Surface degradation features and microstructural properties of ultra-high molecular weight polyethylene (UHMWPe)

被引:13
作者
Cornwall, GB
Hansson, CM
Bowe, AJ
Bryant, JT
机构
[1] QUEENS UNIV, CLIN MECH GRP, KINGSTON, ON K7L 3N6, CANADA
[2] QUEENS UNIV, DEPT MECH ENGN, KINGSTON, ON K7L 3N6, CANADA
基金
加拿大自然科学与工程研究理事会;
关键词
D O I
10.1023/A:1018564412753
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Surface degradation of UHMWPe is recognized as a leading clinical concern, limiting the long-term performance in total knee replacements. Eight retrieved tibial plateaux and six wear screening test samples were evaluated for surface degradation features and microstructural features. The surface degradation features were assessed using stereomicroscopy and scanning electron microscopy. Microstructural features were evaluated using optical microscopy of thin cross-sections and a permanganate etching technique. The pitting mechanism of wear was observed on all eight retrieved TKR and covered an average of 12.6% of the surface area. The size of the pits were similar to the size of grains observed in the microstructural evaluation - approximately 100 to 200 mu m. The presented observations of pitting in retrieved knee implants have shown that the post-processing microstructure may influence this mechanism of surface degradation and hence the wear products.
引用
收藏
页码:303 / 309
页数:7
相关论文
共 37 条
  • [1] *AM SOC TEST MAT, 1994, STAND, V13, P183
  • [2] OBSERVATIONS ON THE SINTERING OF ULTRA-HIGH-MOLECULAR-WEIGHT POLYETHYLENE (UHMWPE) POWDERS
    BARNETSON, A
    HORNSBY, PR
    [J]. JOURNAL OF MATERIALS SCIENCE LETTERS, 1995, 14 (02) : 80 - 84
  • [3] BARTEL DL, 1991, BRIS MYER Z, P61
  • [4] Bassett DC, 1989, COMPREHENSIVE POLYM, V1, P841
  • [5] Birnkraut H. W., 1991, ULTRA HIGH MOL WEIGH, P3
  • [6] BLOEBAUM RD, 1991, CLIN ORTHOP RELAT R, P120
  • [7] POLYETHYLENE WEAR IN UNICONDYLAR KNEE PROSTHESES - 106 RETRIEVED MARMOR, PCA, AND ST-GEORG TIBIAL COMPONENTS COMPARED
    BLUNN, GW
    JOSHI, AB
    LILLEY, PA
    ENGELBRECHT, E
    RYD, L
    LIDGREN, L
    HARDINGE, K
    NIEDER, E
    WALKER, PS
    [J]. ACTA ORTHOPAEDICA SCANDINAVICA, 1992, 63 (03): : 247 - 255
  • [8] BOWE AJ, 1994, THESIS QUEENS U, P56
  • [9] OBSERVATIONS OF RESIDUAL SUBSURFACE SHEAR STRAIN IN THE ULTRAHIGH MOLECULAR-WEIGHT POLYETHYLENE ACETABULAR CUPS OF HIP PROSTHESES
    COOPER, JR
    DOWSON, D
    FISHER, J
    ISAAC, GH
    WROBLEWSKI, BM
    [J]. JOURNAL OF MATERIALS SCIENCE-MATERIALS IN MEDICINE, 1994, 5 (01) : 52 - 57
  • [10] A QUANTITATIVE TECHNIQUE FOR REPORTING SURFACE DEGRADATION PATTERNS OF UHMWPE COMPONENTS OF RETRIEVED TOTAL KNEE REPLACEMENTS
    CORNWALL, GB
    BRYANT, JT
    HANSSON, CM
    RUDAN, J
    KENNEDY, LA
    COOKE, TDV
    [J]. JOURNAL OF APPLIED BIOMATERIALS, 1995, 6 (01) : 9 - 18