolating polynomials with application to mammography [J].
Boone, JM ;
Fewell, TR ;
Jennings, RJ .
MEDICAL PHYSICS, 1997, 24 (12) :1863-1874
[6]   Accurate method for computer-generating tungsten anode x-ray spectra from 30 to 140 kV [J].
Boone, JM ;
Seibert, JA .
MEDICAL PHYSICS, 1997, 24 (11) :1661-1670
[7]   Comparison of x-ray cross sections for diagnostic and therapeutic medical physics [J].
Boone, JM ;
Chavez, AE .
MEDICAL PHYSICS, 1996, 23 (12) :1997-2005
[8]   MAMMOGRAPHY WITH SYNCHROTRON-RADIATION [J].
BURATTINI, E ;
COSSU, E ;
DIMAGGIO, C ;
GAMBACCINI, M ;
INDOVINA, PL ;
MARZIANI, M ;
POCEK, M ;
SIMEONI, S ;
SIMONETTI, G .
RADIOLOGY, 1995, 195 (01) :239-244
[9]   Niobium/molybdenum K-edge filtration in mammography: Contrast and dose evaluation [J].
Calicchia, A ;
Gambaccini, M ;
Indovina, PL ;
Mazzei, F ;
Pugliani, L .
PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (09) :1717-1726
[10]  
CULLEN DE, 1997, TART97 COUPLED NEUTR