A Monte Carlo investigation of the dosimetric characteristics of the VariSource 192Ir high dose rate brachytherapy source

被引:34
作者
Karaiskos, P
Angelopoulos, A
Baras, P
Sakelliou, L
Sandilos, P
Dardoufas, K
Vlachos, L
机构
[1] Univ Athens, Dept Phys, Nucl & Particle Phys Sect, GR-15771 Athens, Greece
[2] Univ Athens, Sch Med, Dept Radiol, Areteion Hosp, GR-11527 Athens, Greece
[3] Hygeia Hosp, Dept Phys Med, Athens, Greece
关键词
VariSource; Ir-192; HDR; brachytherapy; dosimetry; Monte Carlo;
D O I
10.1118/1.598645
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
An analytical Monte Carlo simulation code, incorporating in detail source construction and dimensions, was used to investigate the dosimetric characteristics of the VariSource Ir-192 high dose rate brachytherapy source. Dose-rate constant, radial dose functions, and anisotropy functions, utilized in the AAPM Task Group 43 dose estimation formalism, were calculated with the source centered in a spherical water phantom of 30 cm in diameter. The results, which are in agreement with the corresponding data available in the literature, are compared to those obtained in our previous study for the widely used microSelectron Ir-192 high dose-rate brachytherapy source. The dose-rate constant was found to be equal to 1.043+/-0.005 cGy h(-1) U-1 for the VariSource, compared to a value of 1.116+/-0.006 cGy h(-1) U-1 calculated for the microSelectron. Significant differences in the anisotropy of the two sources are observed only for polar angles close to their long axis and are due to their different dimensions. (C) 1999 American Association of Physicists in Medicine. [S0094-2405(99)03008-4].
引用
收藏
页码:1498 / 1502
页数:5
相关论文
共 13 条
[1]   ACCURATE MONTE-CARLO CALCULATIONS OF THE COMBINED ATTENUATION AND BUILDUP FACTORS, FOR ENERGIES (20-1500 KEV) AND DISTANCES (0-10 CM) RELEVANT IN BRACHYTHERAPY [J].
ANGELOPOULOS, A ;
PERRIS, A ;
SAKELLARIOU, K ;
SAKELLIOU, L ;
SARIGIANNIS, K ;
ZARRIS, G .
PHYSICS IN MEDICINE AND BIOLOGY, 1991, 36 (06) :763-778
[2]   Monte Carlo calculation of dose rate distributions around Ir-192 wires [J].
Ballester, F ;
Hernandez, C ;
PerezCalatayud, J ;
Lliso, F .
MEDICAL PHYSICS, 1997, 24 (08) :1221-1228
[3]   Application of the Monte Carlo integration (MCI) method for calculation of the anisotropy of 192Ir brachytherapy sources [J].
Baltas, D ;
Giannouli, S ;
Garbi, A ;
Diakonos, F ;
Geramani, K ;
Ioannidis, GT ;
Tsalpatouros, A ;
Uzunoglu, N ;
Kolotas, C ;
Zamboglou, N .
PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (06) :1783-1801
[4]   Monte Carlo and TLD dosimetry of an 192Ir high dose-rate brachytherapy source [J].
Karaiskos, P ;
Angelopoulos, A ;
Sakelliou, L ;
Sandilos, P ;
Antypas, C ;
Vlachos, L ;
Koutsouveli, E .
MEDICAL PHYSICS, 1998, 25 (10) :1975-1984
[5]   Dosimetric characteristics of a new high-intensity 192Ir source for remote afterloading [J].
Meigooni, AS ;
Kleiman, MT ;
Johnson, JL ;
Mazloomdoost, D ;
Ibbott, GS .
MEDICAL PHYSICS, 1997, 24 (12) :2008-2013
[6]   Anisotropy of an (192)iridium high dose rate source measured with a miniature ionization chamber [J].
Mishra, V ;
Waterman, FM ;
Suntharalingam, N .
MEDICAL PHYSICS, 1997, 24 (05) :751-755
[7]   ANISOTROPY MEASUREMENTS OF A HIGH-DOSE-RATE IR-192 SOURCE IN AIR AND IN POLYSTYRENE [J].
MULLERRUNKEL, R ;
CHO, SH .
MEDICAL PHYSICS, 1994, 21 (07) :1131-1134
[8]   Perturbation effects in dosimetry .1. Kilovoltage x-rays and electrons [J].
Nahum, AE .
PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (09) :1531-1580
[9]   DOSIMETRY OF INTERSTITIAL BRACHYTHERAPY SOURCES - RECOMMENDATIONS OF THE AAPM RADIATION-THERAPY COMMITTEE TASK GROUP NO 43 [J].
NATH, R ;
ANDERSON, LL ;
LUXTON, G ;
WEAVER, KA ;
WILLIAMSON, JF ;
MEIGOONI, AS .
MEDICAL PHYSICS, 1995, 22 (02) :209-234
[10]   DOSE-RATE DISTRIBUTIONS AROUND CO-60, CS-137, AU-198, IR-192, AM-241, I-125 (MODELS 6702 AND 6711) BRACHYTHERAPY SOURCES AND THE NUCLIDE TC-99(M) [J].
SAKELLIOU, L ;
SAKELLARIOU, K ;
SARIGIANNIS, K ;
ANGELOPOULOS, A ;
PERRIS, A ;
ZARRIS, G .
PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (10) :1859-1872