Coronary artery calcium: Absolute quantification in nonenhanced and contrast-enhanced multi-detector row CT studies

被引:146
作者
Hong, C
Becker, CR
Schoepf, UJ
Ohnesorge, B
Bruening, R
Reiser, MF
机构
[1] Univ Munich, Klinikum Grosshadern, Dept Clin Radiol, D-81377 Munich, Germany
[2] Huazhong Univ Sci & Technol, Dept Radiol, Wuhan 430074, Peoples R China
[3] Siemens Med Syst Inc, Forchheim, Germany
关键词
coronary vessels; calcification; phantoms;
D O I
10.1148/radiol.2232010919
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE: (a) To determine the accuracy of multi-detector row computed tomography (CT) in the measurement of the calcium concentration in a cardiac CT calibration phantom and (b) to assess the correlation of a traditional 3-mm section width CT coronary screening protocol and a 1.25-mm section width CT angiography imaging protocol in the quantification of the absolute mass of coronary calcium in patients who underwent both coronary screening and CT angiography with a multi-detector row CT scanner. MATERIALS AND METHODS: A heart phantom containing calcified cylinders was scanned to determine calibration factors and absolute calcium mass. In 50 patients, the variability (value 1 - value 2/mean value 1 - value 2), limit of agreement (+/-2SD value I - value 2), and systematic error (mean value I - value 2) of the total amount of coronary calcium calculated at traditional 3-mm section width CT and at 1.25-mm section width CT angiography were determined. RESULTS: The correlation coefficient between the 3-mm section width, nonenhanced protocol and the 1.25-mm section width CT angiography protocol was very high (r = 0.977) and the mean variability was low (19.7%) for the absolute mass. There was a systematic error of -6.7 mg and a limit of agreement between 45.0 mg and -58.5 mg. CONCLUSION: Use of the mass quantification algorithm in combination with a calibration phantom allows accurate quantification of coronary calcium. Measurements of calcium mass obtained at 1,25-mm section width CT angiography have the best agreement with those obtained at the traditional 3-mm section width imaging protocol. (C) RSNA, 2002.
引用
收藏
页码:474 / 480
页数:7
相关论文
共 27 条
[1]   QUANTIFICATION OF CORONARY-ARTERY CALCIUM USING ULTRAFAST COMPUTED-TOMOGRAPHY [J].
AGATSTON, AS ;
JANOWITZ, WR ;
HILDNER, FJ ;
ZUSMER, NR ;
VIAMONTE, M ;
DETRANO, R .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1990, 15 (04) :827-832
[2]   Helical and single-slice conventional CT versus electron beam CT for the quantification of coronary artery calcification [J].
Becker, CR ;
Jakobs, TF ;
Aydemir, S ;
Becker, A ;
Knez, A ;
Schoepf, UJ ;
Bruening, R ;
Haberl, R ;
Reiser, MF .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2000, 174 (02) :543-547
[3]   Coronary artery calcium measurement: Agreement of multirow detector and electron beam CP [J].
Becker, CR ;
Kleffel, T ;
Crispin, A ;
Knez, A ;
Young, J ;
Schoepf, UJ ;
Haberl, R ;
Reiser, MF .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2001, 176 (05) :1295-1298
[4]   SMALL LESIONS IN THE HEART IDENTIFIED AT ELECTRON-BEAM CT - CALCIFICATION OR NOISE [J].
BIELAK, LF ;
KAUFMANN, RB ;
MOLL, PP ;
MCCOLLOUGH, CH ;
SCHWARTZ, RS ;
SHEEDY, PF .
RADIOLOGY, 1994, 192 (03) :631-636
[5]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[6]   CORONARY-ARTERY CALCIFICATION DETECTED WITH ULTRAFAST CT AS AN INDICATION OF CORONARY-ARTERY DISEASE - WORK IN PROGRESS [J].
BREEN, JF ;
SHEEDY, PF ;
SCHWARTZ, RS ;
STANSON, AW ;
KAUFMANN, RB ;
MOLL, PP ;
RUMBERGER, JA .
RADIOLOGY, 1992, 185 (02) :435-439
[7]   Measurement of coronary artery calcium with dual-slice helical CT compared with coronary angiography: Evaluation of CT scoring methods, interobserver variations, and reproducibility [J].
Broderick, LS ;
Shemesh, J ;
Wilensky, RL ;
Eckert, GJ ;
Zhou, XH ;
Torres, WE ;
Balk, MA ;
Rogers, WJ ;
Conces, DJ ;
Kopecky, KK .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1996, 167 (02) :439-444
[8]   Sensitivity of two electron beam tomography protocols for the detection and quantification of coronary artery calcium [J].
Callister, T ;
Janowitz, W ;
Raggi, P .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2000, 175 (06) :1743-1746
[9]   Coronary artery disease: Improved reproducibility of calcium scoring with an electron-beam CT volumetric method [J].
Callister, TQ ;
Cooil, B ;
Raya, SP ;
Lippolis, NJ ;
Russo, DJ ;
Raggi, P .
RADIOLOGY, 1998, 208 (03) :807-814
[10]  
Detrano R, 1995, Am J Card Imaging, V9, P167