Optimal interferometer designs for optical coherence tomography

被引:185
作者
Rollins, AM [1 ]
Izatt, JA [1 ]
机构
[1] Case Western Reserve Univ, Dept Biomed Engn & Med, Cleveland, OH 44106 USA
关键词
D O I
10.1364/OL.24.001484
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
We introduce a family of power-conserving fiber-optic interferometer designs for low-coherence reflectometry that use optical circulators, unbalanced couplers, and (or) balanced heterodyne detection. Simple design equations for optimization of the signal-to-noise ratio of the interferometers are expressed in terms of relevant signal and noise sources and measurable system parameters. We use the equations to evaluate the expected performance of the new configurations compared with that of the standard Michelson interferometer that is commonly used in optical coherence tomography (OCT) systems. The analysis indicates that improved sensitivity is expected for all the new interferometer designs, compared with the sensitivity of the standard OCT interferometer, under high-speed imaging conditions. (C) 1999 Optical Society of America.
引用
收藏
页码:1484 / 1486
页数:3
相关论文
共 10 条
[1]   Power-efficient nonreciprocal interferometer and linear-scanning fiber-optic catheter for optical coherence tomography [J].
Bouma, BE ;
Tearney, GJ .
OPTICS LETTERS, 1999, 24 (08) :531-533
[2]   FEMTOSECOND TRANSILLUMINATION OPTICAL COHERENCE TOMOGRAPHY [J].
HEE, MR ;
IZATT, JA ;
JACOBSON, JM ;
FUJIMOTO, JG ;
SWANSON, EA .
OPTICS LETTERS, 1993, 18 (12) :950-952
[3]   STATISTICS OF THERMAL AND LASER RADIATION [J].
HODARA, H .
PROCEEDINGS OF THE INSTITUTE OF ELECTRICAL AND ELECTRONICS ENGINEERS, 1965, 53 (07) :696-&
[4]   OPTICAL COHERENCE TOMOGRAPHY [J].
HUANG, D ;
SWANSON, EA ;
LIN, CP ;
SCHUMAN, JS ;
STINSON, WG ;
CHANG, W ;
HEE, MR ;
FLOTTE, T ;
GREGORY, K ;
PULIAFITO, CA ;
FUJIMOTO, JG .
SCIENCE, 1991, 254 (5035) :1178-1181
[5]   NOISE CHARACTERISTICS OF HIGH-POWER DOPED-FIBER SUPERLUMINESCENT SOURCES [J].
MORKEL, PR ;
LAMING, RI ;
PAYNE, DN .
ELECTRONICS LETTERS, 1990, 26 (02) :96-98
[6]  
Oliver B M., 1961, Proc. IEEE, V49, n, P1960
[7]   Noise analysis of a combined optical coherence tomograph and a confocal scanning ophthalmoscope [J].
Podoleanu, AG ;
Jackson, DA .
APPLIED OPTICS, 1999, 38 (10) :2116-2127
[8]   Real-time in vivo imaging of human gastrointestinal ultrastructure by use of endoscopic optical coherence tomography with a novel efficient interferometer design [J].
Rollins, AM ;
Ung-arunyawee, R ;
Chak, A ;
Wong, RCK ;
Kobayashi, K ;
Sivak, MV ;
Izatt, JA .
OPTICS LETTERS, 1999, 24 (19) :1358-1360
[9]   A SIMPLE INTENSITY NOISE-REDUCTION TECHNIQUE FOR OPTICAL LOW-COHERENCE REFLECTOMETRY [J].
SORIN, WV ;
BANEY, DM .
IEEE PHOTONICS TECHNOLOGY LETTERS, 1992, 4 (12) :1404-1406
[10]   Noise in optical low-coherence reflectometry [J].
Takada, K .