Electrical conductivity imaging using gradient Bz decomposition algorithm in magnetic resonance electrical impedance tomography (MREIT)

被引:74
作者
Park, C
Kwon, O
Woo, EJ
Seo, JK
机构
[1] Kyung Hee Univ, Coll Elect & Informat, Yongin 449701, Kyungki, South Korea
[2] Konkuk Univ, Dept Math, Seoul 143701, South Korea
[3] Yonsei Univ, Dept Math, Seoul 120749, South Korea
关键词
conductivity image; gradient B-z decomposition algorithm; magnetic flux density; MREIT;
D O I
10.1109/TMI.2004.824228
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
In magnetic resonance electrical impedance tomography (MREIT), we try to visualize cross-sectional conductivity (or resistivity) images of a subject. We inject electrical currents into the subject through surface electrodes and measure the z component B. of the induced internal magnetic flux density using an MRI scanner. Here, z is the direction of the main magnetic field of the MRI scanner. We formulate the conductivity image reconstruction problem in MREIT from a careful analysis of the relationship between the injection current and the induced magnetic flux density B-z. Based on the novel mathematical formulation, we propose the gradient B-z decomposition algorithm to reconstruct conductivity images. This new algorithm needs to differentiate B-z only once in contrast to the previously developed harmonic B-z algorithm where the numerical computation of del(2)B(z) is required. The new algorithm, therefore, has the important advantage of much improved noise tolerance. Numerical simulations with added random noise of realistic amounts show the feasibility of the algorithm in practical applications and also its robustness against measurement noise.
引用
收藏
页码:388 / 394
页数:7
相关论文
共 18 条
[1]  
Birgül Ö, 2003, PHYS MED BIOL, V48, P653, DOI 10.1088/0031-9155/48/5/307
[2]   Imaging with electricity: Report of the European Concerted Action on Impedance Tomography [J].
Boone, K ;
Barber, D ;
Brown, B .
JOURNAL OF MEDICAL ENGINEERING & TECHNOLOGY, 1997, 21 (06) :201-232
[3]   Uniqueness and reconstruction in magnetic resonance-electrical impedance tomography (MR-EIT) [J].
Ider, YZ ;
Onart, S ;
Lionheart, WRB .
PHYSIOLOGICAL MEASUREMENT, 2003, 24 (02) :591-604
[4]  
Ider YZ., 1998, ELEKTRIK, V6, P215
[5]   J-substitution algorithm in Magnetic Resonance Electrical Impedance Tomography (MREIT):: Phantom experiments for static resistivity images [J].
Khang, HS ;
Lee, BI ;
Oh, SH ;
Woo, EJ ;
Lee, SY ;
Cho, MY ;
Kwon, O ;
Yoon, JR ;
Seo, JK .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2002, 21 (06) :695-702
[6]   Magnetic resonance electrical impedance tomography (MREIT):: Simulation study of J-substitution algorithm [J].
Kwon, O ;
Woo, EJ ;
Yoon, JR ;
Seo, JK .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2002, 49 (02) :160-167
[7]   Static resistivity image of a cubic saline phantom in magnetic resonance electrical impedance tomography (MREIT) [J].
Lee, BI ;
Oh, SH ;
Woo, EJ ;
Lee, SY ;
Cho, MH ;
Kwon, O ;
Seo, JK ;
Baek, WS .
PHYSIOLOGICAL MEASUREMENT, 2003, 24 (02) :579-589
[8]   Three-dimensional forward solver and its performance analysis for magnetic resonance electrical impedance tomography (MREIT) using recessed electrodes [J].
Lee, BI ;
Oh, SH ;
Woo, EJ ;
Lee, SY ;
Cho, MH ;
Kwon, O ;
Seo, JK ;
Lee, JY ;
Baek, WS .
PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (13) :1971-1986
[9]   Conductivity and current density image reconstruction using harmonic Bz algorithm in magnetic resonance electrical impedance tomography [J].
Oh, SH ;
Lee, BI ;
Woo, EJ ;
Lee, SY ;
Cho, MH ;
Kwon, O ;
Seo, JK .
PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (19) :3101-3116
[10]   Electrical impedance tomography [J].
Saulnier, GJ ;
Blue, RS ;
Newell, JC ;
Isaacson, D ;
Edic, PM .
IEEE SIGNAL PROCESSING MAGAZINE, 2001, 18 (06) :31-43