ICD-antiarrhythmic drug and ICD-pacemaker interactions

被引:35
作者
Brode, SE [1 ]
Schwartzman, D [1 ]
Callans, DJ [1 ]
Gottlieb, CD [1 ]
Marchlinski, FE [1 ]
机构
[1] ALLEGHENY UNIV HOSP,MCP DIV,SCH HLTH SCI,CLIN ELECTROPHYSIOL LAB,PHILADELPHIA,PA 19129
关键词
arrhythmia; implantable cardioverter defibrillator; pacemaker; antiarrhythmic drug;
D O I
10.1111/j.1540-8167.1997.tb00842.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Implantable Cardioverter Defibrillators. Antiarrhythmic drugs and separate bradycardia pacing systems are prescribed commonly in patients with implantable cardioverter defibrillators (ICDs), Adverse effects of antiarrhythmic drugs on ICD function and adverse interactions between ICDs and pacemakers have been documented The effect of antiarrhythmic drugs on the defibrillation threshold (DFT) in patients has not been well assessed, Most studies have been performed in animal models in which cardiac function was normal and drug doses were supraphysiologic, In addition, most studies have utilized monophasic defibrillation shock waveforms and epicardial lead systems, Despite the lack of clinical data applicable to current defibrillation systems, it appears that chronic amiodarone administration causes a significant DFT increase, In addition, antiarrhythmic drugs can influence antitachycardia pacing and tachycardia sensing, Defibrillation shocks can cause transient failure of pacemaker sensing, and pacing, and cause spurious pacemaker reprogramming, Pacemaker function can result in ICD oversensing, leading to inappropriate therapy, or cause ICD undersensing, potentially resulting in failure to deliver therapy during ventricular fibrillation, The susceptibility of an ICD to undersensing appears related to the amplitude of the pacing stimulus artifact recorded by the IGD rate-sensing circuit and to the characteristics of the fibrillation electrogram, Preliminary data suggest that undersensing of ventricular fibrillation by current ICDs is an unlikely event.
引用
收藏
页码:830 / 842
页数:13
相关论文
共 100 条
[51]   UNIPOLAR PACER ARTIFACTS INDUCED FAILURE OF AN AUTOMATIC IMPLANTABLE CARDIOVERTER DEFIBRILLATOR TO DETECT VENTRICULAR-FIBRILLATION [J].
KIM, SG ;
FURMAN, S ;
WASPE, LE ;
BRODMAN, R ;
FISHER, JD .
AMERICAN JOURNAL OF CARDIOLOGY, 1986, 57 (10) :880-881
[52]   AUTOMATIC IMPLANTABLE CARDIOVERTER DEFIBRILLATOR - INADVERTENT DISCHARGES DURING PERMANENT PACEMAKER MAGNET TESTS [J].
KIM, SG ;
FURMAN, S ;
MATOS, JA ;
WASPE, LE ;
BRODMAN, R ;
FISHER, JD .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1987, 10 (03) :579-582
[53]  
KOPP D, 1995, PACE, V18, P872
[54]   EPICARDIAL ADMINISTRATION OF IBUTILIDE FROM POLYURETHANE MATRICES - EFFECTS ON DEFIBRILLATION THRESHOLD AND ELECTROPHYSIOLOGIC PARAMETERS [J].
LABHASETWAR, V ;
UNDERWOOD, T ;
HEIL, RW ;
GALLAGHER, M ;
LANGBERG, J ;
LEVY, RJ .
JOURNAL OF CARDIOVASCULAR PHARMACOLOGY, 1994, 24 (05) :826-840
[55]   PROTECTION OF IMPLANTED PACEMAKERS FROM EXCESSIVE ELECTRICAL ENERGY OF DC SHOCK [J].
LAU, FYK ;
BILITCH, M ;
WINTROUB, HJ .
AMERICAN JOURNAL OF CARDIOLOGY, 1969, 23 (02) :244-&
[56]   REDUCTION IN DEFIBRILLATOR SHOCKS WITH AN IMPLANTABLE DEVICE COMBINING ANTITACHYCARDIA PACING AND SHOCK THERAPY [J].
LEITCH, JW ;
GILLIS, AM ;
WYSE, DG ;
YEE, R ;
KLEIN, GJ ;
GUIRAUDON, G ;
SHELDON, RS ;
DUFF, HJ ;
KIESER, TM ;
MITCHELL, LB .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1991, 18 (01) :145-151
[57]   ADVERSE ACUTE AND CHRONIC EFFECTS OF ELECTRICAL DEFIBRILLATION AND CARDIOVERSION ON IMPLANTED UNIPOLAR CARDIAC PACING SYSTEMS [J].
LEVINE, PA ;
BAROLD, SS ;
FLETCHER, RD ;
TALBOT, P .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1983, 1 (06) :1413-1422
[58]  
LEVINE PA, 1985, MODERN CARDIAC PACIN, P875
[59]  
LUCERI RM, 1985, PACE, V8, P296
[60]  
Man David C., 1996, Journal of the American College of Cardiology, V27, p97A