Noise analysis of a combined optical coherence tomograph and a confocal scanning ophthalmoscope

被引:54
作者
Podoleanu, AG [1 ]
Jackson, DA [1 ]
机构
[1] Univ Kent, Sch Phys Sci, Appl Opt Grp, Canterbury CT2 7NR, Kent, England
关键词
D O I
10.1364/AO.38.002116
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
A fiberized optical coherence tomographic (OCT) system is modified to produce a confocal image similar to that produced by a scanning laser ophthalmoscope. Two possible configurations are presented, one that can deliver either an OCT or a confocal image and another that is capable of producing both the OCT and the confocal images simultaneously. Using the later configuration, we demonstrate such images from the retina in the living eye. The penalty in terms of performance reduction introduced into the optical coherence tomograph when integrated with a confocal receiver and the signal-to-noise ratio analysis for the different confocal receiver configurations are presented. (C) 1999 Optical Society of America.
引用
收藏
页码:2116 / 2127
页数:12
相关论文
共 33 条
[1]  
[Anonymous], 1990, CONFOCAL MICROSCOPY
[2]  
*ANSI, 1986, Z1361 ANSI
[3]   HIGH-RESOLUTION OPTICAL COHERENCE TOMOGRAPHIC IMAGING USING A MODE-LOCKED TI-AL2O3 LASER SOURCE [J].
BOUMA, B ;
TEARNEY, GJ ;
BOPPART, SA ;
HEE, MR ;
BREZINSKI, ME ;
FUJIMOTO, JG .
OPTICS LETTERS, 1995, 20 (13) :1486-1488
[4]  
Corle T.R., 1996, Confocal scanning optical microscopy and related imaging systems
[5]   SPECTRAL REFLECTANCE OF THE HUMAN OCULAR FUNDUS [J].
DELORI, FC ;
PFLIBSEN, KP .
APPLIED OPTICS, 1989, 28 (06) :1061-1077
[6]   ACTIVE OPTICAL DEPTH RESOLUTION IMPROVEMENT OF THE LASER TOMOGRAPHIC SCANNER [J].
DREHER, AW ;
BILLE, JF ;
WEINREB, RN .
APPLIED OPTICS, 1989, 28 (04) :804-808
[7]   OPTICAL COHERENCE TOMOGRAPHY [J].
HUANG, D ;
SWANSON, EA ;
LIN, CP ;
SCHUMAN, JS ;
STINSON, WG ;
CHANG, W ;
HEE, MR ;
FLOTTE, T ;
GREGORY, K ;
PULIAFITO, CA ;
FUJIMOTO, JG .
SCIENCE, 1991, 254 (5035) :1178-1181
[8]   SCANNING INTERFERENCE AND CONFOCAL MICROSCOPY [J].
JUSKAITIS, R ;
WILSON, T .
JOURNAL OF MICROSCOPY-OXFORD, 1994, 176 :188-194
[9]   Comparative study of confocal and heterodyne microscopy for imaging through scattering media [J].
Kempe, M ;
Rudolph, W ;
Welsch, E .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1996, 13 (01) :46-52
[10]   CONFOCAL SCANNING OPTICAL MICROSCOPE USING SINGLE-MODE FIBER FOR SIGNAL-DETECTION [J].
KIMURA, S ;
WILSON, T .
APPLIED OPTICS, 1991, 30 (16) :2143-2150