Investigation of the use of MOSFET for clinical IMRT dosimetric verification

被引:117
作者
Chuang, CF [1 ]
Verhey, LJ [1 ]
Xia, P [1 ]
机构
[1] Univ Calif San Francisco, Ctr Comprehens Canc, Dept Radiat Oncol, San Francisco, CA 94143 USA
关键词
intensity modulated radiotherapy; MOSFET detector; dosimetric verification; phantom measurement;
D O I
10.1118/1.1481520
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
With advanced conformal radiotherapy using intensity modulated beams, it is important to have radiation dose verification measurements prior to treatment. Metal oxide semiconductor field effect transistors (MOSFET) have the advantage of a faster and simpler reading procedure compared to thermoluminescent dosimeters (TLD), and with the commercial MOSFET system, multiple detectors can be used simultaneously. In addition, the small size of the detector could be advantageous, especially for point dose measurements in small homogeneous dose regions. To evaluate the feasibility of MOSFET for routine IMRT dosimetry, a comprehensive set of experiments has been conducted, to investigate the stability, linearity, energy, and angular dependence. For a period of two weeks, under a standard measurement setup, the measured dose standard deviation using the MOSFETs was +/-0.015 Gy with the mean dose being 1.00 Gy. For a measured dose range of 0.3 Gy to 4.2 Gy, the MOSFETs present a linear response, with a linearity coefficient of 0.998. Under a 10x10 cm(2) square field, the dose variations measured by the MOSFETs for every 10 degrees from 0 to 180 degrees is +/-2.5%. The percent depth dose (PDD) measurements were used to verify the energy dependence. The measured PDD using the MOSFETs from 0.5 cm to 34 cm depth agreed to within +/-3% when compared to that of the ionization chamber. For IMRT dose verification, two special phantoms were designed. One is a solid water slab with 81 possible MOSFET placement holes, and another is a cylindrical phantom with 48 placement holes. For each IMRT phantom verification, an ionization chamber and 3 to 5 MOSFETs were used to measure multiple point doses at different locations. Preliminary results show that the agreement between dose measured by MOSFET and that calculated by Corvus is within 5% error, while the agreement between ionization chamber measurement and the calculation is within 3% error. In conclusion, MOSFET detectors are suitable for routine IMRT dose verification. (C) 2002 American Association of Physicists in Medicine.
引用
收藏
页码:1109 / 1115
页数:7
相关论文
共 11 条
[1]   A new radiotherapy surface dose detector: The MOSFET [J].
Butson, MJ ;
Rozenfeld, A ;
Mathur, JN ;
Carolan, M ;
Wong, TPY ;
Metcalfe, PE .
MEDICAL PHYSICS, 1996, 23 (05) :655-658
[2]   The response of a MOSFET, p-type semiconductor and LiF TLD to quasi-monoenergetic x-rays [J].
Edwards, CR ;
Green, S ;
Palethorpe, JE ;
Mountford, PJ .
PHYSICS IN MEDICINE AND BIOLOGY, 1997, 42 (12) :2383-2391
[3]   AUTOMATED DATA-COLLECTION AND ANALYSIS SYSTEM FOR MOSFET RADIATION DETECTORS [J].
GLADSTONE, DJ ;
CHIN, LM .
MEDICAL PHYSICS, 1991, 18 (03) :542-548
[4]   Improved spatial resolution by MOSFET dosimetry of an x-ray microbeam [J].
Kaplan, GI ;
Rosenfeld, AB ;
Allen, BJ ;
Booth, JT ;
Carolan, MG ;
Holmes-Siedle, A .
MEDICAL PHYSICS, 2000, 27 (01) :239-244
[5]   Dose response of various radiation detectors to synchrotron radiation [J].
Kron, T ;
Duggan, L ;
Smith, T ;
Rosenfeld, A ;
Butson, M ;
Kaplan, G ;
Howlett, S ;
Hyodo, K .
PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (11) :3235-3259
[6]   Evaluation of a MOSFET radiation sensor for the measurement of entrance surface dose in diagnostic radiology [J].
Peet, DJ ;
Pryor, MD .
BRITISH JOURNAL OF RADIOLOGY, 1999, 72 (858) :562-568
[7]   Measurement of radiotherapy x-ray skin dose on a chest wall phantom [J].
Quach, KY ;
Morales, J ;
Butson, MJ ;
Rosenfeld, AB ;
Metcalfe, PE .
MEDICAL PHYSICS, 2000, 27 (07) :1676-1680
[8]   Clinical dosimetry using MOSFETS [J].
Ramani, R ;
Russell, S ;
OBrien, P .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1997, 37 (04) :959-964
[9]   Calibration of a MOSFET detection system for 6-MV in vivo dosimetry [J].
Scalchi, P ;
Francescon, P .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1998, 40 (04) :987-993
[10]   EVALUATION OF A DUAL BIAS DUAL METAL-OXIDE-SILICON SEMICONDUCTOR FIELD-EFFECT TRANSISTOR DETECTOR AS RADIATION DOSIMETER [J].
SOUBRA, M ;
CYGLER, J ;
MACKAY, G .
MEDICAL PHYSICS, 1994, 21 (04) :567-572