Polysulfone as an alternative material to PVC in the design of ion-selective electrodes

被引:15
作者
Gonzalez-Bellavista, Anna
Macanas, Jorge
Munoz, Maria
Fabregas, Esteve
机构
[1] Autonomous Univ Barcelona, Dept Chem, Sensors & Biosensors Grp, Bellaterra 08193, Spain
[2] Autonomous Univ Barcelona, Dept Chem, Grp Separat Tech Chem, Bellaterra 08193, Spain
关键词
polysulfone membrane; ion-selective electrode; nitrate; COMPOSITE MEMBRANES; BIOCOMPATIBILITY; ADHESION; MATRICES; SOLVENT; SENSORS;
D O I
10.1016/j.snb.2005.10.040
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Polysulfone (PS). a non-conductive polymer, is employed as an alternative material to polyvinyl chloride (PVC), the most used polymer in the design of ion-selective electrode (ISE) membranes to date. A nitrate-ISE is chosen as a model electrode to study the efficiency of PS as a polymer matrix of the membrane. Tetra-octyl ammonium nitrate is used as ionophore and 2-nitrophenyl octyl ether as plasticizer. The polymer/plasticizer/ionophore ratio is optimized and the analytical performance parameters of the prepared electrodes are evaluated. The electrodes show a slope between 45 and 60 mV/decade (depending on the calibration medium and membrane composition). A linear range response between 1.0 x 10(-4) and 3.0 x 10(-2) M, and a lifetime of at least I month are obtained. The electrodes exhibit good selectivity for nitrate in respect to other anions. (c) 2005 Elsevier B.V. All rights reserved.
引用
收藏
页码:691 / 696
页数:6
相关论文
共 27 条
[1]  
[Anonymous], [No title captured]
[2]   Selective nitrate poly(vinylchloride) membrane electrode based on bis(2-hydroxyacetophenone)ethylenediimine vanadyl(IV) [J].
Mazloum-Ardakani, M ;
Salavati-Niasari, M ;
Jamshidpoor, M .
SENSORS AND ACTUATORS B-CHEMICAL, 2004, 101 (03) :302-307
[3]   Poly(vinyl chloride), polysulfone and sulfonated polyether-ether sulfone composite membranes for glucose and hydrogen peroxide perm-selectivity in amperometric biosensors [J].
Benmakroha, Y ;
Christie, I ;
Desai, M ;
Vadgama, P .
ANALYST, 1996, 121 (04) :521-526
[4]  
BRAQUERO E, 2001, THESIS U AUTONOMA BA, P152
[5]   RECOMMENDATIONS FOR NOMENCLATURE OF ION-SELECTIVE ELECTRODES - (IUPAC RECOMMENDATIONS 1994) [J].
BUCK, RP ;
LINDNER, E .
PURE AND APPLIED CHEMISTRY, 1994, 66 (12) :2527-2536
[6]   ELECTROCHEMICAL PERFORMANCE, BIOCOMPATIBILITY, AND ADHESION OF NEW POLYMER MATRICES FOR SOLID-STATE ION SENSORS [J].
CHA, GS ;
LIU, D ;
MEYERHOFF, ME ;
CANTOR, HC ;
MIDGLEY, AR ;
GOLDBERG, HD ;
BROWN, RB .
ANALYTICAL CHEMISTRY, 1991, 63 (17) :1666-1672
[7]   PROTEIN EXTRACTIONS WITH HOLLOW FIBERS [J].
DAHURON, L ;
CUSSLER, EL .
AICHE JOURNAL, 1988, 34 (01) :130-136
[8]   DETERMINATION AND APPLICATION OF ION-SELECTIVE ELECTRODE MODEL PARAMETERS USING FLOW-INJECTION AND SIMPLEX OPTIMIZATION [J].
DEVITERI, FJS ;
DIAMOND, D .
ANALYST, 1994, 119 (05) :749-758
[9]  
DURLSELEN LF, 1988, ANAL CHEM, V60, P1455
[10]   PLASTICIZERS FOR LIQUID POLYMERIC MEMBRANES OF ION-SELECTIVE CHEMICAL SENSORS [J].
EUGSTER, R ;
ROSATZIN, T ;
RUSTERHOLZ, B ;
AEBERSOLD, B ;
PEDRAZZA, U ;
RUEGG, D ;
SCHMID, A ;
SPICHIGER, UE ;
SIMON, W .
ANALYTICA CHIMICA ACTA, 1994, 289 (01) :1-13