Real-time in vivo dosimetry using micro-MOSFET detectors during intraoperative electron beam radiation therapy in early-stage breast cancer

被引:40
作者
Ciocca, M
Piazzi, V
Lazzari, R
Vavassori, A
Luini, A
Veronesi, P
Galimberti, V
Intra, M
Guido, A
Tosi, G
Veronesi, U
Orecchia, R
机构
[1] European Inst Oncol, Dept Med Phys, I-20141 Milan, Italy
[2] European Inst Oncol, Div Radiat Oncol, I-20141 Milan, Italy
[3] European Inst Oncol, Div Senol, I-20141 Milan, Italy
关键词
in vivo dosimetry; MOSFET; IORT; breast cancer; mobile linear accelerator; electrons;
D O I
10.1016/j.radonc.2005.11.011
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: In a previous paper we reported the results of off-line in vivo measurements using radiochromic films in IOERT. In the present study, a further step was made, aiming at the improvement of the effectiveness of in vivo dosimetry, based on a real-time check of the dose. Materials and methods: Entrance dose was determined using micro-MOSFET detectors placed inside a thin, sterile, transparent catheter. The epoxy side of the detector was faced towards the beam to minimize the anisotropy. Each detector was plugged into a bias supply (standard sensitivity) and calibrated at 5 Gy using 6 MeV electrons produced by a conventional linac. Detectors were characterized in terms of linearity, precision and dose per pulse dependence. No energy and temperature dependence was found. The sensitivity change of detectors was about 1% per 20 Gy accumulated dose. Correction factors to convert surface to entrance dose were determined for each combination of energy and applicator. From November 2004 to May 2005, in vivo dosimetry was performed on 45 patients affected by early-stage breast cancer, who underwent IOERT to the tumour bed. IOERT was delivered using electrons (4-10 MeV) at high dose per pulse, produced by either a Novac7 or a Liac mobile linac. Results: The mean ratio between measured and expected dose was 1.006 +/- 0.035 (1 SD), in the range 0.92-1.1. The procedure uncertainty was 3.6%. Micro-MOSFETs appeared suitable for in vivo dosimetry in IOERT, although some unfavourable aspects, like the limited lifetime and the anisotropy with no build-up, were found. Prospectively, a realtime action level (+/- 6%) on dose discrepancy was defined. Conclusions: Excellent agreement between measured and expected doses was found. Real-time in vivo dosimetry appeared feasible, reliable and more effective than the method previously published. (C) 2005 Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:213 / 216
页数:4
相关论文
共 22 条
[1]   Preliminary evaluation of implantable MOSFET radiation dosimeters [J].
Beddar, AS ;
Salehpour, M ;
Briere, TM ;
Hamidian, H ;
Gillin, MT .
PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (01) :141-149
[2]   Clinical dosimetry with MOSFET dosimeters to determine the dose along the field junction in a split beam technique [J].
Bloemen-van Gurp, E ;
du Bois, W ;
Visser, P ;
Bruinvis, I ;
Jalink, D ;
Hermans, J ;
Lambin, P .
RADIOTHERAPY AND ONCOLOGY, 2003, 67 (03) :351-357
[3]   Peripheral dose measurement with a MOSFET detector [J].
Butson, MJ ;
Cheung, T ;
Yu, PKN .
APPLIED RADIATION AND ISOTOPES, 2005, 62 (04) :631-634
[4]   Effects of temperature variation on MOSFET dosimetry [J].
Cheung, T ;
Butson, MJ ;
Yu, PKN .
PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (13) :N191-N196
[5]   Investigation of the use of MOSFET for clinical IMRT dosimetric verification [J].
Chuang, CF ;
Verhey, LJ ;
Xia, P .
MEDICAL PHYSICS, 2002, 29 (06) :1109-1115
[6]   In vivo dosimetry using radiochromic films during intraoperative electron beam radiation therapy in early-stage breast cancer [J].
Ciocca, M ;
Orecchia, R ;
Garibaldi, C ;
Rondi, E ;
Luini, A ;
Gatti, G ;
Intra, M ;
Veronesi, P ;
Lazzari, R ;
Tosi, G ;
Veronesi, U .
RADIOTHERAPY AND ONCOLOGY, 2003, 69 (03) :285-289
[7]   Application of commercial MOSFET detectors for in vivo dosimetry in the therapeutic x-ray range from 80 kV to 250 W [J].
Ehringfeld, C ;
Schmid, S ;
Poljanc, K ;
Kirisits, C ;
Aiginger, H ;
Georg, D .
PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (02) :289-303
[8]  
*I SUP SAN, 2003, 031 ISTISAN EN ISS
[9]  
International Atomic Energy Agency, 2000, IAEA TECHN REP SER, V398
[10]   Comparison study of MOSFET detectors and diodes for entrance in vivo dosimetry in 18 MV x-ray beams [J].
Jornet, N ;
Carrasco, P ;
Jurado, D ;
Ruiz, A ;
Eudaldo, T ;
Ribas, M .
MEDICAL PHYSICS, 2004, 31 (09) :2534-2542