Specific and targeted detection of viable Escherichia coli O157:H7 using a sensitive and reusable impedance biosensor with dose and time response studies

被引:42
作者
Dweik, Majed [1 ]
Stringer, R. Cody [1 ]
Dastider, Shibajyoti Ghosh [2 ]
Wu, Yifan [2 ]
Almasri, Mahmoud [2 ]
Barizuddin, Syed [1 ]
机构
[1] Lincoln Univ, Jefferson City, MO 65101 USA
[2] Univ Missouri, Dept Elect & Comp Engn, Columbia, MO 65211 USA
基金
美国国家科学基金会;
关键词
Impedance sensing; E. coli O157:H7; Interdigitated microelectrodes (IDEA); ESCHERICHIA-COLI O157-H7; IN-GROUND BEEF; RAPID DETECTION; IMMUNOMAGNETIC-SEPARATION; ELECTROCHEMICAL DETECTION; PATHOGENIC BACTERIA; SENSOR; PCR; SPECTROSCOPY; ELECTRODES;
D O I
10.1016/j.talanta.2012.02.056
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A gold interdigitated microelectrode (IME) impedance biosensor was fabricated for the detection of viable Escherichia coli O157:H7. This sensor was fabricated using lithography techniques. The surface of the electrode was immobilized with anti-E. coli IgG antibodies. This approach is different from other studies where the change in impedance is measured in terms of growth of bacteria on the electrode, rather then the antibody/antigen bonding. The impedance values were recorded for frequency ranges between 100 Hz and 10 MHz. The working range of the dose response for this device was found to be between 2.5 x 10(4) CFU ml(-1) and 2.5 x 10(7) CFU ml(-1). The time response studies indicated that antibody/antigen binding is not a function of time, but can decrease if excess times are allowed for binding. It was observed that the impedance values for 60 min antibody/antigen binding were higher than the impedance values for 120 min binding time. The main advantages of the reported device are that, it provides for both qualitative and quantitative detection in 3 h while other impedance sensors reported earlier may take up to 24 h for detection. If enrichment steps are required then it may take 3-4 days to infer the results. This sensor can be used to detect different types of bacteria by immobilizing the antigen specific antibody. Most of the sensors are not reusable since they either use enzymes or enrichment steps for detection but this device can be reused, following a cleaning protocol which is easy to follow. Each device was used at least five times. The simplicity of this sensor and the ease of fabrication make this sensor a useful alternate to the microfluidics and enzyme based impedance sensors, which are relatively more difficult to fabricate, need programmable fluidic injection pumps to push the sample through the channel, suffer from limitation of coagulation and are difficult to clean. (C) 2012 Elsevier B.V. All rights reserved.
引用
收藏
页码:84 / 89
页数:6
相关论文
共 39 条
[1]   Biosensing at disk microelectrode arrays. Inter-electrode functionalisation allows formatting into miniaturised sensing platforms of enhanced sensitivity [J].
Baldrich, Eva ;
Javier del Campo, Fco. ;
Xavier Munoz, Francesc .
BIOSENSORS & BIOELECTRONICS, 2009, 25 (04) :920-926
[2]   Automated Targeting of Cells to Electrochemical Electrodes Using a Surface Chemistry Approach for the Measurement of Quantal Exocytosis [J].
Barizuddin, Syed ;
Liu, Xin ;
Mathai, Joseph C. ;
Hossain, Maruf ;
Gillis, Kevin D. ;
Gangopadhyay, Shubhra .
ACS CHEMICAL NEUROSCIENCE, 2010, 1 (09) :590-597
[3]   Detection of pathogenic Bacteria by Electrochemical Impedance Spectroscopy: Influence of the immobilization strategies on the sensor performance [J].
Barreiros dos Santos, M. ;
Sporer, C. ;
Sanvicens, N. ;
Pascual, N. ;
Errachid, A. ;
Martinez, E. ;
Marco, M. -P. ;
Teixeira, V. ;
Samiter, J. .
PROCEEDINGS OF THE EUROSENSORS XXIII CONFERENCE, 2009, 1 (01) :1291-+
[4]   Electrical detection and characterization of bacterial adhesion using electrochemical impedance spectroscopy-based flow chamber [J].
Bayoudh, Sonia ;
Othmane, Ali ;
Ponsonnet, Laurence ;
Ben Ouada, Hafedh .
COLLOIDS AND SURFACES A-PHYSICOCHEMICAL AND ENGINEERING ASPECTS, 2008, 318 (1-3) :291-300
[5]   Microelectrodes for the measurement of cellular metabolism [J].
Burdallo, I. ;
Jimenez-Jorquera, C. .
PROCEEDINGS OF THE EUROSENSORS XXIII CONFERENCE, 2009, 1 (01) :289-292
[6]   Gold screen-printed-based impedimetric immunobiosensors for direct and sensitive Escherichia coli quantisation [J].
Escamilla-Gomez, Vanessa ;
Campuzano, Susana ;
Pedrero, Maria ;
Pingarron, Jose M. .
BIOSENSORS & BIOELECTRONICS, 2009, 24 (11) :3365-3371
[7]   Detection of E. coli O157:H7 in raw ground beef by Pathatrix™ immunomagnetic-separation, real-time PCR and cultural methods [J].
Fedio, Willis M. ;
Jinneman, Karen C. ;
Yoshitomi, Ken J. ;
Zapata, Ruben ;
Wendakoon, Chitra N. ;
Browning, Paul ;
Weagant, Stephen D. .
INTERNATIONAL JOURNAL OF FOOD MICROBIOLOGY, 2011, 148 (02) :87-92
[8]   Development of an evanescent-field fibre optic sensor for Escherichia coli O157:H7 [J].
Ferreira, AP ;
Werneck, MM ;
Ribeiro, RM .
BIOSENSORS & BIOELECTRONICS, 2001, 16 (06) :399-408
[9]   Use of real-time polymerase chain reaction and molecular beacons for the detection of Escherichia coli O157:H7 [J].
Fortin, NY ;
Mulchandani, A ;
Chen, W .
ANALYTICAL BIOCHEMISTRY, 2001, 289 (02) :281-288
[10]   Rapid detection of Escherichia coli O157:H7 by immunomagnetic separation and real-time PCR [J].
Fu, Z ;
Rogelj, S ;
Kieft, TL .
INTERNATIONAL JOURNAL OF FOOD MICROBIOLOGY, 2005, 99 (01) :47-57